Table 1

Results of radiofrequency application during VT

Remote from diastolic pathway*On exitNear diastolic pathwayOn diastolic pathwayTotal
Success3  (9%)5  (21%)4  (80%)10  (77%)22
Partial0 (0%)7 (29%)0 (0%)2 (15%)9
Failure31 (91%)12 (50%)1 (20%)1 (8%)45
Total34 (100%)24 (100%)5 (100%)13 (100%)76
  • Remote, radiofrequency was delivered at a site remote from the diastolic pathway identified by the non-contact system.

  • On exit, radiofrequency was delivered on the VT exit identified by the non-contact system.

  • Near diastolic pathway, radiofrequency was delivered near an incompletely identified diastolic pathway.

  • On diastolic pathway, radiofrequency was delivered by a catheter positioned, using the locator signal, on the diastolic pathway identified by the non-contact system.

  • χ2 analysis for all data p < 0.0001.

  • With respect to the “on diastolic pathway” row: *p < 0.0001; p < 0.005.