Table 3

Agreement in detection of patients with abnormalities between the SonoHeart and a standard echocardiographic device, showing the number of patients

SonoHeart
NormalAbnormal
There were 107 patients. Agreement = 96%, κ = 0.92, sensitivity = 96% (95% confidence interval (CI) 0.89 to 0.99), specificity = 96% (95% CI 0.89 to 0.99), positive predictive value = 96% (95% CI 0.89 to 0.99), negative predictive value = 96% (95% CI 0.89 to 0.99).
Standard echocardiographic device
Normal51 2
Abnormal 252