Abstract
Objective To test the hypothesis that in recipients of primary prophylactic implantable cardioverter-defibrillators (ICDs), the non-planarity of ECG vector loops predicts (a) deaths despite ICD protection and (b) appropriate ICD shocks.
Methods Digital pre-implant ECGs were collected in 1948 ICD recipients: 21.4% females, median age 65 years, 61.5% ischaemic heart disease (IHD). QRS and T wave three-dimensional loops were constructed using singular value decomposition that allowed to measure the vector loop planarity. The non-planarity, that is, the twist of the three-dimensional loops out of a single plane, was related to all-cause mortality (n=294; 15.3% females; 68.7% IHD) and appropriate ICD shocks (n=162; 10.5% females; 87.7% IHD) during 5-year follow-up after device implantation. Using multivariable Cox regression, the predictive power of QRS and T wave non-planarity was compared with that of age, heart rate, left ventricular ejection fraction, QRS duration, spatial QRS-T angle, QTc interval and T-peak to T-end interval.
Results QRS non-planarity was significantly (p<0.001) associated with follow-up deaths despite ICD protection with HR of 1.339 (95% CI 1.165 to 1.540) but was only univariably associated with appropriate ICD shocks. Non-planarity of the T wave loop was the only ECG-derived index significantly (p<0.001) associated with appropriate ICD shocks with multivariable Cox regression HR of 1.364 (1.180 to 1.576) but was not associated with follow-up mortality.
Conclusions The analysed data suggest that QRS and T wave non-planarity might offer distinction between patients who are at greater risk of death despite ICD protection and those who are likely to use the defibrillator protection.
- Defibrillators, Implantable
- Biomarkers
- Electrocardiography
Data availability statement
Data are available upon reasonable request. Data are available upon reasonable request pending the approval by the EU-CERT-ICD Steering Committee.
This is an open access article distributed in accordance with the Creative Commons Attribution 4.0 Unported (CC BY 4.0) license, which permits others to copy, redistribute, remix, transform and build upon this work for any purpose, provided the original work is properly cited, a link to the licence is given, and indication of whether changes were made. See: https://creativecommons.org/licenses/by/4.0/.
Statistics from Altmetric.com
WHAT IS ALREADY KNOWN ON THIS TOPIC
In normal physiological recordings, the vectorcardiographic loops of the QRS complex and of the T wave are known to be practically planar, that is, with no or only little three-dimensional twist. The non-planarity of the vectorcardiographic loops has previously been observed mainly in patients with ischaemic heart disease. Nevertheless, the extent of the twist of these loops has not been investigated in recipients of implantable cardioverter-defibrillators (ICDs) implanted for primary prophylactic reasons.
WHAT THIS STUDY ADDS
The spatial twist of the vectorcardiographic loops of the QRS complex and of the T wave is measurable not only based on standard 12-lead ECGs but also based on restricted electrode sets suitable for prolonged monitoring.
In primary prophylactic ICD recipients, the extent of the twist of the QRS complex and of the T wave was found to be an independent predictor of risk.
The twist of the QRS complex was found to predict both all-cause mortality and, to a lesser extent, appropriate ICD shocks.
The twist of the T wave strongly predicted appropriate ICD shocks.
HOW THIS STUDY MIGHT AFFECT RESEARCH, PRACTICE OR POLICY
The results of the study need to be replicated in independent datasets. If the results are confirmed in prospectively collected ECGs of primary prophylactic ICD recipients, the predictive value of the QRS and T wave spatial twists needs to be assessed in terms of the distinction between patients who do and do not benefit from ICD primary prophylaxis. Such investigations might eventually lead to a change in the selection of patients for ICD primary prophylaxis.
Introduction
QRS micro-fragmentation was recently proposed to characterise depolarisation abnormalities beyond the visual detection on standard 12-lead ECGs.1 This characteristic was shown to provide an independent mortality predictor in different populations.2 It was proposed that the QRS micro-fragmentation expresses localised irregularities of ventricular excitation and that these aberrations signify heart failure including the early subclinical stages.
