Quantification of non-calcified coronary atherosclerotic plaques with dual-source computed tomography: comparison with intravascular ultrasound
  1. Tiziano Schepis,
  2. Mohamed Marwan,
  3. Tobias Pflederer,
  4. Martin Seltmann,
  5. Dieter Ropers,
  6. Werner G Daniel,
  7. Stephan Achenbach
  1. Department of Internal Medicine 2 (Cardiology), University of Erlangen, Erlangen, Germany
  1. Correspondence to Dr Tiziano Schepis, Department of Internal Medicine 2 (Cardiology), University of Erlangen, Ulmenweg 18, 91054 Erlangen, Germany; tiziano.schepis{at}gmx.ch

Abstract

Background The quantification of non-calcified coronary plaques using multidetector computed tomography has not been extensively investigated.

Objective To evaluate the ability of dual-source computed tomography (DSCT) to quantify non-calcified plaque volumes using intravascular ultrasound (IVUS) as the standard of reference.

Methods The datasets of 70 patients with suspected or known coronary artery disease who underwent DSCT (330 ms gantry rotation, 2×64×0.6 mm collimation, 60–90 ml contrast agent) were analysed before invasive coronary angiography, with IVUS performed as part of the diagnostic procedure. 100 individual non-calcified coronary atherosclerotic plaques (one to three plaques per patient) with suitable fiducial markers were matched and selected for plaque volume measurements using manual segmentation. Only DSCT datasets with good or excellent image quality were considered for analysis.

Results Intra and interobserver variability for plaque volume measurements by DSCT were 6±5% and 11±7%, respectively. Mean total plaque volume by DSCT was 89±66 mm3 (range 14–400 mm3). Mean total plaque volume by IVUS was 90±73 mm3 (range 16–409 mm3). The mean difference between DSCT and IVUS was 1±34 mm3 (range −131–85 mm3). Despite the good correlation for plaque volume measurements (r=0.89, p<0.001), agreement between the two methods was only modest (Bland–Altman limits of agreement −67 to +65 mm3).

Conclusions Non-calcified plaque volumes as determined by DSCT yielded good correlation but only modest agreement in comparison with IVUS.

  • Atherosclerosis
  • computed tomography scanning
  • dual-source computed tomography
  • intravascular ultrasound
  • non-calcified coronary plaque

Statistics from Altmetric.com

Footnotes

  • Linked articles 187815.

  • Funding This work was supported by a grant (01 EV 0708) from Bundesministerium für Bildung und Forschung, Berlin, Germany.

  • Competing interests None.

  • Ethics approval This study was conducted with the approval of the University of Erlangen institutional review board.

  • Patient consent Obtained.

  • Provenance and peer review Not commissioned; externally peer reviewed.

Request permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Linked Articles