Ventricular arrhythmia, Cheyne-Stokes respiration, and death: observations from patients with defibrillators
  1. A D Staniforth1,
  2. S C Sporton2,
  3. M J Early3,
  4. J A Wedzicha4,
  5. A W Nathan3,
  6. R J Schilling3
  1. 1Department of Cardiovascular Medicine, Queens Medical Centre, Nottingham, UK
  2. 2The Heart Hospital, University College London Hospitals NHS Trust, London, UK
  3. 3Department of Cardiology, St Bartholomew’s Hospital, London, UK
  4. 4Department of Respiratory Medicine, St Bartholomew’s Hospital, London, UK
  1. Correspondence to:
    Dr Andrew D Staniforth
    Department of Cardiovascular Medicine, D Floor South Block, Queens Medical Centre, Nottingham NG7 2UH, UK; adstaniforthbeeb.net

Abstract

Objective: To determine whether ventricular arrhythmia related to nocturnal hypoxaemia during Cheyne-Stokes respiration (CSR) explains the observation that CSR is an independent marker of death in heart failure.

Design: Prospective, observational study.

Patients: 101 patients at high risk of clinical serious ventricular arrhythmia fitted with an implantable cardioverter-defibrillator (ICD).

Measurements: Patients were studied at baseline for CSR during sleep. Arrhythmia requiring device discharge was used as a surrogate marker for possible sudden cardiac death.

Results: 101 patients (42 with CSR) were followed up for a total of 620 months. Twenty six patients experienced 432 ICD discharge episodes. Twenty four (6%), 210 (49%), 125 (29%), and 73 (17%) episodes occurred across the time quartiles 0000–0559, 0600–1159, 1200–1759, and 1800–2359, respectively. Kaplan-Meier analysis showed a relative risk of 1 (95% confidence interval 0.5 to 2.2, p  =  1) for device discharge in the CSR group. The average (SED) numbers of nocturnal ICD discharges per patient per month of follow up were 0.01 (0.01) and 0.04 (0.02) for patients with and without CSR, respectively (p  =  0.6).

Conclusion: These findings refute the proposition that CSR is related to heart failure death through nocturnal serious ventricular arrhythmia.

  • CPAP, continuous positive airway pressure
  • CRT, cardiac resynchronisation therapy
  • CSR, Cheyne-Stokes respiration
  • ICD, implantable cardioverter-defibrillator
  • MADIT, multicentre automatic defibrillator implantation trial
  • NSVT, non-sustained ventricular tachycardia
  • Sao2, transcutaneous oxygen saturation
  • sleep apnoea
  • ventricular tachyarrhythmia
  • implantable cardioverter-defibrillator

Statistics from Altmetric.com

Cheyne-Stokes respiration (CSR) is an independent marker of poor prognosis in chronic stable heart failure.1–3 Although CSR may just be another marker of bad heart failure, there is evidence that treatments specifically directed against CSR improve prognosis.3 There are two plausible mechanisms by which CSR may exert an adverse effect. Firstly, the activation of the sympathetic nervous system during recurrent arousals and sleep fragmentation due to CSR may further exacerbate the maladaptive neurohormonal disturbances characteristic of the heart failure syndrome. Secondly, apnoea associated nocturnal oxygen desaturation may lead to cardiac hypoxaemia and serious ventricular arrhythmia.

Bradycardia rather than tachycardia was considered to be the hallmark of the sleep apnoea syndromes.4 The results of two recent Holter monitor studies have challenged this assumption regarding heart failure with CSR by showing an increase in the rate of ventricular extrasystoles and non-sustained ventricular tachycardia (NSVT). Further interpretation of these studies is limited by the lower left ventricular ejection fractions in patients with CSR than in patients without CSR,5 and by the observation that the excess of NSVT occurred during the day rather than during sleep.6 Furthermore, although ventricular extrasystoles and NSVT are associated with an increased risk of sudden cardiac death, these “soft” Holter measurements lack specificity and positive predictive value.

