The prediction of the impedance of the thorax to defibrillating current

Med Instrum. 1976 May-Jun;10(3):159-62.

Abstract

In this paper a technique for predicting thoracic impedance to defibrillator pulses is described. The impedance to low-current (1.0 mA) high-frequency (10-500kHz) sinusoidal current is used as an indicator of the impedance of the thorax to high-current, damped sinusoidal waveform pulses. Results from 71 dogs to which defibrillator shocks of 4 to 220 A peak current were applied show that thoracic impedance can be predicted by this method. This information indicates that it is possible to design a defibrillator that can automatically measure chest impedance prior to a defibrillation shock and deliver a predetermined peak current to the subject.

Publication types

  • Comparative Study
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Animals
  • Dogs
  • Electric Conductivity
  • Electric Countershock*
  • Thorax / physiology*