Device complications with addition of defibrillation to cardiac resynchronisation therapy for primary prevention

Abstract

Objective In patients indicated for cardiac resynchronisation therapy (CRT), the choice between a CRT-pacemaker (CRT-P) versus defibrillator (CRT-D) remains controversial and indications in this setting have not been well delineated. Apart from inappropriate therapies, which are inherent to the presence of a defibrillator, whether adding defibrillator to CRT in the primary prevention setting impacts risk of other acute and late device-related complications has not been well studied and may bear relevance for device selection.

Methods Observational multicentre European cohort study of 3008 consecutive patients with ischaemic or non-ischaemic dilated cardiomyopathy and no history of sustained ventricular arrhythmias, undergoing CRT implantation with (CRT-D, n=1785) or without (CRT-P, n=1223) defibrillator. Using propensity score and competing risk analyses, we assessed the risk of significant device-related complications requiring surgical reintervention. Inappropriate shocks were not considered except those due to lead malfunction requiring lead revision.

Results Acute complications occurred in 148 patients (4.9%), without significant difference between groups, even after considering potential confounders (OR=1.20, 95% CI 0.72 to 2.00, p=0.47). During a mean follow-up of 41.4±29 months, late complications occurred in 475 patients, giving an annual incidence rate of 26 (95% CI 9 to 43) and 15 (95% CI 6 to 24) per 1000 patient-years in CRT-D and CRT-P patients, respectively. CRT-D was independently associated with increased occurrence of late complications (HR=1.68, 95% CI 1.27 to 2.23, p=0.001). In particular, when compared with CRT-P, CRT-D was associated with an increased risk of device-related infection (HR 2.10, 95% CI 1.18 to 3.45, p=0.004). Acute complications did not predict overall late complications, but predicted device-related infection (HR 2.85, 95% CI 1.71 to 4.56, p<0.001).

Conclusions Compared with CRT-P, CRT-D is associated with a similar risk of periprocedural complications but increased risk of long-term complications, mainly infection. This needs to be considered in the decision of implanting CRT with or without a defibrillator.

  • implanted cardiac defibrillators
  • pacemakers
  • heart failure
  • quality and outcomes of care

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.