RT Journal Article SR Electronic T1 Inappropriate shocks delivered by implantable cardiac defibrillators during oversensing of activity of diaphagmatic muscle JF Heart JO Heart FD BMJ Publishing Group Ltd and British Cardiovascular Society SP 94 OP 96 DO 10.1136/hrt.81.1.94 VO 81 IS 1 A1 D Babuty A1 L Fauchier A1 P Cosnay YR 1999 UL http://heart.bmj.com/content/81/1/94.abstract AB Two cases are reported (both men, one 72 and one 54 years old) of inappropriate shocks delivered by an implantable cardiac defibrillator (ICD) device, which oversensed the myopotentials induced by deep breathing and Valsalva manoeuvre. No damage to leads was associated with the oversensing of myopotentials. The mechanism of the inappropriate shocks was determined using real time electrograms. Modification of the duration of ventricular detection and decrease in sensitivity made it possible to avoid the oversensing of myopotentials and to deliver ICD treatment.