Histopathological validation of high frequency epicardial echocardiography of the coronary arteries in vitro.
  1. A Kenny,
  2. C A Fuller,
  3. N R Cary,
  4. L M Shapiro
  1. Cardiac Unit, Papworth Hospital, Cambridge.

    Abstract

    The accuracy and reliability of measurement of coronary artery dimensions and detection of atherosclerotic lesions by high frequency epicardial echocardiography were compared with histopathological results. Ten pressure perfused human hearts were examined in vitro with a 10 MHz (Diasonics) transducer and a 7.5 MHz (Vingmed/Sonotron) transducer. There was close agreement between ultrasound and pathological measurements of coronary artery luminal diameter. Qualitative changes in wall structure such as diffuse wall thickening and calcification were readily identified; however, the resolution of the transducers was not high enough accurately to measure wall dimensions in normal coronary arteries. Coefficient of variation measurements for intra and inter observer variability (5.2% and 6.9% respectively) showed excellent reproducibility. The technique was accurate in identifying atherosclerotic lesions, imaging arteries distal to an occlusion, locating deeply sited arteries, and identifying complete obliteration of an artery. Intraoperative video playback and transducer miniaturisation may minimise problems caused by cardiac movement and restricted access. With these developments intraoperative assessment of coronary artery disease may become a real possibility.

    Statistics from Altmetric.com

    Request Permissions

    If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.