Diagnostic performance of an acoustic-based system for coronary artery disease risk stratification
  1. Simon Winther1,
  2. Louise Nissen2,
  3. Samuel Emil Schmidt3,
  4. Jelmer Sybren Westra1,
  5. Laust Dupont Rasmussen2,
  6. Lars Lyhne Knudsen2,
  7. Lene Helleskov Madsen2,
  8. Jane Kirk Johansen4,
  9. Bjarke Skogstad Larsen3,
  10. Johannes Jan Struijk3,
  11. Lars Frost4,
  12. Niels Ramsing Holm1,
  13. Evald Høj Christiansen1,
  14. Hans Erik Botker1,
  15. Morten Bøttcher2
  1. 1 Department of Cardiology, Aarhus University Hospital, Aarhus, Denmark
  2. 2 Department of Cardiology, Hospital Unit West, Herning, Denmark
  3. 3 Department of Health Science and Technology, Aalborg University, Aalborg, Denmark
  4. 4 Department of Cardiology, Regional Hospital of Silkeborg, Silkeborg, Denmark
  1. Correspondence to Dr Simon Winther, Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, DK-8200 Aarhus, Denmark; sw{at}dadlnet.dk

Abstract

Objective Diagnosing coronary artery disease (CAD) continues to require substantial healthcare resources. Acoustic analysis of transcutaneous heart sounds of cardiac movement and intracoronary turbulence due to obstructive coronary disease could potentially change this. The aim of this study was thus to test the diagnostic accuracy of a new portable acoustic device for detection of CAD.

Methods We included 1675 patients consecutively with low to intermediate likelihood of CAD who had been referred for cardiac CT angiography. If significant obstruction was suspected in any coronary segment, patients were referred to invasive angiography and fractional flow reserve (FFR) assessment. Heart sound analysis was performed in all patients. A predefined acoustic CAD-score algorithm was evaluated; subsequently, we developed and validated an updated CAD-score algorithm that included both acoustic features and clinical risk factors. Low risk is indicated by a CAD-score value ≤20.

Results Haemodynamically significant CAD assessed from FFR was present in 145 (10.0%) patients. In the entire cohort, the predefined CAD-score had a sensitivity of 63% and a specificity of 44%. In total, 50% had an updated CAD-score value ≤20. At this cut-off, sensitivity was 81% (95% CI 73% to 87%), specificity 53% (95% CI 50% to 56%), positive predictive value 16% (95% CI 13% to 18%) and negative predictive value 96% (95% CI 95% to 98%) for diagnosing haemodynamically significant CAD.

Conclusion Sound-based detection of CAD enables risk stratification superior to clinical risk scores. With a negative predictive value of 96%, this new acoustic rule-out system could potentially supplement clinical assessment to guide decisions on the need for further diagnostic investigation.

Trial registration number ClinicalTrials.gov identifier NCT02264717; Results.

  • cardiac imaging and diagnostics
  • coronary artery disease

This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Introduction

Diagnosing patients with symptoms suggestive of obstructive coronary artery disease (CAD) continues to require substantial healthcare resources. Despite several attempts to improve clinical pretest risk scoring algorithms for obstructive CAD, the incidence of a normal diagnostic test remains high.1 2 A continued high rate of inappropriate Invasive coronary angiography (ICA) in de novo patients has been criticised in particular because it is costly and carries the risk of complications. Hence, there is a pertinent need for more accurate risk stratification tools providing high negative predictive value (NPV) to accurately rule out obstructive CAD to avoid further diagnostic testing.3

One plausible way to obtain this goal may be to use advanced analysis of sounds originating from blood flow turbulence in the coronary circulation during diastole. Acoustic detection of obstructive CAD is non-invasive, radiation free and economically favourable. This makes the technique attractive for primary risk stratification. Recent advances in sound sensor technology, analytic power and data filtering have enabled the use of acoustic detection to diagnose intracoronary turbulence due to obstructive CAD. In addition to weak murmurs related to intracoronary turbulence recent approaches have analysed further aspects of the heart sounds which might relate to myocardial compliance and resonance frequencies of the coronary artery tree.4–7 Several methodological studies describe the frequency range and timing of turbulence associated with obstructive CAD, and several companies have developed European Conformity marked equipment with different algorithms to diagnose obstructive CAD.8 These developments have paved the way for investigation of the technique in clinical practice.5 8 9 Hence, small-scale pilot studies of different acoustic devices have demonstrated NPVs in the range 71%–98% suggesting that they may be used as rule-out modalities.4 10 11

