Interventricular and intra-left ventricular electromechanical delays in right ventricular paced patients with heart failure: implications for upgrading to biventricular stimulation
Free
  1. P Bordachar,
  2. S Garrigue,
  3. S Lafitte,
  4. S Reuter,
  5. P Jaïs,
  6. M Haïssaguerre,
  7. J Clementy
  1. Hopital Cardiologique du Haut-Leveque, University of Bordeaux, Pessac, France
  1. Correspondence to:
    Dr S Garrigue
    Cardiac Pacing and Clinical Electrophysiology Department, Hôpital Cardiologique du Haut-Leveque, 19, avenue de Magellan, Pessac Cedex 33604, France; stephane.garriguechu-bordeaux.fr

Abstract

Objective: To correlate, in patients with right ventricular pacing (RVP), the QRS width with electromechanical variables assessed by pulsed Doppler tissue imaging echocardiography. Secondly, to find reliable parameters for selecting RVP patients who would respond to biventricular pacing (BVP).

Methods: 26 randomly selected control patients with RVP (mean (SD) ejection fraction 74 (3)%) (group A) were matched on sex and age criteria with 16 RVP patients with drug resistant heart failure (mean (SD) ejection fraction 27 (5)%) (group B). All patients were pacemaker dependent and all underwent pulsed Doppler tissue imaging echocardiography. This technique provided the intra-left ventricular (LV) electromechanical delay and the interventricular electromechanical delay. The Gaussian curve properties of data from group A patients provided the normal range of ECG and echographic parameters.

Design: Prospective study.

Setting: University hospital (tertiary referral centre).

Results: Data from the control group showed that an interventricular electromechanical delay or an intra-LV electromechanical delay > 50 ms would identify patients with a significantly abnormal ventricular mechanical asynchrony (p < 0.05). In the same manner, a QRS width > 190 ms was considered significantly larger in group B patients (p < 0.05) than in controls. In Group B patients, there was no correlation between the QRS width and the interventricular electromechanical delay (r  =  −0.23, NS) or the intra-LV electromechanical delay (r  =  0.19, NS). Seven group B patients (44%) were misclassified by ECG criteria for ventricular mechanical asynchrony identification: four patients (25%) had a QRS width similar to that of controls but with a significantly prolonged intra-LV electromechanical delay and interventricular electromechanical delay; and three patients (19%) had a QRS width significantly larger than that in controls but without significant ventricular mechanical asynchrony.

Conclusions: The QRS width is not a reliable tool to identify RVP patients with ventricular mechanical asynchrony. In RVP patients, an interventricular electromechanical delay or intra-LV electromechanical delay > 50 ms reflects a significant ventricular mechanical asynchrony and should be required to select patients for upgrading to BVP.

  • biventricular pacing
  • asynchrony
  • QRS width
  • heart failure
  • BVP, biventricular pacing
  • LV, left ventricle
  • NYHA, New York Heart Association
  • RVP, right ventricular pacing

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Biventricular pacing (BVP) has recently been proposed as a new treatment option for patients with severe heart failure and large QRS width.1–4 By providing early stimulation in the late activated segments, BVP improves systolic wall motion coordination and optimises left ventricular (LV) filling.5–7 A QRS width above 200 ms has been arbitrarily required to upgrade right ventricular pacing (RVP) in patients with heart failure to BVP8,9 because such a wide QRS has been suggested to correspond with notable inter- or intra-LV mechanical asynchrony.8–10 However, the relation between electrical ventricular asynchrony (that is, the QRS duration) and the presence of objective mechanical ventricular asynchrony in RVP patients has never been clearly characterised. New echocardiographic techniques such as tissue Doppler imaging provide reproducible and non-invasive assessment of ventricular electromechanical parameters.5–7,11

The aim of the present study was to assess the potential relation between the QRS duration and the presence of interventricular or intra-LV asynchrony in RVP patients with drug resistant heart failure. Additionally, in the same patients, the correlation between a QRS duration > 200 ms and the presence of interventricular or intra-LV asynchrony was assessed.

