Article Text

PDF

25 An MRI Coil Dual-Tuned by RF Shielding for In-Vivo Detection of Fluorine Labelled Stem-Cells in a Rodent Model of Acute Myocardial Infarction
  1. LE Cochlin1,
  2. RSM Gomes1,2,3,
  3. K Clarke1,
  4. CA Carr1
  1. 1Cardiac Metabolism Research Group, Department of Physiology, Anatomy and Genetics, University of Oxford, UK
  2. 2Biomaterials and Stem-Cells Based Therapeutics Lab, Biocant, Cantanhede, Portugal
  3. 3Centre of Neurosciences and Cell Biology, University of Coimbra, Portugal

Abstract

Stem-cell therapy holds great promise for treatment of disease. However, optimal stem-cell type, number, route, timing of administration have yet to be quantitatively determined. Non-invasive methods of monitoring cell retention are therefore in demand. Fluorine-MRI offers the facility for unambiguous localisation of fluorine-labelled cells as they migrate in-vivo. We achieve high fluorine sensitivity by simple modification to a standard MRI-coil. We monitored stem-cell retention in a rat model of myocardial infarction (MI), for 13 days after administration of fluorine-labelled stem-cells

Methods MRI-coil: 65mm diameter, low-pass birdcage fitted with two RF-shields; 85mm removable, 79mm fixed. Inner shield removal causes the resonant frequency window to shift from proton to fluorine. Labelled cell preparation and administration: cardiac-derived stem-cells were incubated with polylactic-co-glycolic acid (PLGA) nanoparticles containing perfluoro(crown)ether (PFCE) and fluoresceinamine. Five million labelled cells were administered immediately post MI, induced by occlusion of the left descending coronary artery. MRI at 1, 5, 7 and 13 days post MI, using a 7 Tesla magnet. Cardiac-gated gradient-echo cardiac long-axis and short-axis stacks: 1 proton average (Texp~25s), 15 fluorine averages, (Texp~380s) TE=2.9ms, FA=Ernst angle (TR=RRinterval~200ms, T1(PFCE)=1090±15ms in-vivo), FOV=50x50x1.5mm, matrix=128x128.

Results In-vivo, fluorine signal was easily located for setup, and high-resolution fluorine images showed signal co-registered with infarction and operation wound. At day 13, hearts were excised for histological verification of MRI findings.

Conclusion We demonstrate high proton/fluorine sensitivity using a standard MRI-coil, modified for inductance tuning to fluorine. Labelled cardiac stem-cells were imaged for 13 days post MI. Whole-heart proton and fluorine image stacks of matched resolution were acquired within one-hour scan-time.

Statistics from Altmetric.com

Request permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.