Table 2

Univariate predictors of appropriate ICD therapy after generator replacement

VariableRate of events, n/N (%)HR (95% CI)p Value
PresentAbsent
Ischaemic cardiomyopathy21/157 (13)8/109 (7)2.02 (0.90 to 4.57)0.09
LVEF <25%*10/82 (12)12/143 (8)1.48 (0.64 to 3.42)0.36
Renal insufficiency7/59 (12)22/207 (11)1.02 (0.43 to 2.40)0.96
Diabetes mellitus5/55 (9)24/211 (11)0.82 (0.31 to 2.15)0.69
CRT-ICD11/116 (9)18/150 (12)0.77 (0.36 to 1.62)0.49
NYHA ≥III9/106 (8)20/160 (12.5)0.66 (0.30 to 1.44)0.29
Advanced age (age >70 years at replacement)4/68 (6)25/198 (13)0.49 (0.17 to 1.41)0.19
  • CRT, cardiac resynchronisation therapy; ICD, implantable cardioverter-defibrillator; NYHA, New York Heart Association.

  • *LVEF data were available for 225 patients.