In vitro validation of the luminal measurement of a novel catheter based moulding technique

Table 1

Comparison of the results using the catheter with published results for quantitative coronary angiography (QCA) systems

System Accuracy (mm) Precision (mm)
Moulding catheter
1.5–4.0 mm phantoms 0.19 0.34
2.5–4.0 mm phantoms 0.01 0.06
QCA system
Ozaki6 −0.11 to −0.1 0.08 to 0.18
Keane7 −0.08 to 0.07 0.40 to 0.09
−0.12 to −0.03 0.10 to 0.39
Reiber8 0.00 to 0.03 0.020 to 0.189
Keane9 1-150
 50% contrast −0.41 to 0.01 0.06 to 0.20
 100% contrast −0.23 to −0.01 0.07 to 0.28
Reiber10 −0.03 0.09
Uehata11 0.05 0.09
Haase12 1-151
 CMS 100% contrast 0.18 0.14
 CMS 50% contrast 0.19 0.15
 DCI 100% contrast 0.11 0.06
 DCI 50% contrast 0.24 0.13
  • 1-150 Results for 50% and 100% contrast in the same phantoms.

  • 1-151 Results comparing 50% and 100% contrast in cinefilm based Cardiovascular Measurement System (CMS) and Philips digital cardiac imaging system (DCI).

This Article

Latest from Education in Heart

Latest from Education in Heart

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of Heart.
View free sample issue >>

Free archive
The full back archive is now available for Heart. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.