Cardiac output measured by thermodilution in infants and children

https://doi.org/10.1016/S0022-3476(78)80067-5Get rights and content

To determine the accuracy and reproducibility of cardiac output determinations by thermodilution (COT) in children, simultaneous outputs were measured by the Fick technique (COF) (using measured oxygen consumption) and thermodilution in 26 children, ranging in age from 8 to 86 months, who were undergoing cardiac catheterization. There was excellent correlation between mean output by thermodilution and by the Fick technique: COT=1.10 COF−0.2 l/minute, r=0.91. In three-quarters of the patients the COT differed by 15% or less and in none differed by more than 25%. Serial values of thermodilution outputs were reproducible in each patient with a SD of 5.5%. Our observations indicate that COT is accurate, reproducible, and valuable in the care of critically ill infants and children.