METHODS

Inclusion criteria

The patients had to meet the following criteria to be included in the study: apical RVP, complete atrioventricular block or His bundle ablation to consider only pacemaker dependent patients, pacemaker implanted for at least six months, and sinus rhythm with a resting heart rate between 60−65 beats/min. Two groups of patients were considered in the present study. Group A patients (controls) were strictly asymptomatic, without echocardiographic structural cardiac disease or history of ischaemic or primitive cardiomyopathy. These control patients were matched by sex and age with group B patients (see below) and consecutively included in the study. Group B (studied population) patients had ischaemic or idiopathic dilated cardiomyopathy, New York Heart Association (NYHA) functional class III or IV despite maximum tolerated medical treatment (angiotensin converting enzyme inhibitors, β blockers, diuretics, spironolactone), and LV dysfunction (ejection fraction < 35%).

Study protocol

Both patient group underwent the same protocol. The longest QRS duration was measured on a 12 lead surface ECG recorded at a speed of 50 mm/s. The pacemaker was programmed to DDD mode and the atrioventricular delay was adjusted in every patient. According to Ritter and colleagues,12 the atrioventricular delay that provided the longest filling time without truncation of the A wave by means of pulsed Doppler analysis of the transmitral flow was chosen. The first part of the echocardiographic study was a conventional examination (Sequoia, Acuson Inc, Mountain View, California, USA). Five consecutive beats were averaged. Mitral Doppler flow was recorded to measure both E and A wave maximal velocities and the total diastolic filling time. The aortic and pulmonary ejection flows were measured respectively in the four chamber apical and parasternal views. The aortic pre-ejection time interval was defined as the time duration between the stimulation spike on the surface ECG and the onset of the aortic ejection flow, whereas the pulmonary pre-ejection time interval was defined as the time duration between the stimulation spike on the surface ECG and the onset of the pulmonary ejection flow. Tissue Doppler imaging was performed in the pulsed Doppler mode. Acoustic power, gain, dynamic range, and filters were set for each myocardial area analysed once the optimal atrioventricular delay was determined and programmed. The explored areas were respectively the basal LV septum, free wall, and inferior and anterior walls. The LV wall electromechanical delays (defined as the time duration between the stimulation spike on the surface ECG and the onset of the S wave corresponding to the systolic motion of a given LV wall obtained by the tissue Doppler imaging technique) (fig 1) were quantified and compared between group A and B patients.7,11 The intra-LV electromechanical delay was calculated as the time difference between the shortest and the longest LV wall electromechanical delays (fig 1) whereas the interventricular electromechanical delay was the time difference between the aortic and pulmonary pre-ejection time intervals.

Figure 1

The top panel shows, in a patient with sinus rhythm and apical right ventricular complete capture, the value of the shortest electromechanical delay (EMD) calculated for the septal left ventricular (LV) wall (170 ms). The QRS duration reaches 215 ms. The EMD was defined in this example by the interval between the stimulation spike and the onset of the S wave corresponding to the systolic motion of the septal LV wall. The bottom panel shows the longest EMD for the lateral LV wall (253 ms), showing highly heterogeneous systolic LV dynamics and an intra-LV asynchrony reaching 83 ms (range [253–170).

Statistical analysis

Group A patients (control group) were included in the study to characterise the range of the intra-LV and interventricular electromechanical delays in the population of RVP patients with a normal cardiac function. This was defined as the following: the statistical α risk was fixed at 0.05 so that the physiological range of the parameters would be included in the mean (2 SD) range, which accounts for 95% of the control group distribution.13 The international statistical consensus requires a sample size of at least 25 subjects for using these specific Gaussian curve properties.13 Consequently, the upper limit for a statistically normal value of intra-LV and interventricular electromechanical delays was the respective mean (2 SD) value. Any value above this limit in group B patients was considered significantly different from the control group and, accordingly, the patient was classified as presenting with an interventricular or intra-LV significant asynchrony compared with the control population.

The intraobserver correlations for intra-LV and interventricular electromechanical delay calculation were, respectively, 0.93 and 0.95, and the interobserver correlations were, respectively, 0.89 and 0.93 in a sample size of 15 patients.

Differences between patient groups were assessed by using Student’s t test for quantitative variables and Fisher’s test of exact probability for qualitative variables. Correlations between quantitative variables were determined by Pearson’s correlation coefficient. A probability value of p < 0.05 was considered significant. All data are presented as mean (SD) or percentages.

