Heart 2009;95:1091-1097 doi:10.1136/hrt.2008.153486
  • Original article
  • Heart failure and cardiomyopathy

Chronic monitoring of pulmonary artery pressure in patients with severe heart failure: multicentre experience of the monitoring Pulmonary Artery Pressure by Implantable device Responding to Ultrasonic Signal (PAPIRUS) II study

  1. U C Hoppe1,2,
  2. M Vanderheyden3,
  3. H Sievert4,
  4. M C Brandt1,
  5. R Tobar5,
  6. W Wijns3,
  7. Y Rozenman5
  1. 1
    Department of Internal Medicine III, University of Cologne, Cologne, Germany
  2. 2
    Center of Molecular Medicine Cologne (CMMC), Cologne, Germany
  3. 3
    Cardiovascular Center, Aalst, Belgium
  4. 4
    Cardiovascular Center Frankfurt, Frankfurt, Germany
  5. 5
    The Heart Institute, E. Wolfson Medical Center, Holon, The Tel-Aviv University, Sackler School of Medicine, Tel-Aviv, Israel
  1. Professor U C Hoppe, Department of Internal Medicine III, University of Cologne, Kerpener Str. 62, 50937 Cologne, Germany; uta.hoppe{at}uni-koeln.de
  • Accepted 7 October 2008
  • Published Online First 5 February 2009

Abstract

Objective: To evaluate the feasibility and safety of home monitoring of chronic heart failure (CHF) patients using acoustic wireless communication with an implant directly measuring pulmonary artery (PA) pressures.

Design: The PAPIRUS (Pulmonary Artery Pressure by Implantable device Responding to Ultrasonic Signal) II trial was a prospective, multicentre phase I study.

Patients: 31 patients with CHF in New York Heart Association class III-IV.

Interventions: Implantation of a miniature device in the right pulmonary artery (PA) responding to ultrasonic signal that enables wireless recording of a complete PA pressure curve.

Main outcome measures: The primary end points were rates of serious adverse device- or implantation-related events at 6 months. Secondary end points included accuracy of the measured PA pressure, functionality of the system and evaluation of pressure readings at different postures.

Results: The two safety end points were met with no serious adverse events related to the device or implantation. Pressure tracings at 6 months were almost identical to those obtained simultaneously by Millar catheter. Variations of PA diastolic pressure were observed in relation to posture (standing 6.4 (SD 3.4) mm Hg lower than supine, p<0.001). A total of 4627 home measurements were successfully performed by 23 patients using a simple-to-operate hand-held home-unit for daily measurements. The median compliance with daily monitoring was 86%.

Conclusions: Meeting the prespecified safety objective of this study warrants a randomised trial to fully evaluate the potential of home monitoring by this miniature PA implant in guiding long-term management in CHF.

Footnotes

  • Funding: Remon Medical Technologies, (Caesarea, Israel) now part of Boston Scientific (Natick, Massachusetts) funded the trial.

  • Competing interests: YR is a part-time employee of Remon Medical Technologies.

  • Ethics approval: Ethics approval was provided by University of Cologne, Cardiovascular Center Aust, Freiburger Ethik-Kommission International and Tel Aviv University.

  • Patient consent: Obtained.

Relevant Article

Latest from Education in Heart

Latest from Education in Heart

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of Heart.
View free sample issue >>

Free archive
The full back archive is now available for Heart. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.