We report data from 101 patients with heart failure at high risk of ventricular arrhythmia fitted with an implantable cardioverter-defibrillator (ICD). We have prospectively investigated whether patients with nocturnal oxygen desaturation during periods of likely CSR were subsequently at an increased risk of clinical serious ventricular arrhythmia events necessitating ICD discharge. The high reliability and high quality electrogram storage facility of modern ICDs permitted a comparison of the diurnal variation of serious clinical ventricular arrhythmia between patients with and those without nocturnal oxygen desaturation.

METHODS

Patients

We recruited 101 patients undergoing ICD implantation. Patients of any age, sex, heart failure aetiology, and duration and severity of symptoms were enrolled. Patients with known significant pulmonary disorder, obstructive sleep apnoea, stroke, and morbid obesity (body mass index ⩾ 35 kg/m2) were excluded. All patients had stable cardiac failure with no change in medication for two weeks before study and no evidence of fluid retention at the time of study. All patients received standard heart failure treatment (defined as a diuretic, an angiotensin converting enzyme inhibitor or angiotensin receptor blocker, spironolactone, and a β blocker). Documented intolerance to one or more of these agents was not a reason for exclusion from study.

Heart failure was diagnosed on the basis of a suggestive medical history and examination findings together with impaired left ventricular function on two dimensional echocardiography. Left ventricular function was quantified by M mode echocardiography. Only patients with either a left ventricular diastolic diameter greater than 5.5 cm or an ejection fraction less than 35% were eligible for study. All patients underwent coronary angiography.

Assessment of nocturnal oxygen desaturation

All patients underwent overnight pulse oximetry with the Biox 3700 pulse oximeter (Ohmeda Biox Ltd) before ICD implantation. All studies took place in hospital on the night before ICD implantation. The apparatus records the lowest percentage transcutaneous oxygen saturation (Sao2) detected over successive 12 second time intervals up to a maximum of eight hours.7 The data were analysed to calculate the overnight minimum Sao2, mean Sao2, dip frequency (defined as a fall in Sao2 ⩾ =  4%), and the percentage of overnight recording time spent with Sao2 ⩽ 90%. Recordings with a sinusoidal pattern of nocturnal oxygen desaturation together with a 4% dip rate of > 15 dips/hour were defined as “abnormal and suggestive of CSR”.

External validation of oximetry screening

We have previously validated the use of overnight oximetry as a screening tool for CSR.8 Oximetry recordings were found to be reproducible with a mean difference over two successive nights of −0.2 dips/hour (within patient standard deviation of 1.3 dips/hour). With full polysomnography serving as the ideal, receiver operating characteristic analysis with a diagnostic cut off of 15 dips/hour yielded a sensitivity of 87% and a specificity of 81% for the identification of CSR. The area under the receiver operating characteristic curve was 94%.

Internal validation of oximetry screening: polysomnography substudy

Every fourth patient with an abnormal oximetry recording underwent technician supervised overnight polysomnography. All studies took place in hospital within four weeks of ICD implantation. This was to confirm that the overnight breathing abnormalities responsible for oxygen desaturation were identical to those observed in the validation study. Figure 1 shows an overnight oximetry trace from a patient with CSR.

Figure 1

 Sample overnight oxymetry recording from a patient with documented Cheyne-Stokes respiration (CSR). In this patient the 4% oxygen desaturation index was 29 dips/hour, minimum transcutaneous oxygen saturation (Sao2) was 88%, and mean Sao2 was 96%.

The Neuroscience Sleepmaster analysis system (Nellcor Ltd, Pleasanton, California, USA) was used to record the polysomnographs. Surface electrodes were used to record the encephalogram, electro-oculogram, and mental electromyogram. Respiratory movements were recorded with chest wall and abdominal respitrace bands. ECG was recorded with standard limb leads. Oxygen saturation was measured by pulse oximetry. Oral and nasal airflow were detected by thermistors attached to the upper lip.

The data were analysed off line after manual inspection of the raw polygraph recordings. Standard criteria were used to define sleep onset and end.9 The following definitions were used to classify abnormal breathing patterns: Apnoea, an absence of airflow for more than 10 seconds; central apnoea, an apnoea without respiratory movement; obstructive apnoea, an apnoea with respiratory effort; hypopnoea, a reduction in the amplitude of respiratory movement for more than 10 seconds to less than 50% of the maximum amplitude recorded during the preceding breathing cycle; CSR, a central apnoea index of ⩾ 10/hour in combination with the characteristic pattern of crescendo-decrescendo hyperpnoea-central apnoea periodic breathing; obstructive sleep apnoea, an obstructive apnoea index of ⩾ 10/hour; and periodic breathing time, percentage of the total sleep period spent in CSR.