RESULTS

Study population

Forty two patients participated in the study. Twenty six consecutive patients (18 men, mean (SD) age 68 (12) years) constituted group A (control group). By definition, they were strictly asymptomatic. The mean LV ejection fraction was relatively high at 74 (3)%. Nineteen patients (73%) were His ablated for brady–tachy syndrome and seven (27%) had pacemakers implanted for complete atrioventricular block. There was no significant difference between the two groups in terms of sex, age, and pacemaker indication (table 1). Sixteen patients (14 men, mean (SD) age 63 (11)) constituted group B. The mean LV ejection fraction was 27 (5)%. Nine patients had an ischaemic cardiomyopathy and seven an idiopathic cardiomyopathy (table 1). The mean QRS width before pacemaker implantation in group B patients was 142 (23) ms versus 89 (8) ms in group A (p < 0.01). Eleven patients (68%) received a pacemaker for brady–tachy syndrome followed by His bundle ablation and five (32%) for complete atrioventricular block. The mean paced QRS width was 156 (11) ms in group A versus 210 (20) ms in group B (p < 0.01) (table 1). Twelve patients (75%) in group B and none in group A had a QRS duration > 200 ms (p < 0.01). Ten group B patients were in NYHA functional class III and six in class IV after at least six months of RV pacing.

Table 1

Patients characteristics

Echocardiographic data

Control group: group A

In patients without cardiomyopathy, the mean aortic pre-ejection time interval was 150 (22) ms and the mean pulmonary pre-ejection time interval was 117 (22) ms. These measurements resulted in a mean interventricular electromechanical delay reaching 32 (10) ms and a mean intra-LV electromechanical delay of 33 (9) ms. Accordingly (see Statistical analysis section), all group B patients with a value above 50 ms (mean (2 SD) of control patients) were considered to present with a significantly abnormal interventricular or intra-LV asynchrony. All control patients presented with a value of interventricular and intra-LV electromechanical delay < 50 ms. No significant correlation was found between the QRS width and the interventricular electromechanical delay (r  =  0.20, NS) or the intra-LV electromechanical delay (r  =  0.06, NS). There was a significant correlation between the aortic pre-ejection time interval and the interventricular delay (r  =  0.48, p  =  0.03) but not the intra-LV delay (r  =  0.26, NS).

Patients with heart failure: group B

Table 2 summarises the echocardiographic data. In patients with severe heart failure, the mean aortic pre-ejection time interval was 208 (23) ms (p < 0.01 compared with group A). No significant correlation was observed between the QRS width and the aortic pre-ejection time interval (r  =  0.12, NS). The mean pulmonary pre-ejection time interval was 162 (33) ms (p < 0.01 compared with group A) without significant correlation with the QRS width (r  =  0.18, NS). The mean interventricular delay was 55 (14) ms (p < 0.01 versus group A); no significant correlation was found with the QRS width (r  =  −0.23, NS) (fig 2, left panel). The mean intra-LV electromechanical delay was 52 (13) ms (p < 0.01 compared with group A) without significant correlation with the QRS duration (r  =  0.19, NS) (fig 2, right panel). There was a significant but poor correlation between the aortic pre-ejection time interval and the interventricular delay (r  =  0.33, p  =  0.05). The correlation between the aortic pre-ejection time interval and the intraventricular delay was also significant but poor (r  =  0.49, p  =  0.04). Four patients in group B had a QRS duration < 200 ms and all of them presented with an intra-LV or interventricular asynchrony (> 50 ms) (fig 2). In the 12 remaining group B patients who had a QRS duration > 200 ms, three exhibited neither interventricular nor intra-LV asynchrony compared with control patients (fig 2). No significant difference was observed between patients with ischaemic and those with primitive cardiomyopathy in terms of LV ejection fraction, QRS width, aortic and pulmonary pre-ejection time interval, and interventricular electromechanical delay. In contrast, the intra-LV electromechanical delay was significantly longer (p < 0.05) in patients with ischaemic cardiomyopathy. The first activated LV wall was the septal one in 75% of group B patients and in 81% in controls (NS), the anterior wall in 19% v 12% (NS), and the inferior wall in 6% v 7%, respectively (NS). The most delayed LV wall was the free wall in 56% of group B patients and in 84% of controls (p < 0.05), the anterior wall in 12% v 4% (p < 0.05), and the inferior wall in 32% v 12% (p < 0.05), respectively (table 3).

