Heart 98:645-649 doi:10.1136/heartjnl-2011-301455
  • Heart rhythm disorders
  • Original article

Is 7-day event triggered ECG recording equivalent to 7-day Holter ECG recording for atrial fibrillation screening?

  1. Hildegard Tanner
  1. Department of Cardiology, Inselspital, Bern University Hospital and University of Bern, Switzerland
  1. Correspondence to Laurent Roten, Kardiologie, Inselspital, Freiburgstrasse, 3010 Bern, Switzerland; laurent.roten{at}insel.ch
  1. Contributors All authors were involved in designing the study and drafting and approving the final manuscript. Furthermore, Laurent Roten, Manuel Schilling, Andreas Häberlin and Hildegard Tanner were involved in data collection and data analysis.

  • Accepted 31 January 2012
  • Published Online First 7 March 2012

Abstract

Objective Prolonged ECG monitoring is standard for atrial fibrillation (AF) screening. This study investigated whether 7-day event triggered (tECG) ECG recording is equivalent to 7-day continuous Holter (cECG) ECG recording for AF screening.

Design Both a cECG (Lifecard CF) and a tECG (R.Test Evolution 3) were simultaneously worn for 7 days by patients with known or suspected paroxysmal AF.

Results In 100 corresponding recordings, median effective duration of monitoring was 165 h (range 10–170 h) for cECG and 137 h (0–169 h) for tECG (p<0.001). Median number and total duration of arrhythmias (AF, atrial flutter or atrial tachycardia) of ≥30 s duration recorded by cECG were 10 (1–428) and 1030 min (≤1–10 020), respectively. An arrhythmia was recorded in 42 cECGs (42%) versus 37 tECGs (37%, p=0.56). Triggered ECG failed to record an arrhythmia in cECG positive cases because of interrupted monitoring in four cases and because of recording failure in one case. Sensitivity, specificity, and positive and negative predictive values of tECG therefore were 88%, 100%, 100%, and 92%, respectively. Quantitative cECG analysis required a median of 20 min (3–205 min) and qualitative tECG analysis 4 min (1–20 min; p<0.001). Skin irritation was a frequent side effect (42%) resulting in premature removal of devices in 16% of patients.

Conclusion Sensitivity of tECG for AF screening as compared to cECG is lower, mainly because of shorter effective monitoring duration. Qualitative tECG analysis is less time consuming than quantitative cECG analysis. Skin irritation is a frequent side effect and reason for premature device removal.

Footnotes

  • Funding Dr Tanner was supported by a grant from the Swiss Foundation for Pacemaker and Electrophysiology.

  • Competing interests None.

  • Ethics approval Ethics approval was approved by Kantonale Ethikkommission Bern (http://www.kek-bern.ch).

  • Provenance and peer review Not commissioned; externally peer reviewed.

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of Heart.
View free sample issue >>

Free archive
The full back archive is now available for Heart. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.