The essential concept of micro-fragmentation assessment is based on the analysis of simultaneously recorded ECG leads (ie, of the eight mutually independent signals of the standard ECG recording). The use of standard clinical ECGs makes the QRS micro-fragmentation analysis widely applicable to different clinical settings. Nevertheless, reliance on multiple ECG leads makes it inapplicable to situations when fewer ECG leads are available, for example, monitoring systems, standard clinical Holters and wearable ECG devices.
It has previously been reported that in physiological ECG recordings, the vectorcardiographic (VCG) loops of the QRS complex are essentially planarly,3 that is, that the three-dimensional QRS dipole moves practically in a single plane the orientation of which depends on the position of the organ. Similar observations were also made for the VCG loop of the T wave.4
Our understanding of the predictive power of QRS micro-fragmentation suggests that aberrations of the depolarisation sequence might also influence the three-dimensional QRS loop and twist it outside a single two-dimensional plane. Since the VCG loops might be, in principle, constructed from as few as three independent ECG leads, the non-planarity characteristics might be derived from limited electrode sets.
Being guided by these considerations, we tested the hypotheses that QRS complex and T wave non-planarity indices are factors predicting death despite defibrillator protection and appropriate shock therapy by implantable cardioverter-defibrillators (ICDs). These hypotheses were tested in the previously reported population of recipients of prophylactic ICDs collected within the retrospective part of the EU-CERT-ICD Study.5
Methods
Population and follow-up data
As already published,2 5 the EU-CERT-ICD Study included a retrospective part that collected data of ICD recipients implanted in different European centres for primary prophylactic reasons between 2000 and 2014. Baseline characteristics, including age at ICD implantation, pre-implantation left ventricular ejection fraction (LVEF), and the distinction between ischaemic and non-ischaemic heart disease, were collected at each centre. Follow-up data were also provided by participating centres and were quality controlled by the team of University Hospital of Basel, Switzerland.
In each case, the ICD programming corresponded to the clinical needs and to the standard practice of each centre. The follow-up data provided by individual centres included all-cause mortality and ICD shocks that were adjudicated to differentiate between appropriate and inappropriate shocks. For the purposes of this investigation, all-cause mortality and appropriate ICD shocks were used as two separate follow-up event categories. Time of survival was defined as the interval between the ICD implantation and death; patients who did not die were censored at the end of the follow-up by the relevant centre. For patients who experienced an appropriate ICD shock, the interval between the device implantation and the first such shock was considered. As with all-cause mortality, patients who did not experience an appropriate ICD shock were censored (for the purposes of ICD shock prediction) at the time of their death or at the follow-up end. For the purposes of this investigation, follow-up was restricted to the first 5 years after ICD implantation.
ECG recordings
Electronic 12-lead short-term ECGs were obtained in 1948 patients prior to ICD implantation (median 1 day before implantation, IQR 1–6 days). These patients constituted the population of this study and were collected and followed up at the Department of Cardiology, University Hospital of Basel, Switzerland (n=488); Department of Cardiology and Pneumology, University Medical Center, Göttingen, Germany (n=441); Department of Cardiovascular Sciences, University of Leuven, Belgium (n=361); University Central Hospital of Oulu, Finland (n=32); and Department of Medical Physiology, University Medical Center Utrecht, the Netherlands (n=626). The ECG recordings obtained at Oulu were of 8-second duration, other ECGs were of 10-second duration.
ECG measurement
As previously reported,2 the signals of all the ECGs were converted to the same digital format, cubic spline resampled (where appropriate) to 1 kHz frequency and filtered. Automatic QRS detection was visually confirmed and used to construct representative median beatforms of each ECG lead. For each ECG, these representative beatforms of different leads were superimposed on the same isoelectric axis and previously described algorithms were used to detect QRS onset, QRS offset and T wave offset. In each ECG, the delineation positions were visually checked and, where appropriate, manually corrected using a computer display with a single millisecond precision.
