%0 Journal Article %A J-S Kim %A I-K Jang %A J-S Kim %A T H Kim %A M Takano %A T Kume %A N W Hur %A Y-G Ko %A D Choi %A M-K Hong %A Y Jang %T Optical coherence tomography evaluation of zotarolimus-eluting stents at 9-month follow-up: comparison with sirolimus-eluting stents %D 2009 %R 10.1136/hrt.2009.167759 %J Heart %P 1907-1912 %V 95 %N 23 %X Objective: To evaluate the vascular response at 9 months after zotarolimus-eluting stent (ZES; Endeavor) implantation using optical coherence tomography (OCT). These findings were compared with those after implantation of a sirolimus-eluting stent (SES; Cypher Select).Design: Cross-sectional observational study with prospective OCT registry.Setting: Nine months after ZES or SES implantation.Patients and methods: A total of 68 patients (32 ZES and 36 SES) underwent OCT at 9 months after stent implantation. The neointima hyperplasia (NIH) thickness inside each strut and percentage of NIH area at every 1 mm cross section were measured.Main outcome measurement: The degree of neointimal coverage and the prevalence of malapposition at 9 months after ZES and SES implantation using OCT.Results: The mean (SD) NIH thickness (251.2 (110.0) μm vs 85.5 (53.3) μm, p<0.001) and percentage of NIH area (27.9 (9.1)% vs 11.2 (7.1)%, p<0.001) were significantly greater in ZES than in SES. The prevalence of uncovered strut as well as malapposed strut was significantly lower in ZES than in SES (0.3% vs 12.3%, p<0.001 and 0.08% vs 2.6%, p<0.001). Thrombus was not observed in ZES (0.0% in ZES vs 27.8% in SES, p = 0.001).Conclusions: Neointimal coverage in ZES was almost complete and malapposition was very rare at 9-months’ follow-up. %U https://heart.bmj.com/content/heartjnl/95/23/1907.full.pdf