Register for email alerts and news feeds:
This journal | BMJ Group
rss
Published Online First: 4 May 2007. doi:10.1136/hrt.2006.107052
Heart 2007;93:1562-1566
Copyright © 2007 BMJ Publishing Group Ltd & British Cardiovascular Society

INTERVENTIONAL CARDIOLOGY

Stent expansion: a combination of delivery balloon underexpansion and acute stent recoil reduces predicted stent diameter irrespective of reference vessel size

Shahid Aziz1, John L Morris2, Raphael A Perry2, Rodney H Stables2

1 Department of Cardiology, Royal Blackburn Hospital, Blackburn, UK
2 Department of Cardiology, The Cardiothoracic Centre, Liverpool, UK

Dr Shahid Aziz, Royal Blackburn Hospital, Haslingden Road, Blackburn, Lancashire BB2 3HH, UK; saziz1{at}btinternet.com

ABSTRACT

Background: There is a strong inverse relationship between final vessel diameter and subsequent risk of treatment failure after coronary stent deployment. The aim of this study was to investigate the magnitude by which stent delivery balloon underexpansion and stent elastic recoil contributed to suboptimal final vessel geometry.

Methods: A prospective angiographic study recruiting 499 lesions (385 patients) undergoing coronary stent implantation was performed. Quantitative coronary angiography (QCA) was used to measure the minimal lumen diameters of the delivery balloon during stent deployment (MLD1) and of the stented segment following balloon deflation (MLD2). The expected balloon diameter for the deployment pressure was determined from the manufacturer’s reference chart. Delivery balloon deficit was measured by subtracting the MLD1 from the expected balloon size and stent recoil was calculated by subtracting MLD2 from MLD1. Delivery balloon deficit and stent recoil were examined as a function of reference vessel diameter (RVD) and balloon–vessel (BV) ratio.

Results: The final stent MLD was a mean 27.2% (SD = 7.2) less than the predicted diameter. The mean delivery balloon deficit was 0.65 mm (SD = 0.27) and the mean stent recoil was 0.28 mm (SD = 0.17). Percentage delivery balloon deficit and stent recoil were independent of RVD. Delivery balloon deficit increased with higher BV ratios. Stent recoil was independent of BV ratio and the use of predilatation.

Conclusion: Failure to achieve predicted final stent diameter is a real problem with contribution from delivery balloon underexpansion and stent recoil. On average the final stent MLD is only 73% of the expected diameter, irrespective of vessel size.

Abbreviations: BV, balloon–vessel ratio; CSA, cross-sectional area; MD, minimal diameter; MLD, minimal lumen diameter; QCA, quantitative coronary angiography; RVD, reference vessel diameter


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

Relevant Article

Coronary stenting: why size matters
Edoardo De Benedetti and Philip Urban
Heart 2007 93: 1500-1501. [Abstract] [Full Text] [PDF]

This article has been cited by other articles:

  • De Benedetti, E., Urban, P. (2007). Coronary stenting: why size matters. Heart 93: 1500-1501 [Abstract] [Full Text]  

This Article

Services
Citing Articles
Google Scholar
PubMed
Topic Collections
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.