Feasibility and accuracy of point-of-care pocket-size ultrasonography performed by medical students

Abstract

Background

Point-of-care ultrasound performed by clinicians is a useful supplement in the treatment and assessment of patients. We aimed to investigate whether medical students with minimal training were able to successfully acquire and interpret ultrasound images using a pocket-size imaging device (PSID) as a supplement to their clinical practice.

Methods

Thirty 5th year (of six) medical students volunteered to participate. They were each given a personal PSID device to use as a supplement to their physical examination during their allocated hospital terms. Prior to clinical placement the students were given three evenings of hands-on training with PSID by a board certified radiologist/cardiologist, including three short lectures (<20 min). The students were shown basic ultrasound techniques and taught to assess for basic, clinically relevant pathology. They were specifically instructed to assess for the presence or absence of reduced left ventricular function (assessed as mitral annular excursion < 10 mm), pericardial effusion, pleural effusion, lung comets, hydronephrosis, bladder distension, gallstones, abdominal free-fluid, cholecystitis, and estimate the diameter of abdominal aorta and inferior vena cava.

Results

A total of 211 patients were examined creating 1151 ultrasound recordings. Acceptable organ presentation was 73.8% (95% CI 63.1-82.6) for cardiovascular and 88.4% (95% CI: 80.6-93.6) for radiological structures. Diagnostic accuracy was 93.5% (95% CI: 89.0-96.2) and 93.2% (95% CI: 87.4-96.5) respectively.

Conclusion

Medical students with minimal training were able to use PSID as a supplement to standard physical examination and successfully acquire acceptable relevant organ recordings for presentation and correctly interpret these with great accuracy.

Peer Review reports

Background

We are increasingly reliant upon expensive and time-consuming biochemical and radiologic diagnostics to aid us in our evaluation of patients. Unfortunately this still results in major diagnostic errors in up to 30% of patients at autopsy [13]. Furthermore the increasing age and chronicity of the western population highlights the need for improved out of hospital diagnosis and treatment.

Point-of-care ultrasonography allows for the near instantaneous acquisition of real-time dynamic images, which can be correlated directly to the patient’s signs and symptoms [4, 5]. It has been shown to increase diagnostic accuracy, rapidly and cost effectively in the hands of experts and non-experts [613]. Furthermore, portable ultrasonography is a valuable teaching tool in medical anatomy and physiology as well as physical examination [1418]. Despite this most medical students are not routinely educated in the clinical use of point-of-care ultrasonography, as they are in more widely accepted and traditional techniques, such as the stethoscope. This may in part be due to the lack of evidence regarding the bedside use of pocket-size ultrasound by medical students.

We aimed to investigate whether medical students with minimal training were able to successfully acquire and interpret ultrasound images using a pocket-size imaging device (PSID) as a supplement to their clinical practice.

Methods

Medical students

The fifth year (of six) medical students eligible to participate in the study based on planned hospital rotations received verbal and written information from the authors regarding the study. Participation in the study was not part of the students’ curriculum and all participating students were volunteers. The first 30 students whom volunteered were included in the study. There were no further inclusion or exclusion criteria. The number of participating students was limited to the number of available PSID. The medical students had similar limited experience in ultrasound.

Study population

All patients over 18 years of age, encountered in-hospital and at outpatient clinics during the students’ clinical placement periods were eligible for inclusion. The patients were included from a total of seven regional hospitals between January-May 2012. There were no exclusion criteria, and all participating patients provided informed consent.

Training and education of medical students

The medical students received three evenings (nine hours) of combined theoretical and practical training in the use and interpretation of ultrasound images. The theoretical training was given as short didactical lectures by relevant specialists (cardiologists and radiologists) and focused on basic ultrasound physiology, anatomy and examples of normal and pathological ultrasound images. Students were specifically trained to assess for pathology relevant in the immediate emergency care of patients. They were instructed to assess for reduced left ventricular (LV) function defined as mitral annular excursion (MAE) < 10 mm [1921], pericardial effusion, pleural effusion, lung comets, inferior vena cava (IVC) diameter and variation, hydronephrosis, bladder distension, gallstones, signs of cholecystitis, diameter of abdominal aorta (AA) and abdominal free-fluid. Practical hands-on training was given by relevant specialists and senior registrars, with students using their personal PSID. Students were encouraged to perform at least 75 examinations prior to placement.

Written informed consent was obtained from all patients. The Regional Committee for Medical and Health Research Ethics had no objections to the study’s conduction, which was conducted according to the Declaration of Helsinki.

