Noninvasive assessment of cardiac output in children using impedance cardiography

Am Heart J. 1990 Nov;120(5):1166-72. doi: 10.1016/0002-8703(90)90132-h.

Abstract

This study evaluated the effect of intracardiac shunting on the accuracy of thoracic bioimpedance-derived cardiac output determinations. Twenty-six patients, ranging in age from 3 months to 17 years, underwent cardiac catheterization during which simultaneous Fick and impedance measurements of cardiac output were obtained. The subjects were divided into three groups: 10 children with no intracardiac shunts, nine children with predominant left-to-right intracardiac shunts, and seven children with predominant right-to-left intracardiac shunts. Positive correlations between impedance and Fick-derived cardiac output determinations were obtained in the non-shunt group (r = 0.84), with lower correlations in the left-to-right shunt group (r = 0.70). In the right-to-left shunt group, the impedance derived cardiac output correlated with Fick pulmonary flow (r = 0.82), but the variability was unacceptably large. Although further study is needed, impedance cardiography appears to have validity as a methodology in pediatric critical care and cardiovascular health research.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adolescent
  • Cardiac Output*
  • Cardiography, Impedance*
  • Child
  • Child, Preschool
  • Evaluation Studies as Topic
  • Female
  • Heart Septal Defects / physiopathology*
  • Humans
  • Infant
  • Male