In the present study of acoustic detection of CAD, we aimed to test an acoustic device in a low to intermediate risk population using both quantitative coronary angiography (QCA) and invasive fractional flow reserve (FFR) as references. Two algorithms were tested: a previously used acoustic algorithm, CAD-score Version 2 (V2) (per protocol), which was developed using patients included in pilot studies; and a new algorithm, CAD-score V3, by combining an updated acoustic algorithm and clinical risk factors.

Materials and methods

Study design and population

The present study is a predefined substudy of the Danish study of Non-Invasive testing in Coronary Artery Disease (Dan-NICAD). Patients referred to cardiac CT angiography (CTA) as first-line test due to a history of symptoms suggestive of obstructive CAD were consecutively enrolled. Referral decision was taken during an outpatient visit by a cardiologist based on echocardiography, patient history and risk profiling according to national and European Society of Cardiology guidelines.12 Exclusion criteria were: (1) age <40; (2) previous coronary revascularisation; (3) estimated glomerular filtration <40 mL/min; (4) pregnancy; and (5) contraindication for iodine-containing contrast medium, MRI or adenosine (severe asthma, advanced atrioventricular block or critical aorta stenosis). All patients signed a written informed consent form.

All patients underwent a systematic interview to assess risk factors and symptoms to calculate the updated Diamond-Forrester score (DF score).13 In connection with the cardiac CTA, examination was performed with the acoustic CADScor®System (Acarix A/S, Denmark). Patients with suspicion of a coronary stenosis at cardiac CTA underwent ICA with FFR. Patients without stenosis at cardiac CTA were terminated from the study (figure 1). The study design, imaging protocols and analysis strategy have been published.14

Figure 1

Flow chart of patients in the study for final diagnosis of CAD. *A CAD-score Version 2 was not calculated in 190 patients and CAD-score Version 3 was not calculated in 204 patients. Median intertest interval from the cardiac CTA to the ICA was 30 days (10th and 90th percentiles: 14 and 50 days). CACS, coronary artery calcium score; CAD, coronary artery disease; CAG, invasive coronary angiography; CTA, CT angiography; FFR, fractional flow reserve; ICA, invasive coronary angiograph; QCA, quantitative coronary angiography.

CAD-score

Heart sound recordings were obtained transcutaneously with the CADScor®System (Acarix A/S) microphone mounted at the fourth intercostal space just to the left of the sternum using a dedicated adhesive patch. The patient lied down for 3 min during the recording and was asked to hold his/her breath four times, each time for 8 s (figure 2). As the equipment may be influenced by excessive noise, the tests were conducted in an undisturbed room. The CAD-score was estimated immediately after the recording with a fully automated and device embedded algorithm.4 5 9 15 If the recording quality was insufficient the recording device requested a new recording. In these cases, a second attempt was made to calculate the CAD-score. The algorithm combines four measures (Features) of acoustic properties from the diastolic heart sounds which relate to murmurs and other heart sound characteristics into a score (figure 2 and online supplementary material).

Supplementary file 1

Figure 2

Depiction of the CAD-score acquisition and the two CAD-score algorithms. The heart sounds were recorded at the fourth intercostal space with patients in a supine position. The preprocessing part of the algorithm organised the heart sounds for analysis through segmentation and filtering. After preprocessing, the acoustic features were extracted from the sounds and an acoustic score was constructed using LDA. In CAD-score Version 3 (V3), the acoustic score was further combined with gender, age and hypertension using logistic regression. AMI, automutal information; FPR, frequency power ratio; HRV, heart rate variability; LDA, linear discriminant analysis; PCARand, principle component analyses-based measure of the randomness; PCASpec, principle component analysis of the diastolic frequency spectrum; S2freq, frequency distribution of the second heart sounds; S4amp, amplitude of the fourth heart sound; SampEn, sample entropy; SpecSlope, slope of diastolic frequency spectrum; SysFPR, systolic frequency power ratio.

