Objectives: To evaluate optical coherence tomographic elastography as a method for assessing the elastic properties of atherosclerotic plaque and the parameters that influence interpretation.
Methods: Phantoms and aorta were examined in vitro to quantify speckle modulation and measure the displacement and strain maps. A correlation method was used as a speckle tracking technique for measuring axial and lateral displacement vectors and calculation of strain maps. The influence of correlation kernel size on accuracy of the method was evaluated.
Results: In terms of a percentage error between calculated and measured displacements, the best results for phantoms were obtained with a 41 x 41 kernel (1.88% error). For both phantom and aorta images, it was found that, with the increasing size of cross correlation kernel, the axial and lateral displacement maps are less noisy and the displacement vectors are more clearly defined. However, the large kernels tend to average out the differences in displacements of small particles in phantoms and decrease the ability of speckle tracking to make microstructural assessments. Therefore, it is important to select kernel size carefully, based on the image features.
Conclusions: Optical tomographic elastography can be used to assess the microstructural properties of atherosclerotic tissue at micrometre scale resolution, but preselected analysis criteria must be understood in a critical interpretation of the results.
Keywords: optical coherence tomography; elastography; atherosclerotic plaque; strain maps
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
-
Prati, F., Regar, E., Mintz, G. S., Arbustini, E., Di Mario, C., Jang, I.-K., Akasaka, T., Costa, M., Guagliumi, G., Grube, E., Ozaki, Y., Pinto, F., Serruys, P. W.J., for the Expert's OCT Review Document,
(2009). Expert review document on methodology, terminology, and clinical applications of optical coherence tomography: physical principles, methodology of image acquisition, and clinical application for assessment of coronary arteries and atherosclerosis. Eur Heart J
0: ehp433v1-ehp433
[Abstract] [Full Text] -
Raffel, O C, Akasaka, T, Jang, I-K
(2008). Cardiac optical coherence tomography. Heart
94: 1200-1210
[Full Text] -
Honda, Y., Fitzgerald, P. J.
(2008). Frontiers in Intravascular Imaging Technologies. Circulation
117: 2024-2037
[Full Text] -
Rogowska, J, Patel, N, Plummer, S, Brezinski, M E
(2006). Quantitative optical coherence tomographic elastography: method for assessing arterial mechanical properties. Br. J. Radiol.
79: 707-711
[Abstract] [Full Text] -
Stamper, D., Weissman, N. J., Brezinski, M.
(2006). Plaque Characterization With Optical Coherence Tomography.. J Am Coll Cardiol
47: C69-C79
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.