RT Journal Article SR Electronic T1 Intraoperative apical ventricular septal defect closure using a modified Rashkind double umbrella. JF Heart JO Heart FD BMJ Publishing Group Ltd and British Cardiovascular Society SP 367 OP 369 DO 10.1136/hrt.76.4.367 VO 76 IS 4 A1 R. R. Chaturvedi A1 D. F. Shore A1 M. Yacoub A1 A. N. Redington YR 1996 UL http://heart.bmj.com/content/76/4/367.abstract AB OBJECTIVE: To demonstrate the safety and efficacy of intraoperative apical ventricular septal defect (VSD) closure using a modified Rashkind double umbrella. DESIGN: Descriptive study of all patients in whom intraoperative device closure of apical VSDs was attempted. SETTING: A tertiary referral centre. PATIENTS: Four patients with an apical VSD requiring closure, during the period January 1993 to May 1995. INTERVENTIONS: Intraoperative placement of a modified Rashkind umbrella. RESULTS: Four successful placements resulting in apical VSD closure, as judged by transoesophageal colour flow mapping. Three patients received a 17 mm and one a 12 mm umbrella. Early complete closure was achieved in three patients. There was a small residual leak around the 12 mm device that had resolved at 5 month follow up. There was one early death, which was unrelated to VSD closure. CONCLUSION: Apical ventricular septal defects can be closed safely and effectively with intraoperative use of a modified Rashkind umbrella.