Table 2

Echocardiographic Doppler data

Table 3

LV walls with the shortest and the longest electromechanical delays (EMD) in group A and B patients

Figure 2

Left: Distribution of interventricular EMDs according to the QRS width in group B patients. There was no significant correlation between these two variables. An abnormally increased interventricular asynchrony (> 50 ms) was found in 63% of group B patients. Three patients with a QRS width < 200 ms were found to have an abnormally increased interventricular asynchrony (grey diamonds) whereas four patients had a QRS width >200 ms but without interventricular asynchrony (grey circles). Right: Distribution of intra-left ventricular (LV) electromechanical delays according to QRS width. There was no significant correlation between these two variables. An abnormally increased intra-LV asynchrony (> 50 ms) was found in 75% of the patients. Three patients with a QRS width < 200 ms were found to have an abnormally increased intra-LV asynchrony (grey diamonds) whereas three patients had a QRS width >200 ms but without interventricular or intra-LV asynchrony (grey circles).

DISCUSSION

The present study assessed the regional qualitative and quantitative tissue Doppler imaging patterns caused by RVP in patients with and without heart failure. These patterns were graduated in a time scale so that the degree of intra-LV and interventricular electromechanical asynchrony could be defined. Data from the control population permitted us to determine the normal range of RVP induced ventricular asynchrony from which abnormal asynchrony could be identified in the group with heart failure. It is theoretically conceivable that differences in ventricular asynchrony between patients with heart failure and controls may not be caused entirely by heart structural abnormalities. Indeed, 75% of our patients with heart failure (versus none in the control group) were taking β blockers, which may potentially worsen ventricular conduction disturbances. However, the aim of the study was to identify patients with pacemaker dependent heart failure who, despite optimal pharmacological treatment, present with pathological mechanical ventricular asynchrony.

This new way of considering heart failure in paced patients provided useful insights. Firstly, in RVP patients, there is no correlation between electrical features (QRS width) and the degree of electromechanical ventricular asynchrony. Secondly, independent of the presence of heart failure, the last activated LV wall was not always the free wall and the first one activated was not always the septal wall, as has already been shown in patients with spontaneous atrioventricular conduction.5–7,14 Thirdly, in group B more patients had an RVP induced intra-LV than an interventricular asynchrony, suggesting that in these patients the major cause of LV function impairment is likely to be the presence of intra-LV asynchrony. As BVP results in the improvement of intra-LV rather than of interventricular synchrony,5–7 RVP patients who present with an abnormally increased intra-LV asynchrony should benefit from BVP upgrading. Since our data did not show any significant correlation between the degree of intra-LV asynchrony and the QRS width, considering the lower limit of 200 ms of QRS width to be a criterion for upgrading patients from RVP to BVP may not be justified. Indeed, whereas four patients with heart failure (25% of group B) had a QRS width below 200 ms, they all presented with abnormal interventricular or intra-LV asynchrony and should be upgraded to BVP. In contrast, three patients (19% of group B) had a QRS width above 200 ms but had no abnormal ventricular asynchrony compared with the controls and may not be good candidates for BVP despite the large QRS width. In patients with heart failure with a large QRS width, multicentre randomised studies found that about 20% were non-responders even though all the included patients had a wide QRS.1–3 These non-responders may have presented with no important mechanical LV asynchrony despite the large QRS width.

Limitations of the study

The sample size of patients with heart failure was relatively small because these patients had to fit strict inclusion criteria: NYHA functional class III or IV, LV ejection fraction < 35%, sinus rhythm, and a resting heart rate comparable with that of controls. Furthermore, they had to be pacemaker dependent and to have been paced for more than six months.

We used the tissue Doppler imaging technique to quantify the degree of ventricular mechanical asynchrony. Even though this technique has been intensively studied in clinical research,5–6,11,14–16 it is not yet widely used clinically, which limits the application of the method.

Conclusion

The QRS width does not seem to be a good parameter of conduction disturbances or a reliable identifier of paced patients with heart failure likely to benefit from BVP. The need for alternative methods such as new echocardiographic techniques may be required. However, further studies with larger populations are needed to confirm these data.