ICD implantation, programming, and electrogram analysis

ICDs were implanted transvenously under moderate sedation with a standard surgical technique. Seventeen of the ICDs were cardiac resynchronisation devices (CRTs). Device programming was left to the discretion of implanting physicians according to their knowledge of the patients’ individual arrhythmia history. This ensured that ICD discharges would be delivered only for clinically relevant arrhythmia. Subsequent discharges were documented as appropriate or inappropriate by the interrogating cardiologist or cardiac technician after electrogram storage review. The occurrence and timing of appropriate discharges were recorded prospectively at follow up. Non-sustained episodes were specifically not documented.

Data analysis

Numerical data were analysed by Student’s t and Mann-Whitney U tests. Group comparisons of categorical data were evaluated by the χ2 test. Survival to the time of first appropriate discharge was analysed by the Kaplan-Meier method. Differences in survival between patients with and without CSR were investigated with the log rank test.

RESULTS

Patient characteristics and follow up

One hundred and one patients undergoing ICD implantation (89 new systems and 12 elective generator replacements) at our institution between April 2002 and May 2003 consented to participate in this study. During this period 153 devices were implanted (126 new systems and 27 generator replacements). There were 432 episodes of clinically significant serious ventricular arrhythmia necessitating ICD discharge during 620 months of follow up. No patient died. One patient transferred out of the area and was lost to follow up after one month. Forty two of 101 (41%) patients had an overnight oximetry recording suggestive of CSR.

Polysomnography substudy

Twelve patients with oximetry recordings suggestive of CSR underwent full overnight polysomnography to define the nature of their sleep related breathing disorder. The mean (SD) total sleep period was 342 (51) minutes. The mean apnoea-hypopnoea index was 26 (13)/hour with an apnoea index of 16 (15)/hour and a periodic breathing time of 46 (32)%. Among the events, 85 (23)% were central in origin; 11 patients fulfilled the predefined polysomnography definition of CSR. No patient had obstructive sleep apnoea.

Characteristics of patients with and without CSR

The majority of patients were men and had left ventricular dysfunction secondary to coronary artery disease (table 1). Patients with and without CSR were comparable with respect to age, severity of coronary artery disease, and antiarrhythmic use. As would be expected, patients with CSR had more severe left ventricular dysfunction; this was reflected by their greater NYHA classification and uptake of anti-failure medication. Twenty six per cent of ICDs were primary prevention devices (predominantly MADIT (multicentre automatic defibrillator implantation trial) and MADIT-II criteria). Patients without CSR were not more likely than those with CSR to receive an implant for primary prevention. By definition, those with CSR had worse indices of nocturnal oxygen saturation. Seventeen of the ICDs were CRT devices (apart from these, only one patient was bradycardia pacing dependent). Patients with CSR were not more likely than those without CSR to receive a CRT device (9 of 42 v 8 of 59, respectively, p  =  0.3).

Table 1

 Baseline characteristics of implantable cardioverter defibrillator (ICD) recipients with and without Cheyne-Stokes respiration (CSR)

ICD discharges in patients with and without CSR

Eleven of 42 patients (26%) with CSR and 15 of 59 patients (25%) without CSR experienced an appropriate ICD discharge. Appropriate discharge included a therapeutic defibrillation in 19 patients. Kaplan-Meier analysis (fig 2) showed no difference between patients with respect to ICD discharge-free survival. Cox proportional hazard analysis showed a relative risk of ICD discharge in patients with CSR of 1.0 (95% confidence interval 0.5 to 2.2, p  =  1). The mean (SED) number of discharges per patient per month of follow up was 0.5 (0.2) for patients with and 1.3 (0.7) for patients without CSR (p  =  0.9). For patients whose ICD discharged, the mean (SED) times in months to first discharge were 2.2 (0.5) and 2.4 (0.6) for patients with and without CSR (p  =  1). As would be expected, patients who received an appropriate discharge tended to be older (68.4 (1.6) v 66 (1.2) years, p  =  0.3) and have poorer ventricles (ejection fraction 28.5 (2.8)% v 34.5 (1.7)%, p  =  0.08) than those who did not. Patients with a CRT device were no less likely than those with a standard ICD to receive an appropriate discharge (3 of 17 v 23 of 84, p  =  0.55). Twenty patients received an inappropriate discharge; in 18 patients these involved defibrillation.

