Further observations on the effect of balloon size on the short term and intermediate term results of balloon dilatation of the pulmonary valve.
  1. P S Rao
  1. Department of Paediatrics, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia.

    Abstract

    The relation between the size of the balloon used for dilatation of the pulmonary valve and the extent of relief of pulmonary stenosis both immediately after and at intermediate term follow up was studied. Sixty four balloon dilatation procedures in 56 patients were divided into group 1-12 in which the ratio of the diameter of the balloon to that of the pulmonary annulus was less than or equal to 1.0--and group 2-52 in which the ratio was greater than 1.0. Both groups had similar mean (SD) pressure drops across the pulmonary valve before dilatation. Immediately after dilatation there was a significant reduction in the pulmonary valve gradient in both group 1 (84.3 (39.2) v 43.6 (26.8) mm Hg) and group 2 (92.8 (42.1) v 22.4 (13.6)mm Hg). On intermediate term follow up (6-34 months), residual pulmonary valve gradients were significantly lower in group 2 (20.8 (18.5)mm Hg) than in group 1 (75.0 (49.4) mm Hg), suggesting that restenosis was more common after dilatation with small balloons. These data suggest that although the immediate results with either small or large balloons are good, balloons that are larger than the pulmonary valve annulus produce more sustained relief of pulmonary stenosis. Further analysis within the group treated with larger balloons showed that the subgroup with a balloon/annulus ratio of 1.01-1.2 had more recurrences of stenosis (need for repeat balloon dilatation and larger number of patients with residual pulmonary valve gradients greater than 30 mm Hg) than subgroups with balloon/annulus ratios for diameters of 1.21-1.41, greater than 1.41, and greater than 1.5, in which there were no recurrences. Balloons that were > 1.5 times the size of the pulmonary valve annulus had no additional advantage over the other subgroups, namely, 1.21-- 1.4 and > 1.41. These results and reports of damage to the right ventricular outflow tract by oversized (> 1.5) balloons indicate that balloons that give a balloon/annulus ratio 1.2 -- 1.5 are the best for dilatation of the pulmonary valve.

    Statistics from Altmetric.com

    Request Permissions

    If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.