Percutaneous closure of patent foramen ovale: impact of device design on safety and efficacy

  1. M Schwerzmann1,
  2. S Windecker1,
  3. A Wahl1,
  4. H Mehta1,
  5. K Nedeltchev2,
  6. H Mattle2,
  7. C Seiler1,
  8. B Meier1
  1. 1Cardiology, Swiss Cardiovascular Centre Bern, University Hospital Inselspital, Bern, Switzerland
  2. 2Neurology, University Hospital Inselspital, Bern
  1. Correspondence to:
    Dr Bernhard Meier
    Swiss Cardiovascular Centre Bern, University Hospital Inselspital, 3010 Bern, Switzerland; bernhard.meierinsel.ch
  • Accepted 9 June 2003

Abstract

Objective: To compare the safety and efficacy of percutaneous closure of patent foramen ovale (PFO) with the Amplatzer PFO occluder (Amplatzer) or the PFO STAR device (STAR) in patients with presumed paradoxical embolism.

Methods: Implantation characteristics, procedural complications, residual shunt, and recurrence of thromboembolic events were recorded prospectively in 100 consecutive patients undergoing percutaneous PFO closure with the STAR (n  =  50) or Amplatzer (n  =  50) devices between 1998 and 2001. The study was not randomised. Device implantation was successful in all cases.

Results: There were more procedural complications in the STAR than in the Amplatzer group (8/50 v 1/50, p  =  0.01). More than one device placement attempt was an independent predictor of procedural complications (odds ratio (OR) 8.5, 95% confidence interval (CI) 1.3 to 55.8; p  =  0.03). A residual shunt six months after PFO closure, assessed by transoesophageal contrast echocardiography, occurred more often in the STAR than the Amplatzer group (17/50 v 3/50, p  =  0.004), and was predicted in the STAR group by the use of a device with a 5 mm as opposed to a 3 mm disc connector (OR 6.1, 95% CI 1.1 to 34.0; p  =  0.04). The actuarial risk of recurrent thromboembolic events after 3.5 years was 16.8% (95% CI 7.6% to 34.6%) in the STAR and 2.7% (95% CI 0.4% to 17.7%) in the Amplatzer group after three years (p  =  0.08).

Conclusions: Percutaneous PFO closure with the Amplatzer PFO occluder had fewer procedural complications and was more likely to be complete than with the STAR device. These findings underline the importance of device design for successful percutaneous PFO closure.

Footnotes

    Responses to this article

    Latest from Education in Heart

    Latest from Education in Heart

    Register for free content

    The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.