A comparison of outcomes in men 11 years after heart-valve replacement with a mechanical valve or bioprosthesis. Veterans Affairs Cooperative Study on Valvular Heart Disease

N Engl J Med. 1993 May 6;328(18):1289-96. doi: 10.1056/NEJM199305063281801.

Abstract

Background: Mechanical heart valves are durable but thrombogenic, and their use requires that the patient receive anticoagulants. In contrast, bioprosthetic valves are less thrombogenic, but they have limited durability because of tissue deterioration.

Methods: To compare the outcomes of patients who receive these two types of valves, we randomly assigned 575 men scheduled to undergo aortic-valve or mitral-valve replacement to receive either a mechanical or a bioprosthetic valve. The primary end points were death from any cause and any valve-related complication.

Results: During an average follow-up of 11 years, there was no difference between the two groups in the probability of death from any cause (11-year probability for mechanical valves, 0.57; for bioprostheses, 0.62; P = 0.57) or in the probability of any valve-related complication (0.65 and 0.69, respectively; P = 0.39). There was a much higher rate of structural valve failure among patients who received bioprosthetic valves (11-year probability, 0.15 for the aortic valves and 0.36 for the mitral valves) than among those who received mechanical valves (no valve failures; P < 0.001). However, this difference was offset by a higher rate of bleeding complications among patients with mechanical valves than among those with bioprosthetic valves (11-year probability, 0.42 and 0.26, respectively; P < 0.001) and by a greater frequency of peri-prosthetic valvular regurgitation among patients with mechanical mitral valves than among those with mitral bioprostheses (11-year probability, 0.17 and 0.09, respectively; P = 0.05).

Conclusions: After 11 years, the rates of survival and freedom from all valve-related complications were similar for patients who received mechanical heart valves and those who received bioprosthetic heart valves. However, structural failure was observed only with the bioprosthetic valves, whereas bleeding complications were more frequent among patients who received mechanical valves.

Publication types

  • Clinical Trial
  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Aortic Valve / surgery*
  • Bioprosthesis*
  • Cause of Death
  • Follow-Up Studies
  • Heart Valve Prosthesis* / adverse effects
  • Heart Valve Prosthesis* / mortality
  • Hemorrhage / etiology
  • Humans
  • Male
  • Middle Aged
  • Mitral Valve / surgery*
  • Prosthesis Failure
  • Reoperation
  • Survival Rate
  • Treatment Outcome