An updated score, using the CAD-score V3 algorithm, was calculated using postprocessing of the audio recordings obtained with the recording device. The updated algorithm measures eight acoustic properties and the resulting acoustic score was combined with the clinical risk factors gender, age and hypertension, defined as systolic blood pressure ≥140 mm Hg or receiving antihypertension medicine (figure 2 and online supplementary material). The updated CAD-score V3 algorithm was developed using a heart sound database (Acarix A/S) with recordings from patients included in previous pilot studies, published data, and a training cohort of the first 593 consecutively enrolled patients from the present Dan-NICAD study.4 The remaining 1082 patients in the Dan-NICAD study were used as a validation cohort and remained blinded to the developer of the CAD-score V3 algorithm.

Coronary CTA

All cardiac CTA scans were performed using a 320-slice volume CT scanner (Aquilion One, Toshiba Medical Systems, Japan) according to clinical guidelines. CT imaging analysis included an Agatston calcium score and evaluation of luminal diameter stenosis estimation in each segment of the coronary tree using an 18-segment model. Coronary lesions were evaluated blinded to the invasive angiographic findings by an experienced cardiologist. Stenosis severity was classified in all segments with a reference vessel diameter >2 mm as no stenosis: 0% diameter reduction; mild stenosis: 1%–29% diameter reduction; moderate stenosis: 30%–49% diameter reduction; and severe stenosis: 50%–100% diameter reduction. Segments with suspected severe stenosis and non-evaluable segments with a reference vessel diameter >2 mm were defined as having obstructive CAD by cardiac CTA. The cardiac CTA was defined as abnormal if obstructive CAD was not ruled out in all coronary segments. CAD severity was further categorised based on the coronary artery calcium score (CACS) and coronary stenosis severity; non-CAD: CACS=0 and no plaque; mild CAD: CACS≥0 and plaque with maximal mild stenosis; moderate CAD: CACS≥0 and plaque with moderate stenosis; and severe CAD: CACS≥0 and plaque with severe stenosis.

Invasive coronary angiography

The ICA examination was performed according to clinical guidelines. Intracoronary nitroglycerine was administered before the first angiographic acquisition.

QCA analysis was done using QAngioXA V.7.3 software in an independent core lab (ClinFact, Leiden, The Netherlands). Anatomically significant obstructive CAD was defined as a coronary artery with a luminal diameter reduction ≥50% in a segment with a reference vessel diameter >2 mm.

Invasive FFR was measured in lesions with a visualised, estimated 30%–90% diameter stenosis located in vessels with a reference diameter >2.0 mm using a clinical FFR system (Aeris, St. Jude Medical, USA). Haemodynamically significant obstructive CAD was identified in a blinded core lab as: (1) FFR value <0.80, (2) luminal diameter stenosis reduction >90%, or (3) luminal diameter stenosis reduction ≥50% if FFR was indicated but not performed for technical, anatomical or other reasons.

Statistical analysis

Variables are expressed as mean (±SD or total range) or median (range) according to the Gaussian distribution. Categorical variables are reported as frequencies (percentages). The unpaired Student’s t-test and analysis of variance test were used for comparison between Gaussian-distributed variables. Wilcoxon rank-sum test and the χ2 test were used for comparison between non-Gaussian distributed and categorical variables, respectively. Pearson’s test and Spearman’s test were used to analyse correlations of variables of Gaussian and non-Gaussian distributions, respectively. The area under the receiver operating characteristic curve was calculated for continuous variables and compared with the method described by DeLong et al.16 The CAD-score was divided as a binary variable with a cut point of >20; and sensitivity, specificity, positive predictive value, NPV, and positive and negative likelihood ratios were calculated.

For development of CAD-score V3, the cohort was split in training and validation cohorts according to a prespecified date. Diagnostic accuracy of CAD-score V3 for the training and validation cohort is presented in online supplementary table S2 and pooled together in the Results section due to similar diagnostic performance in the two cohorts. Statistical analyses were performed using STATA V.13.1 (StataCorp, USA).