Pocket-size ultrasound examination

The ultrasound examination was performed bedside with a PSID, Vscan (GE Vingmed Ultrasound, Horten, Norway). The device measures 135 × 73 × 28 mm and weighs 390 g, including the phased-arrayed probe. Two-dimensional grey scale and live colour Doppler imaging are offered. The image sector for echocardiographic imaging is 75°. The bandwidth ranges from 1.7 to 3.8 MHz and is automatically adjusted. Storage and looping of a cardiac cycle are possible without ECG signal and looping of other structures is predefined and limited to 2 seconds. The device has separate modes optimized for cardiac and abdominal examinations. All images and recordings were saved on the device’s micro-SD card and later transferred to a computer by commercial software (Gateway; GE Vingmed Ultrasound).

The bedside (point-of-care) cardiovascular ultrasound examination was performed with patients in the left-lateral decubitus and/or supine position. Assessment of LV global function was done from the apical four-chamber view using MAE, where MAE < 10 mm was classified as decreased LV function. Pericardial effusions were classified as present or not. The AA and IVC were assessed from the subcostal position. The AA diameter was assessed proximally to the bifurcation and if exceeding 35 mm classified as an abdominal aortic aneurysm (AAA). IVC diameter was measured end-expiratory within two cm from the right atrium orifice. All measurements of dimensions were done on the PSID. With patients in a supine or upright position, the pleura was assessed from left and right thoracic dorsolateral views, and assessed for the presence of pleural effusions and comet tails.

Other abdominal structures and spaces were assessed from a supine position looking specifically for hydronephrosis, bladder distension, gallstones, and signs of cholecystitis, and abdominal free-fluid.

Accuracy

The students were required to hand-in a log of selected examinations including their own set diagnosis based upon PSID examination. The specialists, one board certified radiologist and 2 board certified cardiologists with special interest in ultrasonography and echocardiography, were asked to categorize the image acquisition of relevant organ as acceptable or unacceptable for clinical interpretation and then determine whether the set diagnosis of the acceptable images were correct or incorrect. The specialists were not blinded to the set diagnosis.

Statistics

Data not following a normal distribution were presented as median and (interquartile) range. For sufficiently large samples logistic mixed model with random intercepts for students was used to examine estimate proportions. Clopper-Pearson estimates were used for small sample analyses. Sensitivity and specificity, negative and positive predictive value calculations were performed using relevant specialists as “gold standard”.

All the statistical analyses were performed using SPSS for Windows/Mac (version 20.0, SPSS, Inc.) or R version 2.13.1.

Results

Thirty 5th year (of six) medical students volunteered to participate in the study. At the end of the study period and their clinical placement, 21 (70%) medical students had performed exams using PSID and recorded their results. A total of 211 patients were examined (43% male, 38% female and 18% unrecorded sex), creating 1151 ultrasound recordings. Each student examined a median of 9 (±8, range 1–27) patients, producing a median of 49 (±49, range 5–169) ultrasound recordings. Acceptable organ presentation (Figure 1) was estimated to 73.8% (95% CI 63.1-82.6) for cardiovascular (heart, lungs and IVC) and 88.4% (95% CI: 80.6-93.6) for radiological (AA, renal system, gallbladder and abdominal free fluid) structures. Specifically, students performed best when acquiring images of the lungs and renal system (>93% (95% CI: 84.3-98.2) and found it most difficult to acquire acceptable images of the heart (71.2% (95% CI: 58.7-81.5)) and free fluid (73.2% (95% CI: 41.4-92.7)). The other categories (AA, IVC and gallbladder) had acceptable presentation in >80% (95% CI: 65.2-92.9) of cases. Diagnostic accuracy (Figure 2) was estimated at 93.5% (95% CI: 89.0-96.2) for cardiovascular structures and 93.2% (95% CI: 87.4-96.5) for radiological structures. The specific diagnostic accuracy was on a whole excellent. Diagnostic accuracy was close to 100% for AA (98.6% (95% CI: 92.7-100)) and free abdominal fluid (100% (95% CI: 76.8-100)) and lowest for gallbladder at 87.6% (95% CI: 73.7-95.1). The remaining categories showed diagnostic accuracy > 93% (95% CI: 83.3-99).

Figure 1
figure 1

Acceptable organ presentation. Cardiovascular all; heart, IVC and Lungs, IVC; Inferior vena cava, Radiological all; includes AA, Renal system, Gallbladder and Abdominal free fluid. AA; Abdominal aorta.