Using these representative beatform delineations, QRS complex and QT interval durations were obtained. Heart rate was derived from the average of all RR intervals in the entire ECG. Using this heart rate, rate-corrected QTc intervals were derived by Fridericia formula. The T-peak to T-end (TpTe) intervals were measured in the vector magnitude of all 12 leads using a previously reported method.6 Using also previously published technique,4 spatial QRS-T angles were measured and expressed in degrees (between 0° and 180°).
Assessment of QRS and T wave planarity
The main plane of ECG vector movement was defined by the means of singular value decomposition (SVD) of the ECG signals. In the same way as previously used for the QRS micro-fragmentation expression, the SVD decomposing signals were obtained.1 The first two components (see the online supplemental material for details) defined the main plane and the two-dimensional vector loop movement within this plane. The third component (orthogonal to the first two) expressed the contribution of the signal components that twisted the vector loop outside the main plane (figure 1). The relative extent (percentage) of the third component expressed the non-planarity (the twist) of the three-dimensional vector loop (figure 1).
Supplemental material
Using this principle, two pairs of assessments of the planarity ECG components were computed. First, all algebraically independent leads of the source ECG (that is, leads I, II and V1–V6) were used to apply the SVD algorithm to the QRS complex (the signal between the verified QRS onset and offset) and to the T wave (the signal between the QRS offset and T wave offset).
Second, to model the situation of ECGs with restricted leads, only signals between electrodes V1 and V6 were considered. Potential differences between four electrodes defined three leads, namely V2-V1, V5-V2 and V6-V5 approximating signal acquisition by a chest belt. These three derived leads were subsequently processed by SVD and the same approach as described expressed the QRS and T wave planarity values.
Statistics and data presentation
Continuous data are presented as medians (IQR). Non-parametric Kruskal-Wallis, Kolmogorov-Smirnov and χ2 tests were used for group comparisons of continuous and categorical data, respectively. Non-parametric Spearman correlation coefficients were used to assess and test pairwise associations between continuous variables. Association of variables with outcome variables was tested by Cox regression analysis which was used both with single variables and for multivariable modelling with backwards stepwise elimination. For the purposes of Cox modelling of continuous variables, the QRS complex and T wave non-planarity values were logarithmically transformed. In addition to Cox regression analysis using continuous variables, models with dichotomised variables were used. For this purpose, age was dichotomised at 75 years, heart rate at 75 beats/min, LVEF at 25% (close to the population median), QRS duration at 120 ms7, QTc interval at 450 ms8, TpTe interval at 100 ms9 and QRS-T angle at 110°.10 The QRS and T wave non-planarity values were dichotomised at their population medians. Two different datasets were used for multivariable Cox regression analysis. Model 1 tested available variables against the QRS complex and T wave non-planarity values derived from all original eight independent ECG leads; model 2 used QRS complex and planarity values derived from the modelled chest belt of electrodes V1, V2, V5 and V6. The association of the dichotomised non-planarity values with outcome was also tested by Kaplan-Meier survival curves; the differences between the curves were tested by log rank test.
SVD computation and the assessment of non-planarity of QRS complex and of T wave were programmed in C++ (Microsoft Visual Studio Professional 2022, 64-bit V.17.3.5). Statistical evaluation was performed by SPSS package (V.27; IBM Corporation); p values below 0.05 were considered statistically significant.
Additional analyses are described in the online supplemental material.
Results
Of the 1948 patients of the study, 294 (15.1%) died during the follow-up restricted to the first 5 years. Defibrillators with cardiac resynchronisation therapy (CRT) function were implanted in 797 patients. In 57 patients (2.9%), the information on ICD shock therapy was not available. Among the remaining 1897 patients, 207 (10.9%) experienced an appropriate ICD shock during the first 5 years of follow-up. Of these, 45 patients (21.7% of patients who received a shock) subsequently died during the first 5 years of follow-up. Clinical characteristics of the population and the measured non-planarity values of ECG components are shown in table 1. Differences between non-planarity values in patients with and without follow-up events are shown in figure 2.
