Skip to main content
Log in

Re-occlusion of residual leaks after transcatheter occlusion of patent ductus arteriosus

  • Cardiology
  • Original Paper
  • Published:
European Journal of Pediatrics Aims and scope Submit manuscript

Abstract

Abstract

Between March 1990 and November 1993 175 patients underwent successful closure of patent ductus arteriosus using the Rashkind double umbrella device. Of those patients seen on follow up, 13 (9 female, 4 male) had a residual leak and were admitted for implantation of a second device alongside the first device. The interval between the procedures ranged from 6 to 22 months. The mean age at the first procedure was 73.9 months (range 24–204 months) and the mean weight was 18.9 kg (range 8.4–64). The mean age at the second procedure was 86.9 months (30–213) and the mean weight was 21.4 kg (8.6–64). The first device was 17 mm in 11 patients and 12 mm in two. The second device was 17 mm in four patients and 12 mm in nine. Four patients required two 17 mm devices in total. There was no difference in the two procedures regarding the fluoroscopy time, procedure time, complications and length of hospital stay. The second procedure was uneventful; however, in one patient the residual ductus had to be dilated before successfully deploying a 12 mm device. Ten patients had immediate total occlusion and three had a trivial leak on echocardiography 24 h later. One of the three patients was found to have total occlusion on Doppler echocardiography 1 year later. The other patients are yet to be seen for follow up. Hence a total occlusion rate was in 11/13 patients (85%). In nine patients pulmonary artery angiography was performed immediately after the implantation of the second device to rule out left pulmonary artery stenosis; three of those developed mild indentation at the origin of the left pulmonary artery but there was no evidence of stenosis on withdrawal pressure recordings Doppler echocardiography in any of the 13 patients.

Conclusion

The re-occlusion of a patent ductus arteriosus using a second Rashkind umbrella device is feasible and has a low risk.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

Abbreviations

PDA :

patent ductus arteriosus

References

  1. Al Yousef S, Khan A, Mullins CE, Sawyer W (1992) Use in children of an additional umbrella for transcatheter occlusion of residual patency of the arterial duct following initial insertion of an umbrella device. Cardiol Young 2: 353–356

    Google Scholar 

  2. Balzer DT, Spray TL, McMullin D, Cottingham W, Canter CE (1993) Endarteritis associated with a clinically silent patent ductus arteriosus. Am Heart J 125: 1192–1193

    PubMed  Google Scholar 

  3. Dyck JD, Benson LN, Smallhorn JF, McLaughlin PR, Freedom RM, Rowe RD (1988) Catheter occlusion of the persistently patent ductus arteriosus Am J Cardiol 62: 1089–1092

    PubMed  Google Scholar 

  4. Galal O, Wilson N, Fadley F, Duran CMG (1992) Novice experience with transcatheter closure of the arterial duct in children, adolescents and adults. Cardiol Young 2: 285–290

    Google Scholar 

  5. Galal O, Abbag F, Fadley F, Redington A, Oufi S (1995) Transcatheter occlusion of patent ductus arteriosus as a day case procedure. Cardiol Young 5: 48–50

    Google Scholar 

  6. Ghani SA, Hashim R (1989) Surgical management of patent ductus arteriosus. A review of 413 cases. J R Coll Surg Edinb 34: 33–36

    PubMed  Google Scholar 

  7. Hosking MCK, Benson LN, Musewe N, Dyck JD, Freedom RM (1991) Transcatheter occlusion of the persistently patent ductus arteriosus: forty month follow-up and prevalence of residual shunting. Circulation 84: 2313–2317

    PubMed  Google Scholar 

  8. Huggon IC, Tabatabaei AH, Qureshi SA, Baker EJ, Tynan M (1993) Use of a second transcatheter Rashkind arterial duct occluder for persistent flow after implantation of the first device: indications and results. Br Heart J 69: 544–550

    PubMed  Google Scholar 

  9. Khan MA, Al-Yousef S, Mullins CE, Sawyer W (1992) Experience with 205 procedures of transcatheter closure of ductus arteriosus in 182 patients, with special reference to residual shunts and long-term follow-up. J Thorac Cardiovasc Surg 104: 1721–1727

    PubMed  Google Scholar 

  10. Musewe NN, Benson LN, Smallhorn JF, Freedom RM (1989) Two-dimensional echocardiographic and color flow Doppler evaluation of ductal occlusion with the Rashkind prosthesis. Circulation 80: 1706–1710

    PubMed  Google Scholar 

  11. Ottenkamp J, Hess J, Talsma MD, Buis-Liem TN (1992) Protrusion of the device: a complication of catheter closure of patent ductus arteriosus. Br Heart J 68: 301–303

    PubMed  Google Scholar 

  12. Sakurai T, Hoshino H, Suzuki Y, Yokoi H, Sakanaka K, Enomoto K, Okada N (1980) Balloon catheter test in patients with atrial septal defect and patent ductus arteriosus. Jpn Heart J 21: 779–792

    PubMed  Google Scholar 

  13. Sørensen KE, Kristensen BO, Hansen OK (1991) Frequency of occurrence of residual ductal flow after surgical ligation by color flow mapping. Am J Cardiol 67: 653–654

    PubMed  Google Scholar 

  14. Tynan M, Huggon I, Anjos R, et al (1992) Transcatheter occlusion of the persistent arterial duct: Report of the European registry. Lancet 340: 1062–1066

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Abbag, F., Galal, O., Fadley, F. et al. Re-occlusion of residual leaks after transcatheter occlusion of patent ductus arteriosus. Eur J Pediatr 154, 518–521 (1995). https://doi.org/10.1007/BF02074825

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02074825

Key words