Table 1

Comparison of automated cardiac flow measurement (ACM) and catheter measurements in patients with atrial septal defects

ACM measurementsCatheter measurements
Patient numberAge (years)SexRVSV (ml)LVSV (ml)Qp/QsQp/Qs
F, female; LVSV, left ventricular stroke volume; M, male; RVSV, right ventricular stroke volume.
    123F70371.92.5
    268M167453.72.7
    325F118294.13.8
    471F102343.04.8
    540F100402.52.3
    660M148334.66.1
    753F157246.57.4
    820M55451.21.4
    957F72421.72.7
1027F85352.43.2
1137F84382.11.7
1247F121422.92.6
1355F87342.43.0
1440F116402.93.6
1552M88362.42.9
1637M128452.83.5
1761F218583.83.4
1840F186375.04.6
1923F114363.13.7
2034M172404.33.9
2126F99412.42.7
2263F113432.62.0