QRS non-planarity values assessed from original 12-lead ECGs and from the modelled chest belt electrodes were significantly correlated (Spearman’s r=0.455, p<0.001). The same was true for the T wave non-planarity (r=0.429, p<0.001). QRS non-planarity values were also correlated with the QRS duration (r=0.204 and 0.189, both p<0.01, for the values derived from the original ECGs and from the chest belt electrodes, respectively). No significant correlations were found between the T wave non-planarity values and the QT interval duration (r=0.012 and r=−0.004, for the two sets of values, respectively). The correlations between the QRS non-planarity and T wave non-planarity were very modest (r=0.154 and r=0.202, respectively).
Table 2 shows the results of the Cox regression analysis based on the continuous variables. The departure from planarity of the QRS complex loop was found to be a strong predictor of all-cause mortality independent of other factors used in the model. While QRS duration and QTc duration were significantly associated with mortality univariably, they were not found significant in a multivariable analysis. Both the departures from QRS planarity assessed in the complete 12-lead ECG and in the precordial band predicted the mortality strongly.
Interestingly, the departure from the planarity of the T wave loop (regardless of whether assessed in the complete ECG or in the precordial band) was the only significant predictor of appropriate ICD shocks that was not eliminated in the multivariable models. QRS planarity assessed from the complete 12-lead ECG was also a univariable predictor of ICD shocks but, in a multivariable analysis, it was eliminated when the T wave loop non-planarity was included.
Consistent results were found with Cox regression models based on dichotomised variables (table 3). When including the distinction between CRT/non-CRT defibrillators, only increased heart rate and QRS non-planarity survived in the multivariable models of mortality prediction.
As expected, female sex was strongly associated with the absence of ICD shocks which remained in the multivariable models when the absence of planarity of the T wave (assessed in either way) was added to the model. Univariably but not multivariably, non-ischaemic heart disease predicted fewer ICD shocks. Interestingly, the TpTe interval above 100 ms was also found significantly associated with the ICD shocks but, contrary to expectations, it predicted fewer rather than more frequent shocks. It did not remain significant in multivariable models.
Figures 3 and 4 show the distinction of Kaplan-Meier curves of probability of follow-up events when dividing the population according to QRS complex and T wave loop planarity. All the distinctions were consistent with the univariable results in table 2.
Additional results are described in the online supplemental material.
Discussion
The analyses show that the non-planarity of the three-dimensional QRS complex and T wave loops provides risk assessment independently of other risk factors.
Physiologically, these observations make sense. Depolarisation abnormalities have repeatedly been associated with poorer survival of cardiac patients11–13 and QRS complex abnormalities have been reported to predict heart failure complications more strongly compared with the arrhythmia prediction. Conversely, repolarisation anomalies have repeatedly been associated with arrhythmic complications and sudden cardiac death.14–16
The measurement of QRS non-planarity is mathematically independent of the assessment of QRS micro-fragmentation.1 We have indeed seen occasional ECGs in which minimal QRS non-planarity was combined with substantial micro-fragmentation and vice versa. Nevertheless, when both QRS non-planarity and QRS micro-fragmentation were used in Cox regression models predicting mortality, the non-planarity was eliminated in the multivariable analysis irrespective of whether continuous or dichotomised variables were used (details not shown). It thus seems that QRS non-planarity, while technically independent, provides less powerful detection of depolarisation abnormalities, although it might be assessed from a limited number of ECG leads. It remains to be seen whether in other populations, QRS non-planarity would significantly contribute to mortality risk predicted by QRS micro-fragmentation.
Since the T wave non-planarity was the only ECG-related parameter that significantly predicted ICD shocks, more advanced analyses of multilead T wave signals of standard ECGs are of interest. Nevertheless, when applying the algorithm of QRS micro-fragmentation to the T wave analysis, we have not obtained any significant prediction of mortality and/or of ICD shocks (details not shown). This is not surprising since propagation of myocardial repolarisation changes is influenced by intercellular electronic interactions17 that eliminate abnormalities that would be detectable by the micro-fragmentation analyses. Nevertheless, different analyses of T wave signals might classify pro-arrhythmic abnormalities of the repolarisation sequence more powerfully compared with the simple non-planarity assessment.
