Table 4 Subanalysis of radiation dose estimates for patients receiving dual-source CT at different heart rate levels
Radiation parametersTube voltage (kV)
120 (n = 45)120 (n = 36)120 (n = 14)100 (n = 16)100 (n = 6)100 (n = 3)
ECG pulsing window (%)70–7070–7070–7070–7070–7070–70
Heart rate (bpm)<6060–70>70<6060–70>70
Mean scan length (mm)135 (26)134 (22)112 (22)123 (10)120 (11)127 (12)
Mean DLP (mGy*cm)594 (259)537 (128)461 (120)237 (84)205 (131)162 (14)
Mean CTDIvol (mGy)35 (13)33 (8)25 (8)15 (5)10 (2)9 (1)
Estimated mean radiation dose (mSv)10.1 (4.4)9.1 (2.2)7.8 (2.2)4.0 (1.5)3.5 (2.5)3.1 (1.1)
  • Results are shown as mean (SD) unless stated otherwise.

  • Minimum tube current for dose estimates calculated is 4% of full tube current level.

  • CTDIvol, volume computed tomography dose index; DLP, dose–length product.