Figure 2
figure 2

Correct diagnosis. Cardiovascular all; heart, IVC and Lungs, IVC; Inferior vena cava, Radiological all; includes AA, Renal system, Gallbladder and Abdominal free fluid. AA; Abdominal ao.

The estimated values for sensitivity, specificity, negative and positive predictive values of PSID are presented in Table 1.

Table 1 Sensitivity, specificity, positive and negative predictive values

Discussion

Medical students, with a limited amount of training, successfully incorporated the use of point-of-care ultrasonography in their clinical placements. They were able to correctly acquire bedside ultrasound images of cardiovascular and radiological structures in 74 and 88% of their patients and correctly interpret these images in 93% of cases.

An attempt to simulate real life scenarios was done when determining the feasibility and accuracy. In our experience, when non-experts use pocket-size ultrasound at the patients point-of-care they may have the need to clarify or present their ultrasound findings to a specialist for review or guidance. The specialists were in this setting used as the gold standard with regards to statistical analysis and were not blinded to the set diagnosis. Optimally this would have been done by higher order, formal imaging, but that was beyond the scope of this study in terms of logistics and economy.

Other studies have shown that medical students are able to quickly acquire ultrasound recordings of good quality on normal test subjects, in optimal conditions with a standard ultrasound machine and PSID after a brief period of training [22, 23]. For the assessment of diagnostic accuracy in our study, only acceptable organ images were used. This may have diluted the true diagnostic accuracy to some extent. However the lack of a formal gold standard/reference made the basis for this, as assessing accuracy in non-acceptable images is useless when no reference is available.

A recent, though smaller study with five final year medical students has shown encouraging results using pocket-size cardiac ultrasonography as an adjunct to standard physical examination in cardiology patients [9]. We have broadened the field, looking at several different organ systems and included diverse groups of hospital and emergency room patients.

The European Association of Echocardiography published a position statement in 2010 regarding with the use of PSID [24]. It supports the use of PSID as a teaching tool in medical schools, as a tool for a fast initial screening in the emergency setting, and as a complement to the standard physical examination.

Previous studies have shown increased accuracy, efficacy and diagnostic impact of pocket-size point-of-care ultrasonography in the hands of experts versus non-experts [68, 11, 13, 25]. Thus the benefits of bedside PSID exams increase with increasing proficiency in its use and proficiency has been shown to increase with increasing use [23]. Additionally, ultrasonography has been shown to increase the skills of medical students in core subjects such as anatomy, physiology, and physical examination [9, 1418]. Therefore standardized training with an appropriate education program in the routine use PSID as an adjunct to standard physical examination should start as early as possible in a physician’s career.

Limitations

The main limitation of this study is the inability to exclude for selection bias. With the use of their logbooks, students were able to select which ultrasound loops were eligible for review. This selection and spectrum bias may have lead to some overestimation of the results for feasibility and accuracy, however the degree of selection bias is in line with similar studies involving unselected residents and nurses [26, 27]. Furthermore one student did not hand in a completed logbook and a further eight students did not perform any examinations with PSID and were therefore excluded from the study. The number of students not performing any examinations was probably influenced by several factors. Firstly the use of PSID in their clinical placement was not a mandatory exercise for the medical students. Secondly, as this was a trial of the use of PSID the students received specific instructions not to let the trial come in the way of their other academic responsibilities. Thirdly, the inclusion of patients was performed by the medical students themselves, which may have created a further barrier for its use. Lastly the use of ultrasound imaging is operator dependant, enthusiastic students will likely acquire more and better images reflecting a more realistic picture of it’s clinical use, i.e. those skilled in ultrasound will also be the ones using it the most.

Conclusion

Medical students with minimal training were able to use PSID as a supplement to standard physical examination and successfully acquire acceptable relevant organ images for presentation and correctly interpret these with great accuracy. Incorporating training of point-of-care ultrasound in medical student education may be one step further towards a more widespread use of ultrasound and a faster and more accurate diagnosis for patients.

Abbreviations

PSID:

Pocket-size imaging device

LV:

Left ventricular

MAE:

Mitral annular excursion

IVC:

Inferior vena cava

AA:

Abdominal aorta

AAA:

Abdominal aortic aneurysm.

References

  1. Burton JL, Underwood J: Clinical, educational, and epidemiological value of autopsy. Lancet. 2007, 369 (9571): 1471-1480.

    Article  Google Scholar 

  2. Kirch W, Schafii C: Misdiagnosis at a university hospital in 4 medical eras. Medicine (Baltimore). 1996, 75 (1): 29-40.