Although the measurements of the QRS and T wave non-planarity in full 12-lead ECGs and in the modelled chest belt were significantly correlated, numerical differences were noticeable. This is not surprising. The SVD analysis represents the multilead ECG signal by a multidimensional vector movement which is a simplification of the electrophysiological processes in the complete organ. Irrespective of this simplification, analyses of both electrode configurations led to similar predictions of follow-up events.
The four electrodes that we selected to model a chest belt are not arranged along a true circle which is of advantage when studying three-dimensional ECG signal properties. The three leads that we obtained from these electrodes were selected arbitrarily since any other combinations (eg, V1–V2, V1–V5 and V1–V6) are only algebraic combinations of the used leads and would thus provide the same SVD decomposition results.
Multiple studies investigated the planarity and non-planarity of the QRS complex. The planarity was assessed mainly in VCG using different expressions including the length, width, thickness, and thickness/length and width/length ratios of the QRS loop in different spatial projections.18 The QRS non-planarity has previously been investigated mainly in relation to ischaemic heart disease and myocardial infarction.3 18 19 Although about 60% of the patients of the study suffered from ischaemic heart disease, multivariable Cox regression analysis showed that the predictive value of QRS loop non-planarity was independent of heart disease aetiology. T wave loop abnormalities have also been repeatedly studied using different characteristics.4 20 Among others, T wave-based prediction of mortality in ischaemic heart disease was repeatedly reported.21–23 Interestingly, in pilot data that preceded the EU-CERT-ICD Study, Seegers et al24 reported that pre-implantation T wave area predicted appropriate ICD shocks. Of these studies, T wave loop planarity was sparsely investigated, mainly in ischaemic heart disease.18
Naturally, every retrospective analysis can only be hypothesis generating. Nevertheless, if our findings are independently confirmed in independent datasets, the assessment of QRS and T loop planarity might be of assistance when considering ICD implantation in patients at the border of criteria of the ICD guidelines. Patients in whom a flat QRS complex loop is combined with a twisted T wave loop might be stronger candidates for ICD prophylaxis compared with patients in whom the planarity of the QRS complex and T wave loops appears to be the other way round. Still, independent confirmations of this possibility are needed before any clinical utility might be proposed.
In future ECG studies, combination of the non-planarity indices of ECG loops needs to be investigated together with other ECG factors25 including those that were derived from long-term recordings26 which were not available for the retrospective EU-CERT-ICD dataset. Analyses of longer recordings, applicable to both monitoring systems and ECG wearables, would also allow to address the variability and intrasubject reproducibility27 of the non-planarity indices.
Limitations
The available data were limited in several aspects; distinction between cardiac and non-cardiac, and sudden and non-sudden death was not available. While it might be assumed that among prophylactic ICD recipients, mortality was mainly cardiac/cardiovascular, the lack of distinction between sudden and thus presumably arrhythmic and non-sudden deaths prevented us from some more detailed investigations, for example, of the MADIT ICD benefit score.28 It might only be assumed that in patients under ICD protection, most cardiac deaths are related to heart failure. CRT utilisation data are not available. Having only one ECG recording in each patient did not allow us to study measurement reproducibility. In the Cox regression analysis, we have intentionally included mostly parameters that might be obtained from ECG analysis. Additional clinical findings, for example, electrophysiological investigation (not available in the investigated dataset), might outperform the risk prediction by the reported indices, although they would not have the advantage of undemanding data acquisition. Finally, these retrospective results have not been tested prospectively. We plan to perform such a prospective evaluation once the follow-up data of the prospective part of the EU-CERT-ICD Study29 have been extended; this follow-up extension is presently ongoing.
Conclusions
Despite these limitations, the study shows that the assessment of QRS complex and T wave loop planarity might also be obtained from limited electrode sets. This makes the assessment plausible also based on data from monitoring systems and wearable devices. In patients with ICD implanted for prophylactic reasons, the estimates of QRS complex and T wave loop planarity appear to differ between those who do and do not use the cardioversion function as well those who do and do not survive despite the ICD anti-tachycardia protection.
Supplemental material
Supplemental material
Data availability statement
Data are available upon reasonable request. Data are available upon reasonable request pending the approval by the EU-CERT-ICD Steering Committee.
