Article Text

Download PDFPDF
A novel method of assessing Starr-Edwards aortic valve stenosis
  1. R J Edwards,
  2. M Al-Bustami,
  3. C D J Ilsley
  1. Harefield Hospital, Middlesex, UK
  1. Correspondence to:
    Dr Richard J Edwards
    Harefield Hospital, Hill End Road, Harefield, Middlesex UB9 6JH, UK; r.edwardsrbh.nthames.nhs.uk

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Catheter assessment of Starr-Edwards prosthetic aortic valve stenosis requires simultaneous recording of left ventricular (LV) and aortic pressures. Passage of conventional Dacron catheters across mechanical valves can damage the mechanism and induce valvar incompetence. The alternative Bockenbrough procedure requires atrial septal puncture with potential lethal complications.1 The availability of a low profile, floppy tipped, 0.014 inch guidewire mounted pressure sensor (PressureWire 4, RADI Medical Systems, Uppsala, Sweden) provides a novel means of crossing this type of valve. The use of fine bore pressure wires may be less likely to disturb the ball cage mechanism than larger fluid filled catheters, and may provide a safe and reproducible method of assessing Starr-Edwards aortic valve stenosis. However, the use of the pressure wire in assessing bileaflet or tilting disc prosthetic valves may be more problematic.

CASE HISTORY

A 54 year old man presented to clinic with a six month history of worsening shortness of breath. He had had a size 10 A Starr-Edwards aortic valve inserted in 1990 and …

View Full Text