Clinical Studies
Follow-up results of transvenous occlusion of patent ductus arteriosus with the buttoned device1

https://doi.org/10.1016/S0735-1097(98)00610-XGet rights and content
Under an Elsevier user license
open archive

Abstract

OBJECTIVES

The purpose of this presentation is to document results of buttoned device (BD) occlusion of patent ductus arteriosus (PDA) in a large number of patients with particular emphasis on long-term follow-up in an attempt to provide evidence for feasibility, safety and effectiveness of this method of PDA closure.

BACKGROUND

Immediate and short-term results of BD occlusion of PDA have been documented in a limited number of children.

METHODS

During a six-year period ending August 1996, transcatheter BD closure of PDA was attempted in 284 patients, ages 0.3 to 92 years (median 7) under a protocol approved by the local institutional review boards and FDA with an investigational device exemption in U.S. cases.

RESULTS

The PDAs measured 1 to 15 mm (median 4) at the narrowest diameter; 20 were larger than 8 mm and 10 larger than 10 mm. They were occluded with devices measuring from 15 to 35 mm delivered via 7F (N = 140) or 8F (N = 144) sheaths. Successful implantation of the device was accomplished in 278 (98%) of 284 patients. The Qp:Qs decreased from 1.8 ± 0.6 (mean ± SD) to 1.09 ± 0.19 (p < 0.001). Effective occlusion defined as no (N = 167 [60%]) or trivial (N = 79 [28%]) residual shunt was achieved in 246 (88%) patients. All types of PDAs, irrespective of the shape (conical, tubular or short), size (small or large) or length (short or long) of the PDA and previously implanted Rashkind devices, could be occluded. Follow-up data, 1 to 60 months (median 24) after device implantation, were available in 234 (84%) patients. Seven (3%) patients required reintervention to treat residual shunt with (N = 2) or without (N = 5) hemolysis. Actuarial reintervention-free rates were 95% at 1 and 5 years. There was gradual reduction of actuarial residual shunts and were 40%, 28%, 21%, 14%, 11%, 10%, 6% and 0% respectively at 1 day, 1, 6, 12, 24, 36, 48 and 60 months after device implantation. Incorporation of folding plug over the button loop in 10 additional patients produced immediate and complete occlusion of PDA.

CONCLUSIONS

This large multiinstitutional experience confirms the feasibility, safety and effectiveness of buttoned device closure of PDAs. All types of PDAs irrespective of the shape, length and diameter can be effectively occluded. Incorporation of folding plug over the button loop produces complete PDA occlusion at the time of device implantation.

Abbreviations

BD
buttoned device
PDA
patent ductus arteriosus

Cited by (0)

1

One of the authors of this article (EBS) is the inventor of the device used in this study.