Case report
Inappropriate shocks delivered by implantable cardiac
defibrillators during oversensing of activity of diaphagmatic muscle
Cardiology B
Department, Hospital Trousseau, 37044 Tours, France
Correspondence to: Dr Babuty.
Accepted for publication 24 June 1998
Two cases are reported (both men, one 72 and one 54 years
old) of inappropriate shocks delivered by an implantable cardiac defibrillator (ICD) device, which oversensed the myopotentials induced
by deep breathing and Valsalva manoeuvre. No damage to leads was
associated with the oversensing of myopotentials. The mechanism of the
inappropriate shocks was determined using real time electrograms.
Modification of the duration of ventricular detection and decrease in
sensitivity made it possible to avoid the oversensing of myopotentials
and to deliver ICD treatment.
© 1999 by Heart
This article has been cited by other articles:
-
Kowalski, M., Ellenbogen, K. A., Wood, M. A., Friedman, P. L.
(2008). Implantable cardiac defibrillator lead failure or myopotential oversensing? An approach to the diagnosis of noise on lead electrograms. Europace
10: 914-917
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.