Figure 2

 Survival to first appropriate implantable cardioverter-defibrillator (ICD) discharge for serious ventricular arrhythmia in patients with and without CSR. Cox proportional hazard analysis showed that a patient with CSR had a relative risk for discharge of 1.0 (95% confidence interval 0.5 to 2.2, p  =  1).

Diurnal pattern of ICD discharges

It was possible that, although the overall arrhythmic burden between patients with and without CSR might have been identical, the groups might have differed with respect to their diurnal distribution of events. To investigate this, we compared the distribution of ICD discharges between the temporal quartiles of 0000 to 0559, 0600 to 1159, 1200 to 1759, and 1800 to 2359 (fig 3, table 2). There was no evidence that patients with CSR were more likely to experience ICD discharges during the night time (sleeping) hours when the degree of oxygen desaturation was greatest. The average (SED) numbers of nocturnal discharges (0000–0559) per patient per month of follow up were 0.01 (0.01) and 0.04 (0.02) for patients with and without CSR, respectively (p  =  0.6).

Table 2

 Patterns of ICD discharges in patients with and without CSR

Figure 3

 Diurnal variation of ICD discharges in patients with and without CSR.

DISCUSSION

CSR is an independent marker of poor prognosis in heart failure.1–3 It remains to be determined whether CSR is more than just a marker of a bad heart. Limited evidence suggests that CSR itself may be harmful. Continuous positive airway pressure (CPAP) ventilation has beneficial haemodynamic effects in heart failure10,11 and is an effective treatment for CSR.12 Sin et al3 reported that CPAP improved ejection fraction and survival in patients with heart failure with CSR; significantly, CPAP did not improve either ejection fraction or survival in a matched group without CSR. Evidence of improved survival after a treatment targeted against CSR strengthens the case for a cause and effect relation between CSR and mortality in patients with heart failure. One mechanism by which CSR may exert an adverse effect is serious ventricular arrhythmia caused by the repeated cycles of nocturnal apnoea related cardiac hypoxaemia.

In this prospective observational study we report on the incidence of serious ventricular arrhythmia in patients with and without patterns of nocturnal oxygen desaturation suggestive of CSR. All patients were fitted with an ICD programmed according to their individual clinical need. The advantage of this design was twofold. Firstly, by using ICDs with modern electrogram storage capacity we could accurately document all episodes of serious ventricular arrhythmia and their diurnal variation. Secondly, by leaving ICD discharge setup to the discretion of the implanting cardiologist according to individual patient clinical need, we could be confident that we were documenting clinically relevant serious ventricular arrhythmia (a more likely surrogate marker for aborted sudden cardiac death). This strategy was superior to counting ventricular extrasystoles or NSVT of dubious clinical relevance on serial 24 hour Holter monitors. It was interesting that no patients died during the follow up period. Although the total follow up was large (620 months), the median follow up was six months and patients with advanced near terminal pump failure would not have been considered for an ICD.

To our surprise, patients with CSR were no more likely than patients without CSR to receive ICD discharges for significant ventricular arrhythmia. The total number of appropriate ICD discharges observed over the follow up period was large (432 episodes in 26 patients). CSR occurs most often and is most severe during sleep, when breathing is purely under the control of a chemical negative feedback loop. We expected that if CSR exerted its lethal effects through hypoxaemia induced ventricular arrhythmia then CSR patients would have greater rates of ICD discharges during the sleeping hours 0000 to 0600. Although our patients with CSR had worse indices of nocturnal oxygen saturation, patients did not differ according to their diurnal variation of ICD discharges. ICD discharges were clustered within the 0600–1200 quartile for both patients with and without CSR. This was in keeping with the known morning increased incidence of ventricular tachycardia and sudden death in heart failure.13,14 We note that our patients with CSR had poorer ventricular function, which may ordinarily be expected to lead to more arrhythmia. This obviously did not apply in our study because, by definition, all patients were already at a high risk of arrhythmia.