    Article  Google Scholar 

  3. Pastores SM, Dulu A, Voigt L, Raoof N, Alicea M, Halpern NA: Premortem clinical diagnoses and postmortem autopsy findings: discrepancies in critically ill cancer patients. Crit Care. 2007, 11 (2): R48.

    Article  Google Scholar 

  4. Moore CL, Copel JA: Point-of-care ultrasonography. N Engl J Med. 2011, 364 (8): 749-757.

    Article  Google Scholar 

  5. Kimura BJ, Amundson SA, Shaw DJ: Hospitalist use of hand-carried ultrasound: preparing for battle. J Hosp Med. 2010, 5 (3): 163-167.

    Article  Google Scholar 

  6. Andersen GN, Haugen BO, Graven T, Salvesen Ø, Mjølstad OC, Dalen H: Feasibility and reliability of point-of-care pocket-sized echocardiography. Eur J Echocardiogr. 2011, 12 (9): 665-670.

    Article  Google Scholar 

  7. Galderisi M, Santoro A, Versiero M, Lomoriello VS, Esposito R, Raia R, Farina F, Schiattarella PL, Bonito M, Olibet M, de Simone G: Improved cardiovascular diagnostic accuracy by pocket size imaging device in non-cardiologic outpatients: the NaUSiCa (Naples ultrasound stethoscope in cardiology) study. Cardiovasc Ultrasound. 2010, 8: 51.

    Article  Google Scholar 

  8. Mjølstad OC, Dalen H, Graven T, Kleinau JO, Salvesen O, Haugen BO: Routinely adding ultrasound examinations by pocket-sized ultrasound devices improves inpatient diagnostics in a medical department. Eur J Intern Med. 2012, 23 (2): 185-191.

    Article  Google Scholar 

  9. Panoulas VF, Daigeler AL, Malaweera AS, Lota AS, Baskaran D, Rahman S, Nihoyannopoulos P: Pocket-size hand-held cardiac ultrasound as an adjunct to clinical examination in the hands of medical students and junior doctors. Eur Heart J Cardiovasc Imaging. 2013, 14 (4): 323-330.

    Article  Google Scholar 

  10. Prinz C, Voigt J-U: Diagnostic accuracy of a hand-held ultrasound scanner in routine patients referred for echocardiography. J Am Soc Echocardiogr. 2011, 24 (2): 111-116.

    Article  Google Scholar 

  11. Skjetne K, Graven T, Haugen BO, Salvesen Ø, Kleinau JO, Dalen H: Diagnostic influence of cardiovascular screening by pocket-size ultrasound in a cardiac unit. Eur J Echocardiogr. 2011, 12 (10): 737-743.

    Article  Google Scholar 

  12. Badano LP, Nucifora G, Stacul S, Gianfagna P, Pericoli M, Del Mestre L, Buiese S, Compassi R, Tonutti G, Di Benedetto L, Fioretti PM: Improved workflow, sonographer productivity, and cost-effectiveness of echocardiographic service for inpatients by using miniaturized systems. Eur J Echocardiogr. 2009, 10 (4): 537-542.

    Article  Google Scholar 

  13. Mjolstad OC, Andersen GN, Dalen H, Graven T, Skjetne K, Kleinau JO, Haugen BO: Feasibility and reliability of point-of-care pocket-size echocardiography performed by medical residents. Eur Heart J Cardiovasc Imaging. 2013, 14 (12): 1195-1202.

    Article  Google Scholar 

  14. Mircea P-A, Badea R, Fodor D, Buzoianu AD: Using ultrasonography as a teaching support tool in undergraduate medical education – time to reach a decision. Med Ultrason. 2012, 14 (3): 211-216.

    Google Scholar 

  15. Mouratev G, Howe D, Hoppmann R, Poston MB, Reid R, Varnadoe J, Smith S, McCallum B, Rao V, Demarco P: Teaching medical students ultrasound to measure liver size: comparison with experienced clinicians using physical examination alone. Teach Learn Med. 2013, 25 (1): 84-88.

    Article  Google Scholar 

  16. Decara JM, Kirkpatrick JN, Spencer KT, Ward RP, Kasza K, Furlong K, Lang RM: Use of hand-carried ultrasound devices to augment the accuracy of medical student bedside cardiac diagnoses. J Am Soc Echocardiogr. 2005, 18 (3): 257-263.

    Article  Google Scholar 

  17. Swamy M, Searle RF: Anatomy teaching with portable ultrasound to medical students. BMC Med Educ. 2012, 12: 99.

    Article  Google Scholar 

  18. Sweetman GM, Crawford G, Hird K, Fear MW: The benefits and limitations of using ultrasonography to supplement anatomical understanding. Anat Sci Educ. 2013, 6 (3): 141-148.

