rss
Heart 2001;85:451-453 doi:10.1136/heart.85.4.451
  • Interventional cardiology surgery

Assessment of the geometric profile of the Amplatzer and Cardioseal septal occluders by three dimensional echocardiography

  1. P Acar,
  2. Z Saliba,
  3. P Bonhoeffer,
  4. D Sidi,
  5. J Kachaner
  1. Services de Cardiologie Pédiatrique, Hôpital Necker/Enfants-malades, 149 rue de Sèvres, 75743 Paris cedex 15, France
  1. Dr Acarphilippearcar{at}yahoo.fr
  • Accepted 13 December 2000

Abstract

OBJECTIVE To apply three dimensional echocardiography to describe the geometric profile of the Amplatzer and Cardioseal occluders after deployment for closure of atrial septal defect.

METHODS 20 patients (mean (SD) age, 14 (5) years) were enrolled for transcatheter closure of a secundum atrial septal defect with the Amplatzer occluder (10) or with the Cardioseal occluder (10). The two populations were matched for the stretched diameter of the defect (mean 18 (6) mm). The profile of the two occluders was examined.

RESULTS Transoesophageal echocardiography did not show any residual shunts after Amplatzer occluder deployment, whereas three patients had a small residual leak after Cardioseal deployment. One patient had transient atrioventricular block with the Amplatzer device. The mean surface area of the Amplatzer occluder was 6.9 (2) cm2, and that of the Cardioseal device 5.4 (3) cm2 (p = 0.03). The mean volume of the Amplatzer occluder was 9.2 (1) cm3, while that of the Cardioseal occluder was 3.5 (1) cm3 (p < 0.0001). From the three dimensional views, the Cardioseal occluder looked like a flat square after deployment whereas the Amplatzer occluder took up a ball shape in the atrial cavity.

CONCLUSIONS Three dimensional views by multiplane transoesophageal echocardiography allow a realistic in vivo description of atrial septal occluders. The Amplatzer occluder, with its high geometric profile, allows complete closure of large atrial septal defects but with some risk of mechanical complications. Use of the Cardioseal device, with its small surface coverage and high residual shunt rate, should be limited to transcatheter closure of a patent foramen ovale or small atrial septal defects.

Footnotes

    Latest from Education in Heart

    Latest from Education in Heart

    Register for free content

    The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

    Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.