TY - JOUR T1 - Primary transcatheter umbrella closure of perimembranous ventricular septal defect. JF - British Heart Journal JO - Heart SP - 368 LP - 371 DO - 10.1136/hrt.72.4.368 VL - 72 IS - 4 AU - M L Rigby AU - A N Redington Y1 - 1994/10/01 UR - http://heart.bmj.com/content/72/4/368.abstract N2 - OBJECTIVES--The starting hypothesis was that some perimembranous ventricular septal defects can be closed safely and effectively with a Bard Rashkind double umbrella introduced through a long transvenous sheath. DESIGN--A descriptive study of all patients who underwent attempted transcatheter umbrella closure of a perimembranous ventricular septal defect. Those patients selected for the study had symptoms of a ventricular septal defect and a perimembranous ventricular septal defect shown by transthoracic echocardiography. The morphological criteria used were a posterior perimembranous defect with a diameter of < or = 8 mm not associated with overriding of the aortic or pulmonary valve or with aortic valve prolapse. The haemodynamic criteria for inclusion in the study were a right to left ventricular systolic pressure ratio of > 0.45, a Doppler derived right ventricular systolic pressure of > 50 mm Hg, and a pulmonary to systemic flow ratio > 3:1. SETTING--A tertiary referral centre. PATIENTS--13 infants, children, and adolescents with a perimembranous ventricular septal defect aged 3 weeks to 16 years and weighing 1.8-46 kg. INTERVENTIONS--A modified Rashkind ductal double umbrella was introduced through a long transvenous sheath and positioned on either side of the ventricular septal defect. Placement was guided by transoesophageal echocardiography. RESULTS--10 out of 13 patients underwent successful partial or complete closure of a perimembranous ventricular septal defect. There were three placement failures. Two of these were associated with a ventricular septal defect too large for the umbrella device. In a third case the umbrella was opened in the left ventricular outflow tract necessitating surgical removal and closure of the ventricular septal defect. CONCLUSION--Transcatheter umbrella closure of a perimembranous ventricular septal defect is technically feasible and can be therapeutically successful, although the procedure is moderately difficult to perform and the mean procedure time is > 120 minutes. It is an alternative to surgery in some cases, but the overall results would not support its routine use even with the introduction of larger devices of the current design. ER -