Comparison of impedance cardiography and dye dilution method for measuring cardiac output

W Spiering, P N van Es, P W de Leeuw

Department of Internal Medicine, University Hospital Maastricht, the Netherlands

Correspondence to: Dr de Leeuw, Department of Internal Medicine, University Hospital Maastricht, PO Box 5800, 6202 AZ Maastricht, the Netherlands.

Accepted for publication 1 September 1997

Objective---To assess the degree of agreement between impedance cardiography, using the NCCOM3-R7 device, and the gold standard---the dye dilution method---both under basal conditions and after stimulation of cardiac output.
Patients---35 paired measurements in five healthy male volunteers.
Interventions---To obtain higher levels of cardiac output, cardiac performance was stimulated with a dopamine infusion.
Results---In 35 paired measurements, the mean of all the impedance values was higher than that of the dye dilution values, at 10.2 v 7.4 l/min (p < 0.0001). The mean discrepancy between the two methods was 3.3 l/min, and the mean bias -2.9 l/min, with limits of agreement of -9.0 and 3.2 l/min. A change in cardiac output could not adequately be predicted by the NCCOM3-R7. In 20 of 25 measurements obtained during continuous intravenous dopamine infusions there was a rise in dye dilution cardiac output (range 0.2 to 5.9 l/min). Neither the magnitude nor the direction of the change in dye dilution values corresponded with the change measured by impedance cardiography. The mean discrepancy here between the two methods was 1.8 l/min, and the mean bias -0.8 l/min, with limits of agreement of -4.9 and 3.3 l/min.
Conclusions---In healthy volunteers, impedance cardiography with NCCOM3-R7 is inadequate for assessing cardiac output when compared with the dye dilution method.

Keywords: cardiac output;  impedance cardiography;  dye dilution


© 1998 by Heart

Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

This article has been cited by other articles:

  • Levy, R. J., Chiavacci, R. M., Nicolson, S. C., Rome, J. J., Lin, R. J., Helfaer, M. A., Nadkarni, V. M. (2004). An Evaluation of a Noninvasive Cardiac Output Measurement Using Partial Carbon Dioxide Rebreathing in Children. Anesth. Analg. 99: 1642-1647 [Abstract] [Full Text]  
  • Cooke, C. R., Wall, B. M., Huch, K. M., Mangold, T. (2001). Cardiovascular effects of vasopressin following V1 receptor blockade compared to effects of nitroglycerin. Am. J. Physiol. Regul. Integr. Comp. Physiol. 281: R887-R893 [Abstract] [Full Text]  
  • Singer, M (1998). Cardiac output in 1998. Heart 79: 425-428 [Full Text]  

This Article

Services
Citing Articles
Google Scholar
PubMed
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.