Skip to main content
Log in

New criteria for estimating baroreflex sensitivity using the transfer function method

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

Computer simulations were carried out to appraise three new criteria for the estimation of baroreflex sensitivity (BRS) using the tranfer function method. The major goal was to identify a computation procedure able to overcome the intrinsic limitations of the classical coherence criterion. Four representative shapes of the gain function and three different average agains (2, 5 and 8 ms(mmHg)−1) in the lowfrequency (LF) band (0.04–0.15 Hz) were considered. The signal-to-noise ratio was made to vary so that the peak coherence in the LF band changed from 0.15 to 0.9 All simulation parameters were derived from previous observations in healthy subjects and heart disease patients. The error of the estimated gain function was obtained from its confidence interval. BRS was computed as average gain in the LF band: (a) including in the average only those points having error≤ threshold (criterion 1, C1); (b) calculating the mean error in the band and accepting BRS measurements only when this error was ≤ threshold (criterion 2, C2); (c) including in the average all points, regardless of the error (criterion 3, C3). The three criteria were compared in terms of measurability (percentage of measured BRS) and accuracy (bias and SD of BRS). Using C1 and C2, measureability dropped to 10% when the peak cohrence in the LF band decreased, respectively, to 0.18–0.41 and to 0.26–0.53, depending on the shape and strength of the gain. In this condition (lower bound of measureability), worst bias and SD (average gain: 8 ms(mmHg)−1) were, respectively, 0.8 ms(mmHg)−1 and 3.3 ms(mmHg)−1 (C1), and 0.1 ms(mmHg)−1 and 1.0 ms(mmHg)−1 (C2), C3, by definition, always ensured 100% measurability and showed bias and SD comparable with, or even lower than, C1 and C2, within the common range of measurable BRS. In the extreme condition of 0.15 coherence, bias and SD were, respectively, 1.7 ms(mmHg)−1 and 2.3 ms(mmHg)−1 (average gain: 8 ms(mmHg)−1). Hence, error checking (C1 and C2) dramatically reduced measurability and did not improve accuracy of BRS measurements compared with performing no error check (C3). In conditions of low signal-to-noise ratio and/or impaired baroreflex gain, leading to markedly reduced coherence, the simple average of the gain function in the LF band allows BRS to be estimated with accuracy adequate for clinical purposes.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

References

  • Basselli, G., Cerutti, S., Civardi, S., Malliani, A., andPagani, M. (1988): ‘Carchovascular variability signals: toward the identification of a closed-loop model of the neural control mechanisms’,IEEE Trans. Biomed. Eng.,35, pp. 1033–1046

    Google Scholar 

  • Bertram, D., Barrés, C., Cuisinaud, D., andJulien, C. (1998): The arterial baroreceptor reflex exhibits positive feedback properties at the frequency of Mayer waves’,J. Physiol.,513.1, pp. 251–261

    Google Scholar 

  • De Boer, R. W., Karemaker, J. M., andStrakee, J. (1987): ‘Hentodynamic fluctuations and baroreflex sensitivity in humans: a beat-to-beat mode’.Am. J. Physiol.,253, pp. 680–689

    Google Scholar 

  • La Rovere, M. T., Bigger, J. T., Marcus, F. I., Mortara, A., andSchwartz, P. J. for the ATRAMI Investigators (1998): ‘Barbreflex sensitivity and heart rate variability in prediction of total cardiac moratility after myocardial infarction’,Lancet,351, pp. 478–484

    Article  Google Scholar 

  • Mandel, J. (1991): Evaluations and control of measurments’ (Marcel Dekker, Inc., New York, 1991)

    Google Scholar 

  • Mortara, A., La Rovere, M. T., Pinna, G. D., Parziale, P., Maestri, R., Capomolia, S., Opasich, C., Cobelli, F., andTavazzi, L. (1997): ‘Depressed arterial baroreflex sensitivity and not reduced heart rate variability identifies patient with chronic heart failure and nonsustained ventricular tachycardia: the effect of high ventricular filliing pressure’,Am. Heart. J.,13, pp. 879–888

    Google Scholar 

  • Pinna, G. D., andMaestri, R. (2001): ‘Reliability of transfer function estimates in cardiovascular variability analysis’,Med. Biol. Eng. Comput.,39, pp. 338–347

    Article  Google Scholar 

  • Polosa, C. (1984): ‘Central nervous system origin of some type of Meyer waves' inMiyakawa, K., Koepchen, H. P., andPolosa, C. (Eds.): ‘Mechanisms of blood pressure waves’ (Springer-Verlag. Tokyo, 1984), pp. 277–292

    Google Scholar 

  • Priestley, M. B. (1981): ‘Spectral analysis and times series’ (Academic Press, London, 1981).

    Google Scholar 

  • Radaelli, A., Perlangeli, S., Cerutti, M. C., Mircoli, L., Mori, I., Boselli, L., Bonaita, M., Terzoli, L., Candotti, G., Signorini, G. andFerrari, A. U. (1999): ‘Altered blood pressure variability in patients with congestive heart failure’.J. Hypertens.,17, pp. 1905–1910.

    Article  Google Scholar 

  • Robbe, H. W. J., Mulder, L. J. M., Ruddel, H., Langewitz, W. A., Veldman, J. B. P., andMulder, G. (1987): ‘Assessment of baroreceptor reflex sensitivity by means of spectral analysis’,Hypertension,10, pp. 538–543.

    Google Scholar 

  • Sleight, P., La Rovere, M. T., Mortara, A., Pinna, G. D., Maestri, R., Leuzzi, S., Bianchini, B., Tavazzi, L., andBernardi, L. (1995): ‘Physiology and pathophysiology of heart rate and blood pressure variability and pathophysiology of heart analysis largely an index of baroreflex gain?’,Clin. Sci.,88, pp. 103–109.

    Google Scholar 

  • Taylor, J. A., andEckberg, D. L. (1995): ‘Fundamental relations between short-term RR interval and arterial pressure oscillations in humans’,Circulation,93 pp. 1527–1532.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to G. D. Pinna.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Pinna, G.D., Maestri, R. New criteria for estimating baroreflex sensitivity using the transfer function method. Med. Biol. Eng. Comput. 40, 79–84 (2002). https://doi.org/10.1007/BF02347699

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02347699

Keywords

Navigation