RT Journal Article SR Electronic T1 Biocorrosion of magnesium alloys: a new principle in cardiovascular implant technology? JF Heart JO Heart FD BMJ Publishing Group Ltd and British Cardiovascular Society SP 651 OP 656 DO 10.1136/heart.89.6.651 VO 89 IS 6 A1 Heublein, B A1 Rohde, R A1 Kaese, V A1 Niemeyer, M A1 Hartung, W A1 Haverich, A YR 2003 UL http://heart.bmj.com/content/89/6/651.abstract AB Objectives: To develop and test a new concept of the degradation kinetics of newly developed coronary stents consisting of magnesium alloys.Methods: Design of a coronary stent prototype consisting of the non-commercial magnesium based alloy AE21 (containing 2% aluminium and 1% rare earths) with an expected 50% loss of mass within six months. Eleven domestic pigs underwent coronary implantation of 20 stents (overstretch injury).Results: No stent caused major problems during implantation or showed signs of initial breakage in the histological evaluation. There were no thromboembolic events. Quantitative angiography at follow up showed a significant (p < 0.01) 40% loss of perfused lumen diameter between days 10 and 35, corresponding to neointima formation seen on histological analysis, and a 25% re-enlargement (p < 0.05) between days 35 and 56 caused by vascular remodelling (based on intravascular ultrasound) resulting from the loss of mechanical integrity of the stent. Inflammation (p < 0.001) and neointimal plaque area (p < 0.05) depended significantly on injury score. Planimetric degradation correlated with time (r = 0.67, p < 0.01).Conclusion: Vascular implants consisting of magnesium alloy degradable by biocorrosion seem to be a realistic alternative to permanent implants.