Ethics statements
Patient consent for publication
Ethics approval
This study involves human participants and separate ethics approval permissions were obtained by each of the centres contributing the ECG recordings and follow-up data to this study (committees and reference numbers: Basel Ethikkommission Nordwest-und Zentralschweiz (PB_2017-00138); Göttingen Ethikkommission der medizinischen Fakulät Universitätsmedizin Göttingen (22/7/13); Leuven EC University Hospitals Leuven (S55991); Oulu Northern Ostrobothnian hospital district (N/A); Utrecht Medisch Ethische Toetsingscommissie, Utrecht (WAG/mb/20/011346)). All patients included in the retrospective data collection provided written informed consent to have their data researched.
References
Supplementary materials
Supplementary Data
This web only file has been produced by the BMJ Publishing Group from an electronic file supplied by the author(s) and has not been edited for content.
Footnotes
Contributors KH—conceptual design of the study, ECG processing algorithms, software coding and computer implementation of ECG analyses, algorithmic measurements of relevant ECG intervals, application of the analytical methods to the EU-CERT-ICD WP02 database of ECG signals, interpretation of results, manuscript draft, manuscript revisions, approval of individual manuscript versions; member of the EU-CERT-ICD Study team. IA—conceptual design of the study, comments to the algorithmic measurements of ECG intervals, visual verification of the measurements of relevant ECG intervals, manual correction of the measurements of relevant ECG intervals, quality control of ECG measurements, interpretation of results, manuscript draft, approval of individual manuscript versions. TN—comments to the algorithmic measurements of ECG intervals, visual verification of the measurements of relevant ECG intervals, manual correction of the measurements of relevant ECG intervals, quality control of ECG measurements, interpretation of results, manuscript draft, approval of individual manuscript versions; member of the EU-CERT-ICD Study team. BV—collection of ECG signals contributed by the Leuven team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for study analyses, quality control of the Leuven part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, follow-up of the Leuven patients submitted to the retrospective WP02 part of EU-CERT-ICD, comments to the manuscript, approval of the final manuscript; member of the EU-CERT-ICD Study team. DS—collection of ECG signals contributed by the Utrecht team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for study analyses, quality control of the Utrecht part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, follow-up of the Utrecht patients submitted to the retrospective WP02 part of EU-CERT-ICD, comments to the manuscript, approval of the final manuscript; member of the EU-CERT-ICD Study team. JJ—collection of ECG signals contributed by the Oulu team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for study analyses, quality control of the Oulu part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, comments to the algorithmic measurements of ECG intervals, follow-up of the Oulu patients submitted to the retrospective WP02 part of EU-CERT-ICD, quality control of the Oulu part of the EU-CERT-ICD WP02 database, comments to the manuscript, approval of the final manuscript; member of the EU-CERT-ICD Study team. TR—collection of ECG signals contributed by the Basel team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for study analyses, quality control of the Basel part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, follow-up of the Basel patients submitted to the retrospective WP02 part of EU-CERT-ICD, comments to the manuscript, approval of the final manuscript. SS—collection of ECG signals contributed by the Göttingen team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for study analyses, quality control of the Göttingen part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, follow-up of the Göttingen patients submitted to the retrospective part of EU-CERT-ICD, comments to the manuscript, approval of the final manuscript; member of the EU-CERT-ICD Study team. MAV—collection of ECG signals contributed by the Utrecht team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for study analyses, quality control of the Utrecht part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, follow-up of the Utrecht patients submitted to the retrospective WP02 part of EU-CERT-ICD, quality control of the Utrecht part of the EU-CERT-ICD WP02 database, comments to the manuscript, approval of the final manuscript; member of the EU-CERT-ICD Study team. TF—statistical aspects of the study design concept, statistical consultations, models of the EU-CERT-ICD WP02 data collection, quality control of the complete EU-CERT-ICD WP02 database, quality control of the ECG links to the EU-CERT-ICD WP02 database, approval of the statistical methods of the data analyses used in the manuscript, comments to the manuscript, approval