RT Journal Article SR Electronic T1 Cost-effectiveness of a risk-stratified approach to cardiac resynchronisation therapy defibrillators (high versus low) at the time of generator change JF Heart JO Heart FD BMJ Publishing Group Ltd and British Cardiovascular Society SP 416 OP 422 DO 10.1136/heartjnl-2017-311749 VO 104 IS 5 A1 Simon Claridge A1 Frederic A Sebag A1 Steven Fearn A1 Jonathan M Behar A1 Bradley Porter A1 Tom Jackson A1 Benjamin Sieniewicz A1 Justin Gould A1 Jessica Webb A1 Zhong Chen A1 Mark O’Neill A1 Jaswinder Gill A1 Christophe Leclercq A1 Christopher A Rinaldi YR 2018 UL http://heart.bmj.com/content/104/5/416.abstract AB Objective Responders to cardiac resynchronisation therapy whose device has a defibrillator component and who do not receive a therapy in the lifetime of the first generator have a very low incidence of appropriate therapy after box change. We investigated the cost implications of using a risk stratification tool at the time of generator change resulting in these patients being reimplanted with a resynchronisation pacemaker.Methods A decision tree was created using previously published data which had demonstrated an annualised appropriate defibrillator therapy risk of 2.33%. Costs were calculated at National Health Service (NHS) national tariff rates (2016–2017). EQ-5D utility values were applied to device reimplantations, admissions and mortality data, which were then used to estimate quality-adjusted life-years (QALYs) over 5 years.Results At 5 years, the incremental cost of replacing a resynchronisation defibrillator device with a second resynchronisation defibrillator versus resynchronisation pacemaker was £5045 per patient. Incremental QALY gained was 0.0165 (defibrillator vs pacemaker), resulting in an incremental cost-effectiveness ratio (ICER) of £305 712 per QALYs gained. Probabilistic sensitivity analysis resulted in an ICER of £313 612 (defibrillator vs pacemaker). For reimplantation of all patients with a defibrillator rather than a pacemaker to yield an ICER of less than £30 000 per QALY gained (current NHS cut-off for approval of treatment), the annual arrhythmic event rate would need to be 9.3%. The budget impact of selective replacement was a saving of £2 133 985 per year.Conclusions Implanting low-risk patients with a resynchronisation defibrillator with the same device at the time of generator change is not cost-effective by current NHS criteria. Further research is required to understand the impact of these findings on individual patients at the time of generator change.