STAND | HAND* | |
---|---|---|
Cost (US$) | 60 000–100 000 | 7000 |
Weight (kg) | 10–15 | 0.390 |
Size | Large laptop | Large smart-phone |
Display resolution (pixels) | 800×600 | 320×240 |
Image sector | 90° | 75° |
Colour Doppler | Yes (adjustable scale) | Yes (fixed scale) |
Spectral Doppler | Yes | No |
Transducers | Multiple | Single |
Frequency | Multiple | Single |
Battery charge time (min) | 120 | 60 |
Battery life (scan time) (min) | 45 | 90 |
Overheat tendency | Rare | Common |
Patient information | Multiple data fields | Voice file only |
Measurements | Multiple (linear, volume, Doppler) | Linear only—cumbersome |
File format | Any, DICOM compatible | .jpg, .mpg, .wav, not DICOM compatible |
File size (MB) | 25–100/study | 1–2/study |
File transfer | Wired, wireless, manual, network | Wired, manual |
Offline viewing | Compatible with any DICOM viewer | Propriety software |
*GE VScan—only handheld echocardiography machine used in published RHD studies.