Doppler echocardiographic measurement of cardiac output using the mitral orifice method.
Cardiac output was determined in 20 patients with various cardiac conditions by measuring the cross sectional area of the mitral orifice by echocardiography and the transmitral flow by the Doppler technique. Cardiac output was calculated by multiplying the corrected mitral orifice area by the maximum diastolic velocity integral recorded by the pulsed mode. The results were compared with that obtained by the Fick method. The correlation for cardiac output by the two techniques was high in the whole group, particularly in patients without mitral regurgitation. There was also a good correlation for stroke volume determined by the two methods. Cardiac output was significantly overestimated by the continuous mode and in patients with mitral regurgitation. These results show that the mitral orifice method provides a new and reliable approach to the non-invasive measurement of cardiac output.
This article has been cited by other articles:
-
Mele, D., Vandervoort, P., Palacios, I., Rivera, J. M., Dinsmore, R. E., Schwammenthal, E., Marshall, J. E., Weyman, A. E., Levine, R. A.
(1995). Proximal Jet Size by Doppler Color Flow Mapping Predicts Severityof Mitral Regurgitation : Clinical Studies. Circulation
91: 746-754
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.