of the final manuscript; member of the EU-CERT-ICD Study team. AB—concepts of ECG data analyses, physiological interpretation of ECG analyses and of their measurement results, design and completeness control of the data analyses, interpretation of results, manuscript draft, manuscript revisions, approval of the final manuscript; member of the EU-CERT-ICD Study team. HVH—collection of ECG signals contributed by the Oulu team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for study analyses, quality control of the Oulu part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, comments to the initial algorithmic measurements of ECG intervals, quality control of the initial algorithmic measurements of ECG intervals, follow-up of the Oulu patients submitted to the retrospective part of EU-CERT-ICD, quality control of the Oulu part of the EU-CERT-ICD WP02 database, comments to the manuscript, approval of the final manuscript; member of the EU-CERT-ICD Study team. RW—collection of ECG signals contributed by the Leuven team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for study analyses, quality control of the Leuven part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, comments to the initial algorithmic measurements of ECG intervals, quality control of the initial algorithmic measurements of ECG intervals, follow-up of the Leuven patients submitted to the retrospective part of EU-CERT-ICD, quality control of the Leuven part of the EU-CERT-ICD WP02 database, comments to the manuscript, approval of the final manuscript; member of the EU-CERT-ICD Study team. GS—physiological interpretation of the concepts of ECG analyses, physiological interpretation of the measurement results, completeness control of the data analyses, interpretation of results, manuscript draft, manuscript revisions, approval of the final manuscript; member of the EU-CERT-ICD Study team. CS—responsibility for the work package 02 (retrospective data) of EU-CERT-ICD, collection of ECG signals contributed by the Basel team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for study analyses, quality control of the Basel part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, follow-up of the Basel patients submitted to the retrospective WP02 part of EU-CERT-ICD, quality control of the complete EU-CERT-ICD WP02 database, comments to the manuscript, approval of the final manuscript; member of the EU-CERT-ICD Study team. MZ—overall responsibility for the EU-CERT-ICD Study, collection of ECG signals contributed by the Göttingen team to the ECG part of the WP02 database of EU-CERT-ICD, submission of these ECGs for the study analyses, quality control of the Göttingen part of the EU-CERT-ICD WP02 ECGs after their conversion into a common open format, follow-up of the Göttingen patients submitted to the retrospective part of EU-CERT-ICD, quality control of the Göttingen part of the EU-CERT-ICD WP02 database, comments to the manuscript, approval of the final manuscript; chair of the EU-CERT-ICD Study team. MM—proposal of the concept of the ECG analysis, sequential structure of ECG analyses, conceptual design of the study, ECG processing algorithms, software coding and computer implementation of ECG analyses, quality control of the algorithmic measurements of relevant ECG intervals, quality control of the manual adjustments of the measurements of relevant ECG intervals, application of the analytical methods to the EU-CERT-ICD W02 database of ECG signals, interpretation of results, manuscript draft, interpretation and team reporting of review comments, manuscript revisions, approval of individual manuscript versions, study and article guarantor; member of the EU-CERT-ICD Study team.
Funding The EU-CERT-ICD Study was funded by the European Community’s 7th Framework Programme (HEALTH-F2-2009-602299). The ECG analyses were supported in part by the British Heart Foundation (NH/16/2/32499), by the Ministry of Health, Czech Republic, conceptual development of research organisation (FNBr/65269705) and by the Specific Research of Masaryk University (MUNI/A/1412/2022).
Competing interests None declared.
Patient and public involvement Patients and/or the public were not involved in the design, or conduct, or reporting, or dissemination plans of this research.
Provenance and peer review Not commissioned; externally peer reviewed.
Supplemental material This content has been supplied by the author(s). It has not been vetted by BMJ Publishing Group Limited (BMJ) and may not have been peer-reviewed. Any opinions or recommendations discussed are solely those of the author(s) and are not endorsed by BMJ. BMJ disclaims all liability and responsibility arising from any reliance placed on the content. Where the content includes any translated material, BMJ does not warrant the accuracy and reliability of the translations (including but not limited to local regulations, clinical guidelines, terminology, drug names and drug dosages), and is not responsible for any error and/or omissions arising from translation and adaptation or otherwise.