Results

A total of 1675 patients were included. Eleven patients (0.7%) were excluded because they did not complete either the CACS or the cardiac CTA. The acoustic CAD-score V2 could not be obtained in 190 patients (11.3%), and CAD-score V3 was missing in another 14 patients (online supplementary table S1). Of these, four patients (0.2%) had missing cardiac CTA data (figure 1).

Baseline and cardiac CTA characteristics are summarised in table 1. Of the patients referred for ICA, 325 patients (91.8%) completed the investigation. Of these, 153 (47.1%) were classified with obstructive CAD by core lab QCA, while 145 patients (44.6%) had haemodynamically obstructive disease based on ICA and FFR (figure 1, table 1). In total, the prevalence of anatomically significant obstructive CAD was 10.6%, and the prevalence of haemodynamically significant obstructive CAD was 10.0%. No adverse events related to the acoustic system were recorded.

Table 1

Patient demographics and imaging study characteristics

CAD-score V2

The diagnostic accuracy of acoustic CAD-score V2 for diagnosing anatomically obstructive stenosis evaluated by area under the curve (AUC) was 58.1% (95% CI 53.6% to 62.6%), which was significantly lower than the DF score, 65.9% (95% CI 61.2% to 70.4%) (P<0.01). Diagnostic accuracy parameters with both anatomically and haemodynamically obstructive stenosis as references are listed in table 2. There were no differences in diagnostic accuracy with the use of anatomically versus haemodynamically obstructive stenosis as a reference.

Table 2

Diagnostic accuracy of CAD-score with a cut-off of >20 according to a reference of (A) anatomically obstructive stenosis diagnosed with core lab QCA and (B) haemodynamically obstructive stenosis diagnosed with invasive FFR as reference

CAD-score V3

The acoustic part of the CAD-score V3 was higher for men than for women, 25.2±10.4 vs 19.3±9.7 (P<0.001). Only a weak correlation was demonstrated with age (r=0.11, P<0.001), body mass index (r=−0.19, P<0.001), systolic blood pressure (r=0.10, P<0.001), diastolic blood pressure (r=0.02, P=0.50) and CACS (r=0.17, P<0.001). In the presence of a coronary stenosis at ICA, a weak correlation with max QCA percentage diameter stenosis was observed (r=0.14, P<0.05). Patients with anatomically obstructive stenosis had a significantly higher acoustic CAD-score V3 than patients without, 24.4±10.2 vs 21.9±10.4 (P<0.01). The AUC for diagnosing anatomically obstructive stenosis of the acoustic part of the CAD-score V3 was 63.3% (95% CI 58.7% to 67.8%). This was significantly higher than the previous acoustic CAD-score V2 (P<0.05), and similar to that of the DF score (P=0.40).

The final CAD-score V3, which includes both an acoustic part and risk factors, correlated moderately with CACS (r=0.38, P<0.001). The CAD-score V3 was associated with CAD severity when stratified by (1) CACS groups (P<0.001), (2) CAD disease severity defined by cardiac CTA (P<0.001) and (3) obstructive disease (P<0.001) identified by cardiac CTA (figure 3).

Figure 3

CAD-score Version 3 divided by: (A) coronary artery calcium score groups; (B) CAD disease severity defined by cardiac CTA; (C) haemodynamically obstructive coronary vessel disease by cardiac CTA. Box plot illustrated median, IQR and adjacent values.  CACS, coronary artery calcium score; CAD, coronary artery disease; CTA, CT angiography; LM, left main.

The mean CAD-score V3 was higher for patients with haemodynamically significant CAD than for patients without, 29.9±12.0 vs 20.4±12.0 (P<0.001). In patients with a stenosis on ICA, CAD-score V3 correlated with maximal QCA diameter stenosis (r=0.23, P<0.001) and increased with QCA diameter stenosis severity (P<0.01) (online supplementary figure S1). However, no such correlation was present for FFR measurements in patients with 30%–90% diameter stenosis on visual estimate (r=0.01, P=0.98). There was a statistically significant stepwise increase in CAD-score V3 with increasing number of diseased vessels. There was no relation between CAD-score V3 and the anatomical location of the stenosis (table 3).

Supplementary file 2

Table 3

Relation between haemodynamically obstructive disease distribution at ICA and mean CAD-score Version 3 (n=320)

The diagnostic accuracy for the training (n=593) versus the validation (n=1082) cohort is summarised in online supplementary table S2.

