Table 4

Intraoperator and interoperator reproducibility and time efficiency for E/e’

Intraoperator reproducibility, n=50Interoperator reproducibility, n=18Time taken (s) to select and measure E/e’ (95% CI), n=18
Bias (limits of agreement)ICC (95% CI)Bias (limits of agreement)ICC (95% CI)
Single index-beat−0.2 (−4.2 to 3.9)0.91 (0.84 to 0.95)−0.3 (−2.9 to 2.2)0.94 (0.87 to 0.98)35.4 (33.1 to 37.8)
3 consecutive beats−0.7 (−6.2 to 4.8)0.74 (0.60 to 0.85)−1.1 (−5.4 to 3.2)0.83 (0.64 to 0.93)Not performed
5 consecutive beats−0.6 (−5.3 to 4.2)0.80 (0.68 to 0.88)−1.1 (−6.5 to 4.2)0.78 (0.56 to 0.91)44.7 (41.8 to 47.5)
10 consecutive beats−0.4 (−4.2 to 3.4)0.88 (0.80 to 0.93)−0.9 (−6.1 to 4.2)0.82 (0.62 to 0.93)98.1 (91.7 to 104.4)
  • E/e’, mitral E wave max/average diastolic tissue Doppler velocity from the septal and lateral annulus; ICC, intraclass correlation coefficient.