Detection and significance of subclinical mitral regurgitation by colour Doppler techniques
  1. CRISPIN DAVIES,
  2. DAVID SAHN*
  1. Division of Cardiololgy
  2. *Clinical Care Center for Congenital Heart Disease
  3. Oregon Health Sciences University
  4. 3181 SW Sam Jackson Park Road
  5. Portland, 97219—3098
  6. Oregon, USA
  7. Daviesc{at}ohsu.edu

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Two papers in this issue demonstrate clinical consequences from the presence of clinically silent mitral regurgitation in paediatric and adolescent populations. Both are potentially important as the development of progressive doxorubicin cardiomyopathy1 and the presence of rheumatic carditis2 have long term implications. A parallel example, which highlights some of the problems inherent in attempting to derive clinical conclusions from the presence of mild regurgitation in an adult population, exists in the recent “phen/fen” controversy.3 Attempting to extract useful prognostic information from the presence of clinically silent, Doppler detected, mitral regurgitation through morphologically normal valves is not new—its negative impact on the prognosis of acute infarction has been recognised for over a decade.4 However, a fundamental prerequisite in attempting to utilise this type of information is an appreciation of the factors which underlie the presence of the colour flow signal. The appearance of mitral regurgitation is dependent upon the mechanical substrate for regurgitation itself, the fluid dynamics of the resultant flow disturbance, and lastly, processing using a colour flow mapping algorithm to produce an image. In particular, we need to question the reproducibility of findings obtained by different examiners and between different echo system implementations of colour flow mapping before we can feel comfortable about extrapolating these findings into clinical practice.

Mechanical substrates for regurgitation

In some very mild cases of rheumatic carditis and subtle forms of the “phen/fen” effect, leaflet damage not apparent on two dimensional echo is the intuitive mechanism of subclinical regurgitation. However, in the majority of cases, the valve is merely responding to changes in ventricular geometry, pressure loading, or both. In a series of elegant studies,5 Levine's group has demonstrated that functional mitral regurgitation is a dynamic entity, with maximal flow occurring in early and late systole accompanied by a characteristic mid-systolic dip,6 suggesting a role for the transmitral systolic pressure gradient in determining the severity of regurgitation.7 In contrast to the study of Allen and colleagues,1 three dimensional studies in animal models8 and two dimensional clinical observations,6 suggest that depressed systolic function in isolation does not result in mitral regurgitation in the absence of ventricular dilatation. The principal requirements for functional mitral regurgitation include outward papillary muscle displacement, annular dilatation, and possibly a reduction in leaflet closing forces.5

Physical properties of regurgitant jet sprays

Despite the presence of ambient counterflows in the atrium, mitral regurgitation behaves as a turbulent free jet.9-11 When flow accelerates towards a restrictive regurgitant orifice, the flow field progressively narrows in proportion to the relative increase in velocity until it passes through the orifice (fig 1). After this, the laminar core continues to focus, reaching its narrowest point (the vena contracta) just inside the left atrium. While accelerating through the isovelocity shells preceding the mitral valve and once within the vena contracta itself, flow is essentially laminar, providing the opportunity for relatively precise quantification of flow in experimental studies.12 ,13 Beyond this point there is no further acceleration and vortices develop as the kinetic energy of the laminar core is progressively dissipated by incorporation of mass from the surrounding atrial blood.14 This region is termed the jet intrusion core or flow development region and typically spans five orifice diameters. The area beyond the jet intrusion zone is termed the fully developed zone and the combination of these two regions produces a characteristic turbulent deceleration zone around and beyond the intrusion core. The jet core (or intrusion distance) varies as a function of velocity × orifice diameter and thus with larger orifices, a greater flow volume will be required to yield a high enough velocity for an equivalently long intrusion distance. Importantly, as the generation of the fully developed core is a function of kinetic energy (pressure) and not simply volume, elevations of ventricular systolic pressure produce larger jet sprays in the absence of any increase in orifice area.15 ,16 Other factors influencing the transformation of the jet core into the fully developed turbulent jet spray region include atrial pressure, pulmonary venous flow, and contact with the atrial walls.

Figure 1

Flow patterns associated with regurgitation through a restrictive orifice.

Effect of instrument settings on flow visualisation

The fundamental, and well recognised, problem is that colour flow imaging is a misnomer as it provides information on the spatial distribution of velocity but not flow itself. Flow is accessible (velocity × area = flow rate), but as we do not currently have easy access to regurgitant orifice measurements in clinical practice, we are tempted into ignoring area and relating velocity fields directly to flow. The next problem is that conventional colour Doppler techniques for the detection of regurgitant flows rely predominantly on the magnitude of the kinetically determined turbulent jet spray deceleration zone and not on quantification of the laminar flow in the acceleration zone or vena contracta. Methods to quantify this laminar flow are unsuitable for mild regurgitation (measurement of the proximal isovelocity surface area) or are currently employed only as research tools (digital analysis of the convergence zone or the use of power Doppler to quantify the vena contracta12). As outlined above, increases in the driving pressure increase the magnitude of regurgitation to the first power; however, this is compounded by the fact that the jet spray area visualised by colour Doppler increases to the second power making the appearance of the regurgitant jet particularly sensitive to changes in blood pressure. Thirdly, variations in detection rates may be attributable to changes in machine settings relating to processing of the Doppler signal or to variability between individual manufacturers. This is a subject that has received relatively little attention in the literature and would be expected to be a particular problem for relatively subtle reverse flow signals unaccompanied by turbulence. In one of the few available studies, Jones and colleagues found that for the same degree of regurgitant flow, visualised jet area increased as a function of instrument gain and transmit frequency, while decreasing with increases in pulse repetition frequency.17 In addition, changes in clutter filter settings (often linked inversely related to gain), transducer frequency, and pulse train packet size all determine the magnitude of the visualised jet spray.11 Subsequent in vitro studies have suggested that, in addition to the effect of a variety of colour maps, differences between individual manufacturers' machines are attributable to differences in the autocorrelation velocity estimation algorithm used to encode variance.18

Practical problems in using valvar regurgitation to predict clinical outcomes in adults

The two papers1 ,2 in this issue ofHeart relate to children and adolescents; in these populations the frequency of valvar regurgitation is low (< 3%)19 ,20 and its presence more likely to be pathologically significant. The problem in adults is that the frequency of mild mitral regurgitation increases with age (19% by a mean age of 54 years in the Framingham cohort,21 exceeding 80% by 80 years22) to an extent that makes it difficult to attribute much significance to its presence. Matters are further compounded by the poorer imaging windows often available in these patients.

Thus the conceptual process involved in attempting to relate pathological events to the presence of mild regurgitant flows on colour Doppler is far from straightforward. Current Doppler techniques visualise the velocity of turbulent jet sprays, not the regurgitant volume, and are heavily pressure dependent. In future, techniques utilising digital Doppler quantification or power analysis of the acceleration zone and vena contracta may provide more quantitative solutions to this problem. In the meantime any study attempting to quantify these events requires rigorous standardisation of Doppler settings, a restriction of the types of machines employed, and simultaneous measurement of blood pressure. Although the presence of mild regurgitation through apparently structurally normal valves may provide clinically useful information in paediatric populations, its high frequency of occurrence renders it of less practical use in adults.