Closure of a large ductus arteriosus in a preterm infant using the ADO II AS device
- Department of Paediatric Cardiology, O.I.R.M - S. Anna Hospital, Turin, Italy
- Correspondence to Dr Gabriella Agnoletti, Servizio di Cardiologia Pediatrica, piazza Polonia 90, 10124 Torino, Italy; gabriella.agnoletti{at}oirmsantanna.piemonte.it
- Received 16 March 2012
- Accepted 1 May 2012
- Published Online First 11 June 2012
- Interventional catheterisation
- paediatrics
- congenital heart disease
- Fallots tetralogy
- paediatric cardiology
- paediatric interventional cardiology
Current catheter technology is not designed for closure of large patent ductus arteriosus (PDA) in low-weight infants. Percutaneous coil occlusion of PDA was previously attempted in a sick premature baby with multiple malformations.1 The new miniaturised device, ADO II AS (Amplatzer Duct Occluder II Additional Size), might avoid the risk of surgical closure of PDA in preterm babies.2
A male child was born with a weight of 1050 g; he …