Heart 2009;95:241-246 doi:10.1136/hrt.2007.139261
  • Original article

Negative vascular remodelling after implantation of bioabsorbable magnesium alloy stents in porcine coronary arteries: a randomised comparison with bare-metal and sirolimus-eluting stents

  1. M Maeng1,2,
  2. L O Jensen3,
  3. E Falk1,2,
  4. H R Andersen1,2,
  5. L Thuesen1,2
  1. 1
    Department of Cardiology, Aarhus University Hospital, Skejby, Denmark
  2. 2
    Institute of Clinical Medicine, Aarhus University Hospital, Denmark
  3. 3
    Department of Cardiology, Odense University Hospital, Denmark
  1. Dr Michael Maeng, Department of Cardiology B, Aarhus University Hospital, Skejby, Brendstrupgaardsvej, DK-8200 Aarhus N, Denmark; michael.maeng{at}ki.au.dk
  • Accepted 8 April 2008
  • Published Online First 22 August 2008

Abstract

Objective: We sought to assess vascular remodelling and neointima formation after implantation of bioabsorbable magnesium alloy stents (AMS).

Design: Randomised experimental study.

Interventions: AMS (n = 11), sirolimus-eluting stents (Cypher; n = 11) and bare-metal stents (BMS; n = 9) were randomly implanted in 31 porcine coronary arteries (n = 11 pigs).

Main outcome measures: Neointima formation was measured by histomorphometry at 90 days. Vascular remodelling defined as change in external elastic membrane area from index intervention to follow-up was assessed by serial intravascular ultrasound (IVUS).

Results: By histomorphometry, lumen (median (quartiles); AMS: 1.75 mm2 (1.07–3.26), Cypher 2.52 mm2 (2.22–5.01), BMS 4.55 mm2 (3.2–7.44); p = 0.013) and external elastic membrane area (AMS: 5.56 mm2 (4.09–6.95), Cypher 7.95 mm2 (6.45–10.92), BMS 9.08 mm2 (7.85–11.63); p = 0.014) were smallest after AMS implantation. By IVUS, external elastic membrane area at follow-up was smallest (AMS: 7.5 (2.8) mm2, Cypher 9.1 (2.7) mm2, BMS 9.9 (3.1) mm2; p = 0.33) and change in external elastic membrane area from index intervention to follow-up (remodelling; AMS: −1.0 (3.1) mm2, Cypher 1.0 (0.8) mm2, BMS 0.9 (1.2) mm2; p = 0.30) was greatest in the AMS group. In a dichotomised IVUS assessment of vascular remodelling, six AMS stents were remodelled (negative remodelling: n = 5; positive remodelling: n = 1) at 90-day follow-up (AMS versus Cypher + BMS: p = 0.001). Neointima formation was smallest in the AMS group (p<0.05 for both histomorphometry and IVUS).

Conclusion: Coronary implantation of absorbable magnesium stents, compared to two non-absorbable stents, was associated with the smallest lumen area at three-month follow-up because of negative vascular remodelling.

Footnotes

  • Competing interests: None.

This Article

  1. All Versions of this Article:
    1. hrt.2007.139261v1
    2. 95/3/241 most recent

Services

  1. Request permissions

Responses

  1. Submit a response
  2. No responses published

Social bookmarking

Latest from Education in Heart

Latest from Education in Heart

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of Heart.
View free sample issue >>

Free archive
The full back archive is now available for Heart. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.