Table 1

Device sizing, arterial injury scores, and metal coverage for all the sections from each group of very small (2 mm) porcine coronary arteries undergoing intervention

BiodivYsio SVNIR SVBiodivYsio standardPOBA
Numbers are mean (SEM).
*p<0.05; **p<0.01 compared with BiodivYsio SV.
NA, not applicable; POBA, plain old balloon angioplasty; SV, small vessel.
Artery diameter (mm)2.08 (0.02)2.09 (0.02)2.01 (0.02)*2.02 (0.04)
Balloon diameter (mm)2.50 (0.00)2.50 (0.00)2.50 (0.00)2.50 (0.00)
Balloon/artery ratio1.20 (0.01)1.20 (0.01)1.25 (0.01)*1.24 (0.02)
Schwartz score1.15 (0.15)1.43 (0.12)1.48 (0.10)**NA
Metal coverage (%)31 (2)40 (2)48 (2)**NA