Artefactual ST segment abnormalities due to electrocardiograph design.

D I Tayler, R Vincent

Specifications recommended for electrocardiographs omit any reference to phase characteristics and place undue emphasis on an extended amplitude response. This does not, however, ensure faithful reproduction of the electrocardiogram and may result in less effective attenuation of unwanted noise than can be achieved with modern filters. The separate effects of phase and amplitude response on electrocardiographic signals were compared, and it was shown that distortion of the electrocardiogram by equipment with an inadequate phase response may cause changes resembling those seen in ischaemic heart disease. Case reports are presented in which distortion produced by commercial equipment impeded diagnosis. Specifications recommended as standard for electrocardiographs should be modified to include phase response and allow greater attenuation of frequencies below 0.5 Hz.


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

This Article

Services
Citing Articles
Google Scholar
PubMed
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.