    Article  Google Scholar 

  19. Alam M, Hoglund C, Thorstrand C: Longitudinal systolic shortening of the left ventricle: an echocardiographic study in subjects with and without preserved global function. Clin Physiol. 1992, 12 (4): 443-452.

    Article  Google Scholar 

  20. Hoglund C, Alam M, Thorstrand C: Atrioventricular valve plane displacement in healthy persons. An echocardiographic study. Acta Med Scand. 1988, 224 (6): 557-562.

    Article  Google Scholar 

  21. Alam M, Hoglund C, Thorstrand C, Philip A: Atrioventricular plane displacement in severe congestive heart failure following dilated cardiomyopathy or myocardial infarction. J Intern Med. 1990, 228 (6): 569-575.

    Article  Google Scholar 

  22. Fernandez-Frackelton M, Peterson M, Lewis RJ, Perez JE, Coates WC: A bedside ultrasound curriculum for medical students: prospective evaluation of skill acquisition. Teach Learn Med. 2007, 19 (1): 14-19.

    Article  Google Scholar 

  23. Cawthorn TR, Nickel C, O’Reilly M, Kafka H, Tam JW, Jackson LC, Sanfilippo AJ, Johri AM: Development and evaluation of methodologies for teaching focused cardiac ultrasound skills to medical students. J Am Soc Echocardiogr. 2014, 27 (3): 302-309.

    Article  Google Scholar 

  24. Sicari R, Galderisi M, Voigt JU, Habib G, Zamorano JL, Lancellotti P, Badano LP: The use of pocket-size imaging devices: a position statement of the European Association of Echocardiography. Eur J Echocardiogr. 2011, 12 (2): 85-87.

    Article  Google Scholar 

  25. Frederiksen CA, Juhl-Olsen P, Larsen UT, Nielsen DG, Eika B, Sloth E: New pocket echocardiography device is interchangeable with high-end portable system when performed by experienced examiners. Acta Anaesthesiol Scand. 2010, 54 (10): 1217-1223.

    Article  Google Scholar 

  26. Ruddox V, Stokke TM, Edvardsen T, Hjelmesaeth J, Aune E, Baekkevar M, Norum IB, Otterstad JE: The diagnostic accuracy of pocket-size cardiac ultrasound performed by unselected residents with minimal training. Int J Cardiovasc Imaging. 2013, 29 (8): 1749-1757.

    Article  Google Scholar 

  27. Henderson SO, Ahern T, Williams D, Mailhot T, Mandavia D: Emergency department ultrasound by nurse practitioners. J Am Acad Nurse Pract. 2010, 22 (7): 352-355.

    Article  Google Scholar 

Pre-publication history

Download references

Acknowledgements

We thank all the participating medical students for their invaluable assistance with the inclusion and data collection for this study. The Norwegian University of Science and Technology (NTNU) and Nord-Trøndelag Health Trust, both Norway, funded this study. GNA, OCM, HD and BOH hold positions at the Medical Imaging Laboratory, NTNU, a Centre of Research-based Innovation that is funded by the Research Council of Norway and industry. One of the industry partners is GE Vingmed Ultrasound. The Centre has a total budget of app. 124 million NOK for the 8 year period 2007–2014, and the contribution from GE Vingmed Ultrasound to this budget is app. 7 million NOK (app. 6%).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Garrett Newton Andersen.

Additional information

Competing interests

GE Healthcare provided 20 of the 30 PSID devices used during this study as a loan. No financial support was received.

Authors’ contributions

GNA participated in the study design and coordination, education of medical students, data processing, performed the statistical analysis and drafted the manuscript. AV participated in the coordination of the study, education of the medical students and review of relevant images. OCM participated in the coordination of the study, education of the medical students and review of relevant images. ØS participated in the design of the study and performed the statistical analysis. HD participated in the design of the study, performed the statistical analysis and helped to draft the manuscript. BOH conceived of the study and participated in its design and coordination, education of the medical students and review of relevant images and helped to draft the manuscript. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.

Authors’ original file for figure 1

Authors’ original file for figure 2

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Andersen, G.N., Viset, A., Mjølstad, O.C. et al. Feasibility and accuracy of point-of-care pocket-size ultrasonography performed by medical students. BMC Med Educ 14, 156 (2014). https://doi.org/10.1186/1472-6920-14-156

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1472-6920-14-156

Keywords