Longitudinal stent fracture 11 months after implantation in the left pulmonary artery and successful management by a stent-in-stent maneuver

Catheter Cardiovasc Interv. 2003 Jan;58(1):116-8. doi: 10.1002/ccd.10379.

Abstract

Fracture of a 12 mm Palmaz stent after implantation in the left pulmonary artery for palliation of postoperative stenosis in a 9-year-old child is described. Successful management by implantation of a second Palmaz stent revealed immediate stabilization and no signs of significant restenosis during 1-year follow-up.

Publication types

  • Case Reports

MeSH terms

  • Blood Vessel Prosthesis Implantation*
  • Catheterization*
  • Child
  • Humans
  • Male
  • Postoperative Complications*
  • Prosthesis Failure*
  • Pulmonary Artery / diagnostic imaging
  • Pulmonary Artery / surgery*
  • Pulmonary Valve Stenosis / diagnostic imaging
  • Pulmonary Valve Stenosis / surgery*
  • Radiography
  • Reoperation
  • Stents / adverse effects*
  • Time Factors