10 Inter-field strength agreement of myocardial strain measurements using tissue tracking cardiovascular magnetic resonance imaging
Free
  1. S Ayton,
  2. L Athithan,
  3. KS Parke,
  4. JV Wormleighton,
  5. C Budgeon,
  6. GP McCann,
  7. GS Gulsin
  1. Department of Cardiovascular Sciences, University of Leicester, and NIHR Leicester Biomedical Research Centre, Glenfield Hospital, Leicester, UK

Abstract

Funding acknowledgement The participants included in this study were recruited as part of a larger study funded by the National Institute for Health Research Career Development Fellowship (CDF 2014-07-045).

Background The agreement of left ventricular (LV) strain measurements using cardiovascular magnetic resonance (CMR) tissue tracking at different field strengths is unknown. We compared inter-field strength agreement of systolic and diastolic strain/strain rates at 1.5T and 3T.

Methods We prospectively enrolled 17 healthy volunteers and 3 participants with diabetes but no cardiovascular disease, from a diet and exercise intervention study. Subjects underwent CMR balanced steady state free precession cine imaging with retrospective ECG gating at 3T followed by 1.5T on the same day. Acquisition voxel size was 1.66 × 1.33 × 8 mm and 1.90 × 1.52 × 8 mm, respectively, temporal resolution 48 ms and 38 ms, TR 3.44 and 2.76 and TE 1.51 and 1.15. Acquisitions were reconstructed to 30 phases. Short axis cine images at 1.5T were obtained following adenosine stress and gadolinium contrast, whereas short axis images at 3T and all long axis images were acquired before contrast administration. Scans were analysed offline by a single experienced observer blinded to patient details, using cmr42 software (Circle Cardiovascular Imaging, Calgary, Canada). End-diastolic epicardial and endocardial and end-systolic endocardial borders were defined manually and propagated by the software throughout the cardiac cycle to derive longitudinal, circumferential and radial global systolic strain (GLS, GCS and GRS) and peak early diastolic strain rates (longitudinal, circumferential and radial PEDSR). Inter-field strength agreement was assessed with intraclass correlation (ICC) and Bland Altman plots.

Results Mean age was 50±6.3 years and 50% were male. LV function, volumetric and strain data are presented in table 1. Mean systolic strain measurements were higher at 1.5T versus 3T (all p<0.05). Inter-field strength agreement was fair for GCS and GRS (ICC=0.49). Despite all images being acquired before contrast, there was no agreement for GLS (ICC=0; table 1). Bland Altman plots are shown in figure 1. All PEDSR measurements had poor agreement (ICC range 0–0.3; table 1).

Abstract 10 Table 1

Left ventricular function, volume and strain measurements

Abstract 10 Figure 1

Bland Altman plots for systolic strain measurements; a. global longitudinal strain (GLS), b. global circumferential strain (GCS), c. global radial strain (GRS)

Conclusion In this pilot study, strain measurements showed fair inter-field strength agreement for GCS and GRS and no agreement for GLS. This may relate to different signal-to-noise characteristics between field strengths. Further work is needed with randomised comparisons, but these data suggest strain measurements obtained at 1.5T and 3T are not interchangeable.