Original contribution
Influences of ultrasonic machine settings, transducer frequency and placement of region of interest on the measurement of integrated backscatter and cyclic variation

https://doi.org/10.1016/S0301-5629(97)00117-8Get rights and content

Abstract

Integrated backscatter and its cyclic variation are potentially important parameters to discriminate normal from diseased myocardium. Cyclic variation of integrated backscatter is expected to be independent of machine settings. Backscatter images of swine hearts were taken using a two-dimensional backscatter system while acoustic power was varied at different time gain control (TGC) settings. Cyclic variation was measured in vivo with various acoustic power and TGC settings using different transducer frequencies. Three different regions were analyzed. For any given TGC setting, the relationship between acoustic power and integrated backscatter in vitro was linear only over a narrow range. In vivo, cyclic variation was present at all regions studied in both long- and short-axis views. However, lower acoustic power (<15 dB) and TGC (<20 dB), or excessive settings of acoustic power (>35 dB) and TGC (>50 dB), produced minimal cyclic variation. Appropriate acoustic power (20–35 dB) and TGC (30–50 dB) produced larger and more consistent cyclic variation at the posterior region of the left ventricle. These data indicate that each region has specific, appropriate machine settings to maximize the magnitude of cyclic variation.

References (43)

  • T Ota et al.

    Echocardiographic cardiac rejection surveillance using acoustic densitometry of two dimensional ultrasound backscatter imaging: Multiple views and regions of interest (abstract)

    J Am Coll Cardiol

    (1995)
  • D Recchia et al.

    Quantification of ultrasonic anisotropy in normal myocardium with lateral gain compensation of two-dimensional integrated backscatter images

    Ultrasound Med Biol

    (1993)
  • H Rijsterborgh et al.

    Ultrasound myocardial integrated backscatter signal processing: Frequency domain versus time domain

    Ultrasound Med Biol

    (1993)
  • DJ Skorton et al.

    Clinical potential of ultrasound tissue characterization in cardiomyopathies

    J Am Soc Echocardiogr

    (1988)
  • JE Stuhlmuller et al.

    Reproducibility of quantitative backscatter echocardiographic imaging in normal subjects

    Am J Cardiol

    (1992)
  • AFW van der Steen et al.

    Influence of attenuation on measurements of ultrasonic myocardial integrated backscatter during cardiac cycle (an in vitro study)

    Ultrasound Med Biol

    (1991)
  • BF Vandenberg et al.

    Cyclic variation of ultrasound backscatter in normal myocardium is view dependent: Clinical studies with a real-time backscatter imaging system

    J Am Soc Echocardiogr

    (1989)
  • Z Vered et al.

    Ultrasound integrated backscatter tissue characterization of remote myocardial infarction in human subjects

    J Am Coll Cardiol

    (1989)
  • B Barzilai et al.

    Effects of myocardial contraction on ultrasonic backscatter before and after ischemia

    Am J Physiol

    (1984)
  • B Barzilai et al.

    Detection of remote myocardial infarction with quantitative real-time ultrasonic characterization

    J Am Soc Echocardiogr

    (1988)
  • PS Douglas et al.

    Ultrasonic integrated backscatter: A new method for imaging and analysis (abstract)

    Circulation

    (1994)
  • Cited by (14)

    • Physical Properties of Tissues Relevant to Arterial Ultrasound Imaging and Blood Velocity Measurement

      2007, Ultrasound in Medicine and Biology
      Citation Excerpt :

      This parameter originated as a practical index of backscatter strength used with clinical imaging systems (O’Donnell et al. 1979, 1981a, 1981b) and it has mainly been used in in-vivo cardiac studies (Mimbs et al. 1981; Wickline et al. 1986; Milunski et al. 1989; Takiuchi et al. 1998). It has been recognized that IBS is dependent on machine factors, making it difficult to compare IBS measured using different machines (Ota et al. 1997), and that assumed values for attenuation of overlying tissues will affect estimated IBS (Bridal et al. 1998). Measurements in excised arteries have demonstrated higher backscatter in calcified and fibrous plaque compared with normal arteries or fatty plaque (Wickline 1993; Barzali et al. 1987; Bridal et al. 1998; Takiuchi et al. 2000; Waters et al. 2003) and anisotropic behavior with higher IBS in the radial orientation (Nguyen et al. 2002; de Kroon et al. 1991a).

    • Depth variation bias and interaction with gain setting in ultrasonic tissue characterization by integrated backscatter analysis

      2003, Journal of the American Society of Echocardiography
      Citation Excerpt :

      Our study also shows that even tough differences in IBS as a result of individual chest attenuation can be partially offset by subtraction of an internal reference (such as pericardium), in this case a significant effect of gain setting occurs. As previously described,12 the effect of gain setting is also a major source of variability in IBS analysis. However, in our study, the effect of gain variation could be almost fully predicted by the equation obtained in the plastic phantom, taking into account the nonlinear behavior of gain as to depth of a given structure.

    • Cyclic variation of integrated ultrasound backscatter in the left ventricle during the early neonatal period

      2000, American Heart Journal
      Citation Excerpt :

      The intensity of IBS at any one point is dependent on transmit power, transducer frequency, sample depth, and attenuation by the chest wall. By contrast, the CV of IBS is theoretically independent of these factors.8,12 CV was expressed as the difference between end-diastolic (peak) and end-systolic (nadir) IBS values.

    View all citing articles on Scopus
    View full text