Biocompatibility of septal defect closure devices
  1. Matthias Sigler*,
  2. Christian Jux*
  1. Department of Paediatric Cardiology and Paediatric Intensive Care Medicine, Georg-August University, Goettingen, Germany
  1. Correspondence to:
    Dr M Sigler
    Department of Paediatric Cardiology and Paediatric Intensive Care Medicine, Georg-August University Goettingen, Robert Koch Strasse 40, Goettingen D 37099, Germany; msigler{at}gwdg.de

Abstract

Objective: Despite their clinical introduction 10 years ago, no human series on the healing response to Amplatzer and Starflex devices in humans have been reported yet. We sought to investigate the biocompatibility of Amplatzer and Cardioseal/Starflex septal occluder devices in humans and compare the findings to results in experimental animals.

Methods: The healing response of Amplatzer and Cardioseal/Starflex septal occluder devices in humans (n = 12, follow-up periods from 5 days to 4 years) and in experimental animals (n = 32, follow-up periods from 4 days to 1 year) was studied using a uniform work up protocol. Histological sections of paraffin-wax-embedded or methacrylate-embedded specimen and scanning electron microscopy were used for biocompatibility screening.

Results: Neoendothelialisation of all examined devices was complete after 3 months in vivo. Protruding metal frame parts, like screw threads and spring arms, were covered last. The initial deposition of fibrin and blood cells on the polyester fabric was subsequently organised by ingrown fibroblastic cells. Loosely arranged and poorly vascularised young granulation tissue was transformed time-dependently into quiescent fibre-rich connective repair tissue poor of cellular and capillary vessel components. Consistently, a mild chronic inflammatory response directed against textile fibres of both types of implants characterised by lymphocytic infiltration and multinucleated foreign body giant cells was observed equally in human and animal explants.

Conclusions: Systematic biocompatibility screening in a series of explanted human septal occluder devices showed results corresponding to findings in animal studies with regard to neoendothelialisation, cellular organisation of initial thrombus and persisting immune response.