Table 2 features the CAD-score V3 diagnostic accuracy with both anatomically and haemodynamically obstructive stenosis as references. The CAD-score V3 for diagnosing anatomically obstructive stenosis had an AUC of 72.4% (95% CI 68.0% to 76.8%). This was significantly higher than the AUC of both the DF score 65.9% (P<0.01) and the CAD-score V2 58.1% (P<0.001), and the acoustic part of the CAD-score V3 63.3% (P<0.001) when analysed in the full cohort. The AUCs of these algorithms for haemodynamically obstructive stenosis with FFR as a reference are illustrated in figure 4. Frequency plot, 2×2 table, and sensitivity and specificity curves with haemodynamically obstructive stenosis as reference are illustrated in figure 5. In total, 27 patients had false-negative CAD-scores and of these, 48.1% were men, mean age was 52.7 years, mean body mass index was 27.1, and mean left ventricular ejection fraction was 60.4%.

Figure 4

The area under the receiver operating characteristic curve for Diamond-Forrester score, acoustic CAD-score Versions 2 and 3, and CAD-score Version 3 with haemodynamically obstructive stenosis diagnosed as reference.

Figure 5

Plots of CAD-score Version 3 accuracy of haemodynamically obstructive coronary stenosis diagnosed with fractional flow reserve (FFR) as reference. Illustrated are a frequency plot, 2×2 table according to a binary CAD-score cut-off >20, and sensitivity and specificity curves with 95% CI bands shown.

Discussion

In this large, first-of-its-kind prospective trial, we evaluated a novel acoustic technique for primary risk stratification and rule-out of obstructive CAD. We studied a cohort of symptomatic patients with low to intermediate risk of CAD referred to cardiac CTA. This patient cohort represents a relevant clinical setting where the use of an accurate rule-out device could have clinical impact. All patients in the trial had either a cardiac CTA or an ICA to rule out coronary stenosis. Furthermore, suspected stenosis at ICA was verified with both core lab QCA and invasive FFR.

CAD-score V2

The isolated acoustic CAD-score V2 had a significantly lower diagnostic accuracy for predicting coronary stenosis than a DF score which is recommended in clinical guidelines. However, both the CAD-score V2 and the DF score performed poorly in this low-risk cohort of patients referred to cardiac CTA (prevalence of CAD=10%) compared with a previous study with a moderate/high-risk cohort consisting of patients referred to either cardiac CTA or ICA (prevalence of CAD=28%).4 Hence, present versus previously published AUCs with anatomically obstructive stenosis as reference were: CAD-score V2: 58% vs 72%; and DF score: 65% vs 79%. The CAD-score V2 was developed using 228 primarily moderate/high-risk patients and previously only validated by cross validation. These facts might have induced an overfit of the algorithm to cohorts with a higher CAD prevalence.

CAD-score V3

Data from the updated CAD-score V3 are presented from both the training and the validation cohorts owing to similar accuracy between the two cohorts. A moderate association between CAD-score V3 and CACS was detected in this study; we demonstrated a significant difference in the mean CAD-score between the CACS groups. The acoustic risk score as determined by either the CADScor®System device or the Cardiac Sonospectrographic Analyzer (CSA, SonoMedica, Vienna, Virginia) has previously shown weak correlation with CACS.4 10 Hence, the inclusion of risk factors (gender, age and hypertension) to the algorithm increased the correlation with CACS, which previously demonstrated high prognostic value.17 18

Comparable accuracy was detected when using QCA and FFR as reference standard of disease despite the fact that the CAD-score was optimised in order to detect stenosis accessed by QCA. Compared with the DF score, the CAD-score V3 was more accurate in detecting anatomically obstructive CAD, and a trend was seen towards a high accuracy for haemodynamically obstructive CAD. In addition, CAD-score V3 increased with the number of affected coronary vessels by cardiac CTA and ICA. Importantly, CAD-score V3 was not related to the location of the stenosis.