The reason why nocturnal desaturation was not associated with serious ventricular arrhythmia is unclear. The degree of oxygen desaturation and hypoxia seen during central sleep apnoea is typically less severe than that observed in obstructive sleep apnoea. This may be of relevance when considering the individual roles of the various sleep apnoea syndromes as a substrate for arrhythmia (see below). Nocturnal oxygen desaturation in the setting of an unstable coronary syndrome is a proven stimulus for ischaemia and lethal ventricular arrhythmia.15 Our patients had chronic stable heart failure; none had an unstable coronary syndrome. A single study has reported that CRT reduces the frequency of ICD discharge.16 The use of CRT devices did not explain the lack of association between CSR and serious ventricular arrhythmia because patients with CSR were just as likely as non-CSR patients to receive a CRT device. Furthermore, patients with CRT devices did not experience less arrhythmia and excluding CRT patients from the study analysis did not change its findings.

Previous investigations of cardiac arrhythmia in sleep apnoea

A large observational study has reported on the frequency of nocturnal arrhythmia in 400 consecutive patients with sleep apnoea (all types and aetiology) referred to the Stanford sleep disorders clinic.4 The usual rhythm disturbance was bradycardia, not tachycardia. Ventricular extrasystoles were common but ventricular tachycardia occurred in only 3% of patients. No ventricular tachycardia was observed in the subgroup (10% of patients) with central sleep apnoea. Other studies examining the prevalence of sleep apnoea5,6 and nocturnal oxygen desaturation8,17,18 in patients with chronic stable heart failure have reported a higher occurrence of arrhythmia on 24 hour Holter monitoring. Reliance on single Holter recording significantly reduces the power of these observational studies to discern the true nature of the arrhythmia burden of this patient group. All of these studies were consequently limited to reporting on the frequency of “soft” end points such as ventricular extrasystoles and NSVT. In the recent study by Lanfranchi et al,6 CSR during sleep inexplicably increased daytime (but not night-time) NSVT. Patients with heart failure and CPAP responsive sleep apnoea (obstructive and central) have been shown to experience a reduction in ventricular extrasystoles and couplets.19 This treatment effect does not prove that hypoxaemia was responsible for ventricular arrhythmia in CSR. CPAP not only improves overnight oxygenation but also reduces sympathetic nervous system activation, reduces left ventricular afterload, and increases cardiac output.

Study limitations

Overnight oximetry (not polysomnography) was used to identify patients with likely CSR in the majority of patients. This does not limit our ability to comment on a link between nocturnal oxygen desaturation and arrhythmogenesis in chronic stable heart failure; indeed, we have previously validated the use of oximetry as a screening tool for CSR. We also performed a polysomnography substudy to confirm that the majority of apnoea events documented were indeed central rather than obstructive in origin. In the absence of oesophageal pressure balloon monitoring, some minor misclassification may have occurred. Oximetry was performed once only at baseline immediately before ICD implantation. The severity of nocturnal oxygen desaturation may have been reduced over the observation period, even though ICDs have no clinical effect on CSR, the use of CRT devices was equally split between the two groups, and medical treatment was not up-titrated as part of the study protocol. None of these detract from the major strength of this study: the deliberate use of high risk patients with ICDs implanted for arrhythmia to accurately document and define true clinical arrhythmia episodes and their temporal relation to likely nocturnal oxygen desaturation.

Conclusion

We have shown that nocturnal oxygen desaturation during CSR was not associated with an excess of serious ventricular arrhythmia. This implies that, if CSR exerts a lethal effect, it does so through a non-arrhythmic mechanism. These observations counter the findings and conclusions of two recent Holter monitor studies in CSR heart failure. An ICD alone will not therefore protect patients against the deleterious effects of CSR, for which further targeted treatments need to be considered.