Acoustic recording and diagnostic algorithm

Several acoustic methods are currently being tested for detection of coronary stenosis.8 The recording system used in this study is mounted by a single-use adhesive patch at the fourth intercostal space, and recordings are performed automatically during four short breath holds over a 3 min recording period. The result is subsequently calculated and presented on the device display. Other systems use multiple record positions, handheld and elastic strap attachment, and one system performs records simultaneously with an electrocardiographic signal. The methodological differences have inherent limitations and advantages. The adhesive patches may reduce microphone noise and inter-recording variability. Single versus multiple recording positions may limit the ability to differentiate between stenosis locations. Individual body compositions may be crucial for correct placement of the recording devices and sound quality. Similarly, there are major differences in software algorithms and which spectral frequency ranges are included.8 These differences need to be highlighted to further optimise both the equipment and the algorithm. Finally, studies of how patient characteristics, coronary plaque length, location, compositions and flow impact the acoustic patterns are lacking and insights from such studies could potentially enhance system performance.

The latest UK NICE (National Institute for Health and Care Excellence) guideline update CG95: Chest pain of recent onset, includes the acoustic device as a clinically relevant prediction model with an AUC ranging from 72% to 82%, alone or in combination with the DF score.

Comparing the sensitivity of this acoustic device with the sensitivity of other large-scale, presently used techniques is also relevant; for example, exercise ECG (sensitivity 45%–50%), exercise stress echocardiography (sensitivity 80%–85%) and exercise stress SPECT (73%–92%).12 Thus, the above-mentioned techniques have false negative rates above or similar to those reported in the present study. Nonetheless, these other diagnostic modalities have previously all been proven to hold substantial prognostic value, which still needs to be demonstrated for acoustic rule-out algorithms.

The health economic impact of acoustic rule-out of patients with symptoms suggestive of obstructive CAD could be substantial. As mentioned above, the acoustic system seems to perform similarly to other established techniques in ruling out CAD, but its use comes at a potentially much lower cost.

Limitations

The cohort was almost exclusively Caucasian (99%), and with a documented low to intermediate prevalence of CAD. Thus, generalisation to other populations with other ethnic backgrounds, higher prevalence and other healthcare environments may not be possible. Higher CAD-scores were observed for a low number of patients with heart valve disease, however only when adding the clinical risk factors, indicating that patients with heart valve disease are diagnosed accurately. Finally, the development of CAD-score V3 required splitting of the cohort into a training and a validation cohort, which could have overfitted the results in the total cohort analysis.

Conclusion

The sound-based recording device, the CADScorSystem, enables risk stratification in patients with suspected CAD. With an NPV of 96%, this new acoustic rule-out system could potentially supplement clinical assessment to guide decisions on the need for further investigation and thereby reduce the demand for more advanced and costly diagnostic modalities.

Key messages

What is already known on this subject?

 Recent advances in sound sensor technology, analytic power and data filtering have enabled the use of acoustic detection to diagnose intracoronary turbulence due to obstructive coronary artery disease (CAD).

What might this study add? 

In this large, first-of-its-kind prospective trial, we demonstrated that acoustic detection of CAD enables risk stratification in patients with suspected CAD.

How might this impact on clinical practice?

The negative predictive value of this device suggests that it may be used as a rule-out modality, which is highly relevant in daily clinical practice.

References

Footnotes

  • Contributors Substantial contribution to conception and design: SW, LN, SES, NRH, HEB and MB. Collection of data: SW, LN, JSW, LDR, LLK, LHM, JKJ and LF. Analysis and interpretation of data: SW, LN, SES, JSW, BSL, JJS, EHC, HEB and MB. All authors have worked on drafting the article or revising it critically and approved the final version.

  • Funding The research reported here is partly financed by The Danish Heart Foundation (grant no. 15-R99-A5837-22920), The Hede Nielsen Foundation, and by an unrestricted institutional research grant from Acarix A/S.

  • Competing interests The current research is financed partly by Acarix A/S by an unrestricted grant. SES is a minor shareholder and part-time consultant in Acarix A/S. BSL is an industrial student at Acarix A/S. MB is part of the Acarix A/S advisory board. MB and SW received an unrestricted institutional research grant from Acarix A/S.

  • Ethics approval The study was approved by The Central Denmark Region Committees on Health Research Ethics and The Danish Data Protection Agency.

  • Provenance and peer review Not commissioned; externally peer reviewed.